Dynamic determination of equivalent CT source models for personalized dosimetry
With improvements in CT technology, the need for reliable patient-specific dosimetry increased in the recent years. The accuracy of Monte-Carlo simulations for absolute dose estimation is related to scanner specific information on the X-ray spectra of the scanner as well as the form filter geometrie...
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
De Gruyter
2017-09-01
|
Series: | Current Directions in Biomedical Engineering |
Subjects: | |
Online Access: | https://doi.org/10.1515/cdbme-2017-0167 |
Summary: | With improvements in CT technology, the need for reliable patient-specific dosimetry increased in the recent years. The accuracy of Monte-Carlo simulations for absolute dose estimation is related to scanner specific information on the X-ray spectra of the scanner as well as the form filter geometries and compositions. In this work a mobile measurement setup is developed, which allows both to determine the X-ray spectra and equivalent form filter of a specific scanner from just one helical scan in less than 2 minutes. |
---|---|