Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges
Ιn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to righ...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2021-05-01
|
Series: | Indian Pacing and Electrophysiology Journal |
Subjects: | |
Online Access: | http://www.sciencedirect.com/science/article/pii/S0972629221000292 |
_version_ | 1818651878917931008 |
---|---|
author | Fani Zagkli Panagiotis Chronopoulos John Chiladakis |
author_facet | Fani Zagkli Panagiotis Chronopoulos John Chiladakis |
author_sort | Fani Zagkli |
collection | DOAJ |
description | Ιn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to right ventricular pace-sense lead fracture. The need to ensure adequate ICD antibradycardia backup pacing in pacing-dependent patients when deleterious sensing errors occur, cannot be overemphasized. |
first_indexed | 2024-12-17T02:13:07Z |
format | Article |
id | doaj.art-1b3b9afebd3f40c4b73af08ac0c283e4 |
institution | Directory Open Access Journal |
issn | 0972-6292 |
language | English |
last_indexed | 2024-12-17T02:13:07Z |
publishDate | 2021-05-01 |
publisher | Elsevier |
record_format | Article |
series | Indian Pacing and Electrophysiology Journal |
spelling | doaj.art-1b3b9afebd3f40c4b73af08ac0c283e42022-12-21T22:07:29ZengElsevierIndian Pacing and Electrophysiology Journal0972-62922021-05-01213182185Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and dischargesFani Zagkli0Panagiotis Chronopoulos1John Chiladakis2University Hospital of Patras, Department of Cardiology, GreeceUniversity Hospital of Patras, Department of Cardiology, GreeceCorresponding author.; University Hospital of Patras, Department of Cardiology, GreeceΙn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to right ventricular pace-sense lead fracture. The need to ensure adequate ICD antibradycardia backup pacing in pacing-dependent patients when deleterious sensing errors occur, cannot be overemphasized.http://www.sciencedirect.com/science/article/pii/S0972629221000292BradycardiaICD lead FractureOversensingInappropriate shocks |
spellingShingle | Fani Zagkli Panagiotis Chronopoulos John Chiladakis Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges Indian Pacing and Electrophysiology Journal Bradycardia ICD lead Fracture Oversensing Inappropriate shocks |
title | Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges |
title_full | Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges |
title_fullStr | Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges |
title_full_unstemmed | Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges |
title_short | Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges |
title_sort | unexpected inhibition of bradycardia pacing due to oversensing in icd lead fracture associated with spurious tachyarrhythmia detection and discharges |
topic | Bradycardia ICD lead Fracture Oversensing Inappropriate shocks |
url | http://www.sciencedirect.com/science/article/pii/S0972629221000292 |
work_keys_str_mv | AT fanizagkli unexpectedinhibitionofbradycardiapacingduetooversensinginicdleadfractureassociatedwithspurioustachyarrhythmiadetectionanddischarges AT panagiotischronopoulos unexpectedinhibitionofbradycardiapacingduetooversensinginicdleadfractureassociatedwithspurioustachyarrhythmiadetectionanddischarges AT johnchiladakis unexpectedinhibitionofbradycardiapacingduetooversensinginicdleadfractureassociatedwithspurioustachyarrhythmiadetectionanddischarges |