Comparison of Diagon Coag XL and Cobas t 511 analyzers
INTRODUCTION: We compared prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, and Ddimer test results measured using the Diagon Coag XL coagulation analyzer with Cobas t 511 analyzer. METHODS: Imprecision studies were performed for the PT, aPTT, fibrinogen, and D-dimer t...