Using a Scanner to Measure Absorbed Doses with Radiochromic Film Dosimeters
The article provides a sequence of steps for using RISØ calorimeters for calibration and subsequent use of B3 radiochromic film dosimeters (GEX corporation) and a scanner for measuring absorbed doses. Calibration was carried out with the help of electron beam accelerator in the range of absorbed do...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
V.N. Karazin Kharkiv National University Publishing
2022-03-01
|
Series: | East European Journal of Physics |
Subjects: | |
Online Access: | https://periodicals.karazin.ua/eejp/article/view/18444 |
Summary: | The article provides a sequence of steps for using RISØ calorimeters for calibration and subsequent use of B3 radiochromic film
dosimeters (GEX corporation) and a scanner for measuring absorbed doses. Calibration was carried out with the help of electron
beam accelerator in the range of absorbed doses of 3 – 40 kGy (measurement range of RISØ calorimeters).
In the course of the work, the following was carried out:
– calibration of B3 radiochromic dosimetry films using RISØ calorimeters;
– plotting a calibration curve for B3 radiochromic dosimetric films;
– calculation of approximation functions;
– development of a technique for using a flatbed scanner to measure absorbed doses;
– estimation of the measurement uncertainties of absorbed doses.
Accelerator operation parameters: scanning frequency of the accelerated electron beam – 5 Hz, pulse frequency – 120 Hz, electron
energy – 5 MeV, electron beam current – 60 μA. The measurement error of the absorbed dose is 5.8 %.
|
---|---|