Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V

Objectives: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA1c) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. Methods: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 guide...

Full description

Bibliographic Details
Main Authors: José María Maesa, Patricia Fernández-Riejos, Catalina Sánchez Mora, María de Toro, Paloma Menéndez Valladares, Concepción González-Rodriguez
Format: Article
Language:English
Published: Elsevier 2016-08-01
Series:Practical Laboratory Medicine
Online Access:http://www.sciencedirect.com/science/article/pii/S2352551716300154
_version_ 1819152520877965312
author José María Maesa
Patricia Fernández-Riejos
Catalina Sánchez Mora
María de Toro
Paloma Menéndez Valladares
Concepción González-Rodriguez
author_facet José María Maesa
Patricia Fernández-Riejos
Catalina Sánchez Mora
María de Toro
Paloma Menéndez Valladares
Concepción González-Rodriguez
author_sort José María Maesa
collection DOAJ
description Objectives: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA1c) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. Methods: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 guidelines. We selected 100 samples from the routine laboratory workload and analyzed them in duplicate with the three analyzers. The imprecision study was performed according to CLSI EP5-A2 guidelines for both inter-assay and intra-assay variability. Bias was assessed with external quality control material. To establish linearity, CLSI EP6-A protocol was followed. Results: Method comparison (95% confidence intervals in parentheses): D-100 vs G8: Passing-Bablok regression; y=0.973(0.963–0.983)×−0.07(−0.07−0.069); r=0.9989. Bland-Altman mean difference: −0.229%HbA1c (−0.256: −0.202); Relative bias plot: D-100/G8 vs D100-G8 mean ratio=0.971(0.967−0.975). D-100 vs HA-8180V: Passing-Bablok regression; y=0.944(0.932–0.958)×−0.078(0.024−0.173); r=0.9989. Bland-Altman mean difference: −0.363%HbA1c (−0.401: −0.325); Relative bias plot D-100/HA-8180V vs D100-HA-8180V mean ratio=0.955(0.952−0.958). Inter-assay coefficient of variation (CV): 0.81%. Intra-assay CV: 1.04% (low level), and 0.78% (high level). Bias against target value=2.332%. Linearity: r2=0.998 in the concentration range 4.4−13.9%HbA1c. Carry-over: 0.0024%. Conclusions: The Bio-Rad D-100 shows good correlation with G8 and HA-8180V. There is a small proportional systematic difference (2.7% and 5.6%, respectively) in both comparisons. Inter and intra-assay CVs are both lower than the lowest CV obtained in studies performed with D-100 and other instruments. Keywords: Glycated hemoglobin, High performance liquid chromatography, Analyzer, Intra-assay variability, Inter-assay variability
first_indexed 2024-12-22T14:50:36Z
format Article
id doaj.art-05bb801199d74dfb82d89fd0e9a0335d
institution Directory Open Access Journal
issn 2352-5517
language English
last_indexed 2024-12-22T14:50:36Z
publishDate 2016-08-01
publisher Elsevier
record_format Article
series Practical Laboratory Medicine
spelling doaj.art-05bb801199d74dfb82d89fd0e9a0335d2022-12-21T18:22:19ZengElsevierPractical Laboratory Medicine2352-55172016-08-0155764Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180VJosé María Maesa0Patricia Fernández-Riejos1Catalina Sánchez Mora2María de Toro3Paloma Menéndez Valladares4Concepción González-Rodriguez5Clinic Biochemistry Department, Virgen Macarena University Hospital, SpainCorresponding author.; Clinic Biochemistry Department, Virgen Macarena University Hospital, SpainClinic Biochemistry Department, Virgen Macarena University Hospital, SpainClinic Biochemistry Department, Virgen Macarena University Hospital, SpainClinic Biochemistry Department, Virgen Macarena University Hospital, SpainClinic Biochemistry Department, Virgen Macarena University Hospital, SpainObjectives: To evaluate the Bio-Rad D-100®, an HPLC analyzer for glycated hemoglobin (HbA1c) determination, and to compare its performance with the Menarini HA-8180V® and Sysmex G8®. Methods: Method comparison was performed according to Clinical and Laboratory Standards Institute (CLSI) EP9-A2 guidelines. We selected 100 samples from the routine laboratory workload and analyzed them in duplicate with the three analyzers. The imprecision study was performed according to CLSI EP5-A2 guidelines for both inter-assay and intra-assay variability. Bias was assessed with external quality control material. To establish linearity, CLSI EP6-A protocol was followed. Results: Method comparison (95% confidence intervals in parentheses): D-100 vs G8: Passing-Bablok regression; y=0.973(0.963–0.983)×−0.07(−0.07−0.069); r=0.9989. Bland-Altman mean difference: −0.229%HbA1c (−0.256: −0.202); Relative bias plot: D-100/G8 vs D100-G8 mean ratio=0.971(0.967−0.975). D-100 vs HA-8180V: Passing-Bablok regression; y=0.944(0.932–0.958)×−0.078(0.024−0.173); r=0.9989. Bland-Altman mean difference: −0.363%HbA1c (−0.401: −0.325); Relative bias plot D-100/HA-8180V vs D100-HA-8180V mean ratio=0.955(0.952−0.958). Inter-assay coefficient of variation (CV): 0.81%. Intra-assay CV: 1.04% (low level), and 0.78% (high level). Bias against target value=2.332%. Linearity: r2=0.998 in the concentration range 4.4−13.9%HbA1c. Carry-over: 0.0024%. Conclusions: The Bio-Rad D-100 shows good correlation with G8 and HA-8180V. There is a small proportional systematic difference (2.7% and 5.6%, respectively) in both comparisons. Inter and intra-assay CVs are both lower than the lowest CV obtained in studies performed with D-100 and other instruments. Keywords: Glycated hemoglobin, High performance liquid chromatography, Analyzer, Intra-assay variability, Inter-assay variabilityhttp://www.sciencedirect.com/science/article/pii/S2352551716300154
spellingShingle José María Maesa
Patricia Fernández-Riejos
Catalina Sánchez Mora
María de Toro
Paloma Menéndez Valladares
Concepción González-Rodriguez
Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
Practical Laboratory Medicine
title Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
title_full Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
title_fullStr Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
title_full_unstemmed Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
title_short Evaluation of Bio-Rad D-100 HbA1c analyzer against Tosoh G8 and Menarini HA-8180V
title_sort evaluation of bio rad d 100 hba1c analyzer against tosoh g8 and menarini ha 8180v
url http://www.sciencedirect.com/science/article/pii/S2352551716300154
work_keys_str_mv AT josemariamaesa evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v
AT patriciafernandezriejos evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v
AT catalinasanchezmora evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v
AT mariadetoro evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v
AT palomamenendezvalladares evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v
AT concepciongonzalezrodriguez evaluationofbioradd100hba1canalyzeragainsttosohg8andmenariniha8180v