Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature

Abstract Objective To review current literature and experience with Gore‐TexⓇ implant extrusion following medialization thyroplasty, as well as to report the unique case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Methods Review of existing literature and description...

Full description

Bibliographic Details
Main Authors: Ariel Frost, Natasha Mirza
Format: Article
Language:English
Published: Wiley 2022-09-01
Series:World Journal of Otorhinolaryngology-Head and Neck Surgery
Subjects:
Online Access:https://doi.org/10.1016/j.wjorl.2020.08.001
_version_ 1811185585232019456
author Ariel Frost
Natasha Mirza
author_facet Ariel Frost
Natasha Mirza
author_sort Ariel Frost
collection DOAJ
description Abstract Objective To review current literature and experience with Gore‐TexⓇ implant extrusion following medialization thyroplasty, as well as to report the unique case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Methods Review of existing literature and description of personal experience with unique case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Results Review of existing literature found no prior reported cases of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Risk factors for implant extrusion include the pressure of the implant on insertion and the inability to secure the implant. Cases of implant extrusion can be managed operatively via an endoscopic or via an external open approach. Conclusion This is the first reported case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Careful consideration should be given in revision medialization thyroplasty as additional implant material may cause increased pressure, a risk factor for implant extrusion.
first_indexed 2024-04-11T13:32:55Z
format Article
id doaj.art-197054818f384344b2f2d0a4a29488d7
institution Directory Open Access Journal
issn 2095-8811
2589-1081
language English
last_indexed 2024-04-11T13:32:55Z
publishDate 2022-09-01
publisher Wiley
record_format Article
series World Journal of Otorhinolaryngology-Head and Neck Surgery
spelling doaj.art-197054818f384344b2f2d0a4a29488d72022-12-22T04:21:44ZengWileyWorld Journal of Otorhinolaryngology-Head and Neck Surgery2095-88112589-10812022-09-018321321610.1016/j.wjorl.2020.08.001Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literatureAriel Frost0Natasha Mirza1Department of Otorhinolaryngology ‐ Head & Neck Surgery University of Pennsylvania Philadelphia PA USADepartment of Otorhinolaryngology ‐ Head & Neck Surgery University of Pennsylvania Philadelphia PA USAAbstract Objective To review current literature and experience with Gore‐TexⓇ implant extrusion following medialization thyroplasty, as well as to report the unique case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Methods Review of existing literature and description of personal experience with unique case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Results Review of existing literature found no prior reported cases of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Risk factors for implant extrusion include the pressure of the implant on insertion and the inability to secure the implant. Cases of implant extrusion can be managed operatively via an endoscopic or via an external open approach. Conclusion This is the first reported case of Gore‐TexⓇ implant extrusion following revision medialization thyroplasty. Careful consideration should be given in revision medialization thyroplasty as additional implant material may cause increased pressure, a risk factor for implant extrusion.https://doi.org/10.1016/j.wjorl.2020.08.001ComplicationsGore‐TexⓇImplant extrusionLaryngoplastyMedialization thyroplastyRevision
spellingShingle Ariel Frost
Natasha Mirza
Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
World Journal of Otorhinolaryngology-Head and Neck Surgery
Complications
Gore‐TexⓇ
Implant extrusion
Laryngoplasty
Medialization thyroplasty
Revision
title Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
title_full Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
title_fullStr Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
title_full_unstemmed Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
title_short Gore‐Tex implant extrusion following revision medialization thyroplasty: Case report and review of the literature
title_sort gore tex implant extrusion following revision medialization thyroplasty case report and review of the literature
topic Complications
Gore‐TexⓇ
Implant extrusion
Laryngoplasty
Medialization thyroplasty
Revision
url https://doi.org/10.1016/j.wjorl.2020.08.001
work_keys_str_mv AT arielfrost goreteximplantextrusionfollowingrevisionmedializationthyroplastycasereportandreviewoftheliterature
AT natashamirza goreteximplantextrusionfollowingrevisionmedializationthyroplastycasereportandreviewoftheliterature