Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform
Abstract Mycobacterium chimaera, belonging to the Mycobacterium avium complex, is an opportunistic environmental mycobacterium which has been isolated from medical device water samples such as Heater Cooler Units (HCU). Laboratories currently use culture-based diagnostic methods to detect M. chimaer...
Main Authors: | , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Nature Portfolio
2024-12-01
|
Series: | Scientific Reports |
Subjects: | |
Online Access: | https://doi.org/10.1038/s41598-024-82440-5 |
_version_ | 1826913697784659968 |
---|---|
author | Francesco Bisognin Michele Borghi Matteo Faccio Vincenzo Ferraro Federica Sorella Cristina Maria Crovara Pesce Alessandra Primavera Paola Dal Monte |
author_facet | Francesco Bisognin Michele Borghi Matteo Faccio Vincenzo Ferraro Federica Sorella Cristina Maria Crovara Pesce Alessandra Primavera Paola Dal Monte |
author_sort | Francesco Bisognin |
collection | DOAJ |
description | Abstract Mycobacterium chimaera, belonging to the Mycobacterium avium complex, is an opportunistic environmental mycobacterium which has been isolated from medical device water samples such as Heater Cooler Units (HCU). Laboratories currently use culture-based diagnostic methods to detect M. chimaera, but these take a long time to obtain results. The aim of this study is to test and define the Limit of Detection (LoD) of a Real-Time Polymerase Chain Reaction test (RT-PCR) specific for M. chimaera, using a sample-to-result InGenius platform, performed on medical device water samples collected at the referral centre for the detection of mycobacteria from environmental specimens of Bologna, Italy. A total of 285 water samples were included in this study. The sensitivity and specificity of RT-PCR compared to culture were 60.5% and 98.8% respectively, with an overall agreement of 82.1% and a positive predictive value of 97.4%. The LoD calculated was approximately 2900 CFU/ml. In conclusion, this study confirmed that detection of M. chimaera with RT-PCR could support culture-based methods in reducing the time necessary to identify highly colonised HCUs, with high positive predictive values. Therefore, we suggest performing this customised RT-PCR on concentrated decontaminated water samples, shutting down and thoroughly disinfecting positive HCUs, to reduce the risk of patient infection. |
first_indexed | 2025-02-17T10:48:47Z |
format | Article |
id | doaj.art-1d3910c6a00744f0bf9588ce6b51a397 |
institution | Directory Open Access Journal |
issn | 2045-2322 |
language | English |
last_indexed | 2025-02-17T10:48:47Z |
publishDate | 2024-12-01 |
publisher | Nature Portfolio |
record_format | Article |
series | Scientific Reports |
spelling | doaj.art-1d3910c6a00744f0bf9588ce6b51a3972024-12-29T12:23:37ZengNature PortfolioScientific Reports2045-23222024-12-011411610.1038/s41598-024-82440-5Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platformFrancesco Bisognin0Michele Borghi1Matteo Faccio2Vincenzo Ferraro3Federica Sorella4Cristina Maria Crovara Pesce5Alessandra Primavera6Paola Dal Monte7Microbiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaMicrobiology Unit, IRCCS Azienda Ospedaliero-Universitaria of BolognaAbstract Mycobacterium chimaera, belonging to the Mycobacterium avium complex, is an opportunistic environmental mycobacterium which has been isolated from medical device water samples such as Heater Cooler Units (HCU). Laboratories currently use culture-based diagnostic methods to detect M. chimaera, but these take a long time to obtain results. The aim of this study is to test and define the Limit of Detection (LoD) of a Real-Time Polymerase Chain Reaction test (RT-PCR) specific for M. chimaera, using a sample-to-result InGenius platform, performed on medical device water samples collected at the referral centre for the detection of mycobacteria from environmental specimens of Bologna, Italy. A total of 285 water samples were included in this study. The sensitivity and specificity of RT-PCR compared to culture were 60.5% and 98.8% respectively, with an overall agreement of 82.1% and a positive predictive value of 97.4%. The LoD calculated was approximately 2900 CFU/ml. In conclusion, this study confirmed that detection of M. chimaera with RT-PCR could support culture-based methods in reducing the time necessary to identify highly colonised HCUs, with high positive predictive values. Therefore, we suggest performing this customised RT-PCR on concentrated decontaminated water samples, shutting down and thoroughly disinfecting positive HCUs, to reduce the risk of patient infection.https://doi.org/10.1038/s41598-024-82440-5Nontuberculous mycobacteriaM. chimaeraMedical device water samplesHeater-cooler unitsReal-time PCRELITe InGenius® platform |
spellingShingle | Francesco Bisognin Michele Borghi Matteo Faccio Vincenzo Ferraro Federica Sorella Cristina Maria Crovara Pesce Alessandra Primavera Paola Dal Monte Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform Scientific Reports Nontuberculous mycobacteria M. chimaera Medical device water samples Heater-cooler units Real-time PCR ELITe InGenius® platform |
title | Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform |
title_full | Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform |
title_fullStr | Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform |
title_full_unstemmed | Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform |
title_short | Detection of Mycobacterium chimaera in medical device water samples by customised real time PCR using a InGenius platform |
title_sort | detection of mycobacterium chimaera in medical device water samples by customised real time pcr using a ingenius platform |
topic | Nontuberculous mycobacteria M. chimaera Medical device water samples Heater-cooler units Real-time PCR ELITe InGenius® platform |
url | https://doi.org/10.1038/s41598-024-82440-5 |
work_keys_str_mv | AT francescobisognin detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT micheleborghi detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT matteofaccio detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT vincenzoferraro detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT federicasorella detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT cristinamariacrovarapesce detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT alessandraprimavera detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform AT paoladalmonte detectionofmycobacteriumchimaerainmedicaldevicewatersamplesbycustomisedrealtimepcrusingaingeniusplatform |