Could radiofrequency echographic multispectrometry (REMS) overcome the overestimation in BMD by dual-energy X-ray absorptiometry (DXA) at the lumbar spine?

Abstract Background Osteoarthritis (OA) and vertebral fractures at the lumbar spine lead to an overestimation of bone mineral density (BMD). Recently, a new approach for osteoporosis diagnosis, defined as radiofrequency echographic multi-spectrometry (REMS), represents an innovative diagnostic tool...

Full description

Bibliographic Details
Main Authors: Carla Caffarelli, Maria Dea Tomai Pitinca, Antonella Al Refaie, Michela De Vita, Simone Catapano, Stefano Gonnelli
Format: Article
Language:English
Published: BMC 2022-05-01
Series:BMC Musculoskeletal Disorders
Subjects:
Online Access:https://doi.org/10.1186/s12891-022-05430-6
Description
Summary:Abstract Background Osteoarthritis (OA) and vertebral fractures at the lumbar spine lead to an overestimation of bone mineral density (BMD). Recently, a new approach for osteoporosis diagnosis, defined as radiofrequency echographic multi-spectrometry (REMS), represents an innovative diagnostic tool that seems to be able to investigate bone quality and provide an estimation of fracture risk independent of BMD. The aim of this paper was to evaluate whether the use of REMS technology can favor the diagnosis of osteoporosis in subjects with an apparent increase in BMD. Methods In a cohort of 159 postmenopausal (66.2 ± 11.6 yrs) women with overestimated BMD by DXA at the lumbar spine, we performed an echographic scan with the REMS technique. Results The mean values of BMD at different skeletal sites obtained by the DXA and REMS techniques showed that the BMD T-scores by REMS were significantly lower than those obtained by the DXA technique both at the lumbar spine (p < 0.01) and at all femoral subregions (p < 0.05). In OA subjects, the percentage of women classified as “osteoporotic” on the basis of BMD by REMS was markedly higher with respect to those classified by DXA (35.1% vs 9.3%, respectively). Similarly, the REMS allows a greater number of fractured patients to be classified as osteoporotic than DXA (58.7% vs 23.3%, respectively). Conclusions REMS technology by the analysis of native raw unfiltered ultrasound signals appears to be able to overcome the most common artifacts, such as OA and vertebral fracture of the lumbar spine, which affect the value of BMD by DXA.
ISSN:1471-2474