Rapid assessment of iron in blood plasma and serum by spectrophotometry with cloud-point extraction [version 1; referees: 2 approved]
Rapid photometric assessment of iron in blood plasma and serum by a simple procedure after the extraction of iron(II) complex with 1-nitroso-2-naphthol in the micellar phase of a nonionic surfactant at the cloud point upon heating (pH range is 4.5–6.3) is proposed. The procedure trueness was verifie...
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
F1000 Research Ltd
2015-08-01
|
Series: | F1000Research |
Subjects: | |
Online Access: | http://f1000research.com/articles/4-623/v1 |
Summary: | Rapid photometric assessment of iron in blood plasma and serum by a simple procedure after the extraction of iron(II) complex with 1-nitroso-2-naphthol in the micellar phase of a nonionic surfactant at the cloud point upon heating (pH range is 4.5–6.3) is proposed. The procedure trueness was verified using a standard reference protocol using bathophenanthroline. The advantages of the procedure are higher sensitivity than the reference protocol: the limit of detection is 0.03 μg/mL, the limit of quantitation is 0.1 μg/mL, the determination range is 0.1 – 2.8 μg/mL (RSD 0.02–0.10). Copper does not interfere with the iron assessment. |
---|---|