Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study
Background: Measuring implant stability is an important issue in predicting treatment success. Dental implant stability is usually measured through resonance frequency analysis (RFA). Osstell® RFA devices can be used with transducers (Smartpeg™) that correspond to the implants used as well as with...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
College of Dentistry/ University of Baghdad
2023-09-01
|
Series: | Journal of Baghdad College of Dentistry |
Subjects: | |
Online Access: | https://jbcd.uobaghdad.edu.iq/index.php/jbcd/article/view/3455 |
_version_ | 1797683247090499584 |
---|---|
author | Alamin Y Dhahi Salwan Y Bede Haig Khachadourian |
author_facet | Alamin Y Dhahi Salwan Y Bede Haig Khachadourian |
author_sort | Alamin Y Dhahi |
collection | DOAJ |
description |
Background: Measuring implant stability is an important issue in predicting treatment success. Dental implant stability is usually measured through resonance frequency analysis (RFA). Osstell® RFA devices can be used with transducers (Smartpeg™) that correspond to the implants used as well as with transducers designed for application with Penguin® RFA devices (Multipeg™). Aims: This study aims to assess the reliability of a MultiPeg™ transducer with an Osstell® device in measuring dental implant stability. Materials and Methods: Sixteen healthy participants who required dental implant treatment were enrolled in this study. Implant stability was measured by using an Osstell® device with two transducers, namely, Smartpeg™ and Multipeg™. Insertion torque was also measured and recorded as >50 and ≤50 N·cm. Unpaired t-test and Mann–Whitney U test were conducted to assess the relationships of the implant stability values obtained by the two transducers with insertion torque, whereas Pearson and Spearman's correlations were utilized to investigate correlations between the two transducers. Interclass correlation coefficients were applied to assess the reliability between the two transducers. Results: Implant stability measurements (primary and secondary) showed strong positive correlations between Smartpeg™ and Multipeg™. The reliability values between both transducers in primary and secondary implant stability measurements were 0.922 and 0.981, respectively. The use of both transducers revealed higher implant stability measurements for implants inserted with insertion torque > 50 N·cm than those inserted with insertion torque ≤ 50 N·cm. Conclusions: This study demonstrated that the Multipeg™ transducer is reliable in measuring the stability of dental implants using an Osstell® device.
|
first_indexed | 2024-03-12T00:11:47Z |
format | Article |
id | doaj.art-1354c068ac8b4826acecaaf2c361e87c |
institution | Directory Open Access Journal |
issn | 1817-1869 2311-5270 |
language | English |
last_indexed | 2024-03-12T00:11:47Z |
publishDate | 2023-09-01 |
publisher | College of Dentistry/ University of Baghdad |
record_format | Article |
series | Journal of Baghdad College of Dentistry |
spelling | doaj.art-1354c068ac8b4826acecaaf2c361e87c2023-09-15T22:06:06ZengCollege of Dentistry/ University of BaghdadJournal of Baghdad College of Dentistry1817-18692311-52702023-09-0135310.26477/jbcd.v35i3.3455Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical studyAlamin Y Dhahi0Salwan Y Bede1Haig Khachadourian2Department of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, IraqDepartment of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, IraqSpecialist Oral Surgeon, Lead Dentist, Primary Health Care Corporation, Doha, Qatar Background: Measuring implant stability is an important issue in predicting treatment success. Dental implant stability is usually measured through resonance frequency analysis (RFA). Osstell® RFA devices can be used with transducers (Smartpeg™) that correspond to the implants used as well as with transducers designed for application with Penguin® RFA devices (Multipeg™). Aims: This study aims to assess the reliability of a MultiPeg™ transducer with an Osstell® device in measuring dental implant stability. Materials and Methods: Sixteen healthy participants who required dental implant treatment were enrolled in this study. Implant stability was measured by using an Osstell® device with two transducers, namely, Smartpeg™ and Multipeg™. Insertion torque was also measured and recorded as >50 and ≤50 N·cm. Unpaired t-test and Mann–Whitney U test were conducted to assess the relationships of the implant stability values obtained by the two transducers with insertion torque, whereas Pearson and Spearman's correlations were utilized to investigate correlations between the two transducers. Interclass correlation coefficients were applied to assess the reliability between the two transducers. Results: Implant stability measurements (primary and secondary) showed strong positive correlations between Smartpeg™ and Multipeg™. The reliability values between both transducers in primary and secondary implant stability measurements were 0.922 and 0.981, respectively. The use of both transducers revealed higher implant stability measurements for implants inserted with insertion torque > 50 N·cm than those inserted with insertion torque ≤ 50 N·cm. Conclusions: This study demonstrated that the Multipeg™ transducer is reliable in measuring the stability of dental implants using an Osstell® device. https://jbcd.uobaghdad.edu.iq/index.php/jbcd/article/view/3455resonance frequency analysis, osstell, smartpeg, multipeg |
spellingShingle | Alamin Y Dhahi Salwan Y Bede Haig Khachadourian Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study Journal of Baghdad College of Dentistry resonance frequency analysis, osstell, smartpeg, multipeg |
title | Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study |
title_full | Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study |
title_fullStr | Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study |
title_full_unstemmed | Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study |
title_short | Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study |
title_sort | reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device osstell r an observational clinical study |
topic | resonance frequency analysis, osstell, smartpeg, multipeg |
url | https://jbcd.uobaghdad.edu.iq/index.php/jbcd/article/view/3455 |
work_keys_str_mv | AT alaminydhahi reliabilityofthemultipegtransducerinmeasuringdentalimplantstabilitybyusingaresonancefrequencyanalysisdeviceosstellanobservationalclinicalstudy AT salwanybede reliabilityofthemultipegtransducerinmeasuringdentalimplantstabilitybyusingaresonancefrequencyanalysisdeviceosstellanobservationalclinicalstudy AT haigkhachadourian reliabilityofthemultipegtransducerinmeasuringdentalimplantstabilitybyusingaresonancefrequencyanalysisdeviceosstellanobservationalclinicalstudy |