The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University

Introduction: A screen-film system has been replaced by computed radiography (CR) system in Thailand. The advantages of CR are that it provides faster process and better contrast resolution. The major role of CR system is that it allows user to define an imaging technique which does not only provide...

Full description

Bibliographic Details
Main Authors: Banjong Kheonkaew, Wichai Wichatorntrakul, Eimorn Mairiang, Panatsda Awikhunprasert, Khathayuth Nikapruek
Format: Article
Language:English
Published: Chaing Mai University 2014-01-01
Series:Journal of Associated Medical Sciences
Subjects:
Online Access:https://www.tci-thaijo.org/index.php/bulletinAMS/article/view/59986
_version_ 1818928500184186880
author Banjong Kheonkaew
Wichai Wichatorntrakul
Eimorn Mairiang
Panatsda Awikhunprasert
Khathayuth Nikapruek
author_facet Banjong Kheonkaew
Wichai Wichatorntrakul
Eimorn Mairiang
Panatsda Awikhunprasert
Khathayuth Nikapruek
author_sort Banjong Kheonkaew
collection DOAJ
description Introduction: A screen-film system has been replaced by computed radiography (CR) system in Thailand. The advantages of CR are that it provides faster process and better contrast resolution. The major role of CR system is that it allows user to define an imaging technique which does not only provide good image quality, but also reduce the patient radiation dose as the ALARA principle. Radiation exposure index is a numerical value recommended by a manufacturer (manufacturer recommended range; MRR). As for the international requirements, this value is used to remind a radiologic technician to determine an appropriate radiation exposure for the imaging. An excessive radiation exposure can cause an unnecessary radiation dose. Conversely, insufficient radiation exposure can cause a noise on the image which reduces the image quality. This might cause an adversely effect on the diagnosis. Objectives: The purpose of this study was to investigate whether the exposure indicators of posterior–anterior computed radiography chest are in the MRR ranges and also whether higher or not. Materials and methods: The retrospective study was performed in 1005 patients at Srinagarind Hospital, aged above 16 years old from June 2012 to March 2013. The S values measured from Fuji FCR 5000 on the PA radiographic chest were recorded. Results: The results showed that only 69 percent of all the images were in the range specified by the manufacturer (S value 200-600). Fifteen percent of the measures were greater than the MRR ranges (S value >600) which is defined as underexposure. Sixteen percent of the S-values were less than the MRR range (S value <600) which is defined as overexposure. It was considered that the radiation exposure tends to increase. Conclusions: The results of the study demonstrated that an accuracy of the exposure index calibration is required as the number can correctly represent the amount of radiation used to create the image and the quality of image. In order to reduce the radiation dose on patients and obtain a high image quality, the radiation exposure charts are required to determine an appropriate radiation exposure for different organs. Bull Chiang Mai Assoc Med Sci 2014; 47(1): 23-29
first_indexed 2024-12-20T03:29:53Z
format Article
id doaj.art-01c256f323f648b1b7e10ba35e341f92
institution Directory Open Access Journal
issn 2539-6056
2539-6056
language English
last_indexed 2024-12-20T03:29:53Z
publishDate 2014-01-01
publisher Chaing Mai University
record_format Article
series Journal of Associated Medical Sciences
spelling doaj.art-01c256f323f648b1b7e10ba35e341f922022-12-21T19:55:01ZengChaing Mai UniversityJournal of Associated Medical Sciences2539-60562539-60562014-01-01471232359986The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen UniversityBanjong KheonkaewWichai WichatorntrakulEimorn MairiangPanatsda AwikhunprasertKhathayuth NikapruekIntroduction: A screen-film system has been replaced by computed radiography (CR) system in Thailand. The advantages of CR are that it provides faster process and better contrast resolution. The major role of CR system is that it allows user to define an imaging technique which does not only provide good image quality, but also reduce the patient radiation dose as the ALARA principle. Radiation exposure index is a numerical value recommended by a manufacturer (manufacturer recommended range; MRR). As for the international requirements, this value is used to remind a radiologic technician to determine an appropriate radiation exposure for the imaging. An excessive radiation exposure can cause an unnecessary radiation dose. Conversely, insufficient radiation exposure can cause a noise on the image which reduces the image quality. This might cause an adversely effect on the diagnosis. Objectives: The purpose of this study was to investigate whether the exposure indicators of posterior–anterior computed radiography chest are in the MRR ranges and also whether higher or not. Materials and methods: The retrospective study was performed in 1005 patients at Srinagarind Hospital, aged above 16 years old from June 2012 to March 2013. The S values measured from Fuji FCR 5000 on the PA radiographic chest were recorded. Results: The results showed that only 69 percent of all the images were in the range specified by the manufacturer (S value 200-600). Fifteen percent of the measures were greater than the MRR ranges (S value >600) which is defined as underexposure. Sixteen percent of the S-values were less than the MRR range (S value <600) which is defined as overexposure. It was considered that the radiation exposure tends to increase. Conclusions: The results of the study demonstrated that an accuracy of the exposure index calibration is required as the number can correctly represent the amount of radiation used to create the image and the quality of image. In order to reduce the radiation dose on patients and obtain a high image quality, the radiation exposure charts are required to determine an appropriate radiation exposure for different organs. Bull Chiang Mai Assoc Med Sci 2014; 47(1): 23-29https://www.tci-thaijo.org/index.php/bulletinAMS/article/view/59986Exposure indexS valueCR
spellingShingle Banjong Kheonkaew
Wichai Wichatorntrakul
Eimorn Mairiang
Panatsda Awikhunprasert
Khathayuth Nikapruek
The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
Journal of Associated Medical Sciences
Exposure index
S value
CR
title The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
title_full The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
title_fullStr The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
title_full_unstemmed The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
title_short The assessment of exposure index on the posterior-anterior computed radiography chest in Srinagarind Hospital, Khon Kaen University
title_sort assessment of exposure index on the posterior anterior computed radiography chest in srinagarind hospital khon kaen university
topic Exposure index
S value
CR
url https://www.tci-thaijo.org/index.php/bulletinAMS/article/view/59986
work_keys_str_mv AT banjongkheonkaew theassessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT wichaiwichatorntrakul theassessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT eimornmairiang theassessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT panatsdaawikhunprasert theassessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT khathayuthnikapruek theassessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT banjongkheonkaew assessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT wichaiwichatorntrakul assessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT eimornmairiang assessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT panatsdaawikhunprasert assessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity
AT khathayuthnikapruek assessmentofexposureindexontheposterioranteriorcomputedradiographychestinsrinagarindhospitalkhonkaenuniversity