On the inaccuracies of dental radiometers.

This study investigated the accuracy of sixteen models of commercial dental radiometers (DR) in measuring the output of thirty-eight LED light curing units (LCUs) compared with a 'gold standard' laboratory-grade spectrometer integrating-sphere (IS) assembly. Nineteen Type I (fiber-bundle l...

Full description

Bibliographic Details
Main Authors: Adrian C Shortall, Mohammed A Hadis, William M Palin
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2021-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0245830
_version_ 1818919845958254592
author Adrian C Shortall
Mohammed A Hadis
William M Palin
author_facet Adrian C Shortall
Mohammed A Hadis
William M Palin
author_sort Adrian C Shortall
collection DOAJ
description This study investigated the accuracy of sixteen models of commercial dental radiometers (DR) in measuring the output of thirty-eight LED light curing units (LCUs) compared with a 'gold standard' laboratory-grade spectrometer integrating-sphere (IS) assembly. Nineteen Type I (fiber-bundle light guide) and nineteen Type II (light source in head) LED LCUs were tested, some using different output modes and light guides, resulting in 61 test subsets per radiometer. Gold standard (GS) output measurements (n = 3) were taken using the IS and confirmed with two types of laboratory-grade power meter (PowerMax-Pro 150 HD and PM10-19C; Coherent). One DR (Bluephase Meter II, Ivoclar; BM II) allowed power (mW) as well as irradiance (mW/cm2) recordings. Irradiance readings (n = 3) for each DR/LCU were compared with the IS derived irradiance. Individual LCU irradiance values were normalized against IS data. The GS method yielded reproducible data with a 0.4% pooled coefficient of variation for the LCUs. Mean power values ranged from 0.19 W to 2.40 W. Overall power values for the laboratory-grade power meters were within 5% of GS values. Individual LCU/DR normalized irradiance values ranged from 7% to 535% of the GS; an order of magnitude greater than previous reports. BM II was the only radiometer to average within 20% of normalized pooled GS irradiance values, whereas other radiometers differed by up to 85%. Ten radiometers failed to provide any reading for 1 LCU. When tested with the PowerMax-Pro in high speed (20 kHz) mode, eight LCUs demonstrated pulsing outputs undetectable at the standard (10 Hz) data acquisition rate. Sufficient light exposure is critical for the successful curing of dental resin-based materials. Substantial discrepancies may occur between actual and estimated radiometric data using current DRs. More accurate DRs need to be developed. Manufacturers' accuracy claims for DRs should specify compatible LCUs and testing parameters.
first_indexed 2024-12-20T01:12:20Z
format Article
id doaj.art-19baea9075dc4e89b3c3a44043353422
institution Directory Open Access Journal
issn 1932-6203
language English
last_indexed 2024-12-20T01:12:20Z
publishDate 2021-01-01
publisher Public Library of Science (PLoS)
record_format Article
series PLoS ONE
spelling doaj.art-19baea9075dc4e89b3c3a440433534222022-12-21T19:58:39ZengPublic Library of Science (PLoS)PLoS ONE1932-62032021-01-01161e024583010.1371/journal.pone.0245830On the inaccuracies of dental radiometers.Adrian C ShortallMohammed A HadisWilliam M PalinThis study investigated the accuracy of sixteen models of commercial dental radiometers (DR) in measuring the output of thirty-eight LED light curing units (LCUs) compared with a 'gold standard' laboratory-grade spectrometer integrating-sphere (IS) assembly. Nineteen Type I (fiber-bundle light guide) and nineteen Type II (light source in head) LED LCUs were tested, some using different output modes and light guides, resulting in 61 test subsets per radiometer. Gold standard (GS) output measurements (n = 3) were taken using the IS and confirmed with two types of laboratory-grade power meter (PowerMax-Pro 150 HD and PM10-19C; Coherent). One DR (Bluephase Meter II, Ivoclar; BM II) allowed power (mW) as well as irradiance (mW/cm2) recordings. Irradiance readings (n = 3) for each DR/LCU were compared with the IS derived irradiance. Individual LCU irradiance values were normalized against IS data. The GS method yielded reproducible data with a 0.4% pooled coefficient of variation for the LCUs. Mean power values ranged from 0.19 W to 2.40 W. Overall power values for the laboratory-grade power meters were within 5% of GS values. Individual LCU/DR normalized irradiance values ranged from 7% to 535% of the GS; an order of magnitude greater than previous reports. BM II was the only radiometer to average within 20% of normalized pooled GS irradiance values, whereas other radiometers differed by up to 85%. Ten radiometers failed to provide any reading for 1 LCU. When tested with the PowerMax-Pro in high speed (20 kHz) mode, eight LCUs demonstrated pulsing outputs undetectable at the standard (10 Hz) data acquisition rate. Sufficient light exposure is critical for the successful curing of dental resin-based materials. Substantial discrepancies may occur between actual and estimated radiometric data using current DRs. More accurate DRs need to be developed. Manufacturers' accuracy claims for DRs should specify compatible LCUs and testing parameters.https://doi.org/10.1371/journal.pone.0245830
spellingShingle Adrian C Shortall
Mohammed A Hadis
William M Palin
On the inaccuracies of dental radiometers.
PLoS ONE
title On the inaccuracies of dental radiometers.
title_full On the inaccuracies of dental radiometers.
title_fullStr On the inaccuracies of dental radiometers.
title_full_unstemmed On the inaccuracies of dental radiometers.
title_short On the inaccuracies of dental radiometers.
title_sort on the inaccuracies of dental radiometers
url https://doi.org/10.1371/journal.pone.0245830
work_keys_str_mv AT adriancshortall ontheinaccuraciesofdentalradiometers
AT mohammedahadis ontheinaccuraciesofdentalradiometers
AT williammpalin ontheinaccuraciesofdentalradiometers