Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device

The aim of this paper is to analyze in depth the coupling between leakage electromagnetic fields (EMFs) generated by an electric vehicle wireless power transfer (EV-WPT) device under misaligned operating conditions and metallic coronary stents, which could be potentially hazardous to human electroma...

Full description

Bibliographic Details
Main Authors: Tianhao Wang, Bo Li, Kaifeng Zhao, Quanyi Yu, Linlin Xu, Yaodan Chi, Shanshan Guan
Format: Article
Language:English
Published: MDPI AG 2023-10-01
Series:Electronics
Subjects:
Online Access:https://www.mdpi.com/2079-9292/12/20/4231
_version_ 1797574015955501056
author Tianhao Wang
Bo Li
Kaifeng Zhao
Quanyi Yu
Linlin Xu
Yaodan Chi
Shanshan Guan
author_facet Tianhao Wang
Bo Li
Kaifeng Zhao
Quanyi Yu
Linlin Xu
Yaodan Chi
Shanshan Guan
author_sort Tianhao Wang
collection DOAJ
description The aim of this paper is to analyze in depth the coupling between leakage electromagnetic fields (EMFs) generated by an electric vehicle wireless power transfer (EV-WPT) device under misaligned operating conditions and metallic coronary stents, which could be potentially hazardous to human electromagnetic safety. In this paper, we established that a standing human with a coronary stent implant and a sitting human with a coronary stent implant are exposed to the leaked EMFs of an EV-WPT device with a transmission power of 11 kW and 22 kW and a transmission frequency of 85 kHz, and we quantified the induced <i>E</i>-field strength of the human body by considering the <i>x</i>- and <i>z</i>-axis misalignment ranges of the WPT device as [−75 mm, 75 mm]. The results showed that the enhancement of the induced <i>E</i>-field in the human tissues near the sharp edges of the implant is significant. Meanwhile, the larger the WPT device misalignment generated, the higher the transmission power, and the closer the human body is to the WPT device, the larger the induced <i>E</i>-field<sub>max</sub> value of the human body is, which has the possibility of exceeding the ICNIRP safety limit, and there is a potential human body electromagnetic safety problem.
first_indexed 2024-03-10T21:17:14Z
format Article
id doaj.art-14f3faeb09da45a9ba5b8fa9f256d828
institution Directory Open Access Journal
issn 2079-9292
language English
last_indexed 2024-03-10T21:17:14Z
publishDate 2023-10-01
publisher MDPI AG
record_format Article
series Electronics
spelling doaj.art-14f3faeb09da45a9ba5b8fa9f256d8282023-11-19T16:18:42ZengMDPI AGElectronics2079-92922023-10-011220423110.3390/electronics12204231Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer DeviceTianhao Wang0Bo Li1Kaifeng Zhao2Quanyi Yu3Linlin Xu4Yaodan Chi5Shanshan Guan6The College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaThe College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaThe College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaThe College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaThe College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaJilin Provincial Key Laboratory of Architectural Electricity and Comprehensive Energy Saving, Jilin Jianzhu University, Changchun 130118, ChinaThe College of Instrumentation and Electrical Engineering, Jilin University, Changchun 130026, ChinaThe aim of this paper is to analyze in depth the coupling between leakage electromagnetic fields (EMFs) generated by an electric vehicle wireless power transfer (EV-WPT) device under misaligned operating conditions and metallic coronary stents, which could be potentially hazardous to human electromagnetic safety. In this paper, we established that a standing human with a coronary stent implant and a sitting human with a coronary stent implant are exposed to the leaked EMFs of an EV-WPT device with a transmission power of 11 kW and 22 kW and a transmission frequency of 85 kHz, and we quantified the induced <i>E</i>-field strength of the human body by considering the <i>x</i>- and <i>z</i>-axis misalignment ranges of the WPT device as [−75 mm, 75 mm]. The results showed that the enhancement of the induced <i>E</i>-field in the human tissues near the sharp edges of the implant is significant. Meanwhile, the larger the WPT device misalignment generated, the higher the transmission power, and the closer the human body is to the WPT device, the larger the induced <i>E</i>-field<sub>max</sub> value of the human body is, which has the possibility of exceeding the ICNIRP safety limit, and there is a potential human body electromagnetic safety problem.https://www.mdpi.com/2079-9292/12/20/4231electric vehicle (EV)wireless power transfer (WPT)precision medical implantsinduced electric field strengthelectromagnetic field exposure
spellingShingle Tianhao Wang
Bo Li
Kaifeng Zhao
Quanyi Yu
Linlin Xu
Yaodan Chi
Shanshan Guan
Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
Electronics
electric vehicle (EV)
wireless power transfer (WPT)
precision medical implants
induced electric field strength
electromagnetic field exposure
title Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
title_full Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
title_fullStr Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
title_full_unstemmed Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
title_short Evaluation of Electromagnetic Exposure of the Human with a Coronary Stent Implant from an Electric Vehicle Wireless Power Transfer Device
title_sort evaluation of electromagnetic exposure of the human with a coronary stent implant from an electric vehicle wireless power transfer device
topic electric vehicle (EV)
wireless power transfer (WPT)
precision medical implants
induced electric field strength
electromagnetic field exposure
url https://www.mdpi.com/2079-9292/12/20/4231
work_keys_str_mv AT tianhaowang evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT boli evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT kaifengzhao evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT quanyiyu evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT linlinxu evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT yaodanchi evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice
AT shanshanguan evaluationofelectromagneticexposureofthehumanwithacoronarystentimplantfromanelectricvehiclewirelesspowertransferdevice