Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics

The aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosth...

Full description

Bibliographic Details
Main Authors: Andrea Giovanni Cutti, Maria Grazia Santi, Andrew H. Hansen, Stefania Fatone
Format: Article
Language:English
Published: MDPI AG 2024-04-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/24/7/2350
_version_ 1797211919893921792
author Andrea Giovanni Cutti
Maria Grazia Santi
Andrew H. Hansen
Stefania Fatone
author_facet Andrea Giovanni Cutti
Maria Grazia Santi
Andrew H. Hansen
Stefania Fatone
author_sort Andrea Giovanni Cutti
collection DOAJ
description The aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosthetics. Four limb models with different shapes were fabricated and scanned with a metrological 3D scanner (EINScan Laser FreeScan 5X, SHINING 3D Tech. Co., Ltd., Hangzhou, China) by a professional operator (OP0). Limb models were then mailed to three sites where two operators (OP1, OP2) scanned them using their own structured-light 3D scanner (same model). OP1 scanned limb models twice (OP1-A, OP1-B). OP0, OP1-A, and OP2 scans were compared for accuracy, OP1-A and OP1-B for repeatability, and OP1-A and OP2 for reproducibility. Among all comparisons, the mean radial error was <0.25 mm, mean angular error was <4°, and root mean square error of the radial distance was <1 mm. Moreover, limits of agreement were <3.5% for perimeters and volumes. By comparing these results with respect to clinically-relevant thresholds and to the literature available on other 3D scanners, we conclude that the EINScan Pro 2X Plus 3D Scanner with High Definition Prime Pack has good accuracy, repeatability, and reproducibility, supporting its use in multi-site settings.
first_indexed 2024-04-24T10:34:09Z
format Article
id doaj.art-05a0c879c7c8436e971c8a3ef0a85afc
institution Directory Open Access Journal
issn 1424-8220
language English
last_indexed 2024-04-24T10:34:09Z
publishDate 2024-04-01
publisher MDPI AG
record_format Article
series Sensors
spelling doaj.art-05a0c879c7c8436e971c8a3ef0a85afc2024-04-12T13:26:50ZengMDPI AGSensors1424-82202024-04-01247235010.3390/s24072350Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb ProstheticsAndrea Giovanni Cutti0Maria Grazia Santi1Andrew H. Hansen2Stefania Fatone3INAIL, Via Rabuina 14, Vigorso di Budrio, 40054 Bologna, ItalyDepartment of Industrial Engineering, University of Padova, 35131 Padova, ItalyMinneapolis VA Health Care System, One Veterans Drive, Minneapolis, MN 55417, USANUOPC, Department of Physical Medicine and Rehabilitation, Northwestern University, 680 N Lake Shore Dr, Suite 1100, Chicago, IL 60611, USAThe aim of this work was to assess the accuracy, repeatability, and reproducibility of a hand-held, structured-light 3D scanner (EINScan Pro 2X Plus with High Definition Prime Pack, SHINING 3D Tech. Co., Ltd., Hangzhou, China), to support its potential use in multi-site settings on lower limb prosthetics. Four limb models with different shapes were fabricated and scanned with a metrological 3D scanner (EINScan Laser FreeScan 5X, SHINING 3D Tech. Co., Ltd., Hangzhou, China) by a professional operator (OP0). Limb models were then mailed to three sites where two operators (OP1, OP2) scanned them using their own structured-light 3D scanner (same model). OP1 scanned limb models twice (OP1-A, OP1-B). OP0, OP1-A, and OP2 scans were compared for accuracy, OP1-A and OP1-B for repeatability, and OP1-A and OP2 for reproducibility. Among all comparisons, the mean radial error was <0.25 mm, mean angular error was <4°, and root mean square error of the radial distance was <1 mm. Moreover, limits of agreement were <3.5% for perimeters and volumes. By comparing these results with respect to clinically-relevant thresholds and to the literature available on other 3D scanners, we conclude that the EINScan Pro 2X Plus 3D Scanner with High Definition Prime Pack has good accuracy, repeatability, and reproducibility, supporting its use in multi-site settings.https://www.mdpi.com/1424-8220/24/7/2350structured-light 3D scannerhand-held 3D scanneraccuracyrepeatabilityreproducibilityamputation
spellingShingle Andrea Giovanni Cutti
Maria Grazia Santi
Andrew H. Hansen
Stefania Fatone
Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
Sensors
structured-light 3D scanner
hand-held 3D scanner
accuracy
repeatability
reproducibility
amputation
title Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
title_full Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
title_fullStr Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
title_full_unstemmed Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
title_short Accuracy, Repeatability, and Reproducibility of a Hand-Held Structured-Light 3D Scanner across Multi-Site Settings in Lower Limb Prosthetics
title_sort accuracy repeatability and reproducibility of a hand held structured light 3d scanner across multi site settings in lower limb prosthetics
topic structured-light 3D scanner
hand-held 3D scanner
accuracy
repeatability
reproducibility
amputation
url https://www.mdpi.com/1424-8220/24/7/2350
work_keys_str_mv AT andreagiovannicutti accuracyrepeatabilityandreproducibilityofahandheldstructuredlight3dscanneracrossmultisitesettingsinlowerlimbprosthetics
AT mariagraziasanti accuracyrepeatabilityandreproducibilityofahandheldstructuredlight3dscanneracrossmultisitesettingsinlowerlimbprosthetics
AT andrewhhansen accuracyrepeatabilityandreproducibilityofahandheldstructuredlight3dscanneracrossmultisitesettingsinlowerlimbprosthetics
AT stefaniafatone accuracyrepeatabilityandreproducibilityofahandheldstructuredlight3dscanneracrossmultisitesettingsinlowerlimbprosthetics