Two-Channel Detecting Sensor with Signal Cross-Correlation for FTIR Instruments
This paper’s purpose was to demonstrate a performance of a novel approach in a low-noise optical sensor for an FTIR spectrometer. Methods: Compared to the standard FTIR detection setup, our sensor ensures a higher signal-to-noise ratio (SNR) and lower signal standard deviation by reducing the uncorr...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2022-11-01
|
Series: | Sensors |
Subjects: | |
Online Access: | https://www.mdpi.com/1424-8220/22/22/8919 |
Summary: | This paper’s purpose was to demonstrate a performance of a novel approach in a low-noise optical sensor for an FTIR spectrometer. Methods: Compared to the standard FTIR detection setup, our sensor ensures a higher signal-to-noise ratio (SNR) and lower signal standard deviation by reducing the uncorrelated noise components (e.g., thermal and 1/<i>f</i> noises of the detection module). Its construction is based on two-channel detection modules and a processing unit with implemented cross-correlation signal analyses. Each module was built of LWIR HgCdTe photodiodes and low-noise transimpedance amplifiers. Results: the experiments demonstrated a decrease in a signal standard deviation of about 1.7 times with a 10 dB-improvement in the SNR. Conclusion: this result indicates our sensor’s main benefit, especially in registered “weak” and noisy interferograms. |
---|---|