Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway
We attempted to develop different sized catheter flow sensors for evaluating breathing characteristics at different levels of the tracheobronchial airway in a lung system. Two catheter flow sensors with 1.8 and 3.5 mm outer diameters were designed for measuring breathing characteristics in bronchus-...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2017-08-01
|
Series: | Proceedings |
Subjects: | |
Online Access: | https://www.mdpi.com/2504-3900/1/4/356 |
_version_ | 1827092796247375872 |
---|---|
author | Tomohiro Fujinori Yoshihiro Hasegawa Miyoko Matsushima Tsutomo Kawabe Mitsuhiro Shikida |
author_facet | Tomohiro Fujinori Yoshihiro Hasegawa Miyoko Matsushima Tsutomo Kawabe Mitsuhiro Shikida |
author_sort | Tomohiro Fujinori |
collection | DOAJ |
description | We attempted to develop different sized catheter flow sensors for evaluating breathing characteristics at different levels of the tracheobronchial airway in a lung system. Two catheter flow sensors with 1.8 and 3.5 mm outer diameters were designed for measuring breathing characteristics in bronchus- and bronchiole- regions in the lung system, respectively. They were fabricated by using photolithography and heat shrinkable tube packaging processes. We experimentally investigated the fundamental sensor characteristic and found that the sensor output depended on the mean flow velocity. The developed catheter flow sensors also successfully detected the oscillating airflow produced by an artificial ventilator. |
first_indexed | 2024-04-13T03:06:11Z |
format | Article |
id | doaj.art-1db1c94e41c34d34aaa36da5e9948a44 |
institution | Directory Open Access Journal |
issn | 2504-3900 |
language | English |
last_indexed | 2025-03-20T06:17:53Z |
publishDate | 2017-08-01 |
publisher | MDPI AG |
record_format | Article |
series | Proceedings |
spelling | doaj.art-1db1c94e41c34d34aaa36da5e9948a442024-10-02T16:26:31ZengMDPI AGProceedings2504-39002017-08-011435610.3390/proceedings1040356proceedings1040356Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial AirwayTomohiro Fujinori0Yoshihiro Hasegawa1Miyoko Matsushima2Tsutomo Kawabe3Mitsuhiro Shikida4Hiroshima City University, 3-4-1, Ozuka-Higashi, Asa-Minami-Ku, Hiroshima 731-3194, JapanHiroshima City University, 3-4-1, Ozuka-Higashi, Asa-Minami-Ku, Hiroshima 731-3194, JapanNagoya University, 1-1-20, Daiko-Minami, Higashi-ku, Nagoya 464-8673, JapanNagoya University, 1-1-20, Daiko-Minami, Higashi-ku, Nagoya 464-8673, JapanHiroshima City University, 3-4-1, Ozuka-Higashi, Asa-Minami-Ku, Hiroshima 731-3194, JapanWe attempted to develop different sized catheter flow sensors for evaluating breathing characteristics at different levels of the tracheobronchial airway in a lung system. Two catheter flow sensors with 1.8 and 3.5 mm outer diameters were designed for measuring breathing characteristics in bronchus- and bronchiole- regions in the lung system, respectively. They were fabricated by using photolithography and heat shrinkable tube packaging processes. We experimentally investigated the fundamental sensor characteristic and found that the sensor output depended on the mean flow velocity. The developed catheter flow sensors also successfully detected the oscillating airflow produced by an artificial ventilator.https://www.mdpi.com/2504-3900/1/4/356COPDcatheter flow sensorbreathing measurement |
spellingShingle | Tomohiro Fujinori Yoshihiro Hasegawa Miyoko Matsushima Tsutomo Kawabe Mitsuhiro Shikida Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway Proceedings COPD catheter flow sensor breathing measurement |
title | Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway |
title_full | Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway |
title_fullStr | Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway |
title_full_unstemmed | Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway |
title_short | Development of Catheter Flow Sensor for Breathing Measurements at Different Levels of Tracheobronchial Airway |
title_sort | development of catheter flow sensor for breathing measurements at different levels of tracheobronchial airway |
topic | COPD catheter flow sensor breathing measurement |
url | https://www.mdpi.com/2504-3900/1/4/356 |
work_keys_str_mv | AT tomohirofujinori developmentofcatheterflowsensorforbreathingmeasurementsatdifferentlevelsoftracheobronchialairway AT yoshihirohasegawa developmentofcatheterflowsensorforbreathingmeasurementsatdifferentlevelsoftracheobronchialairway AT miyokomatsushima developmentofcatheterflowsensorforbreathingmeasurementsatdifferentlevelsoftracheobronchialairway AT tsutomokawabe developmentofcatheterflowsensorforbreathingmeasurementsatdifferentlevelsoftracheobronchialairway AT mitsuhiroshikida developmentofcatheterflowsensorforbreathingmeasurementsatdifferentlevelsoftracheobronchialairway |