Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction

Defibrillation threshold (DFT) testing is usually recommended after device implantation to confirm appropriate implantable cardioverter defibrillator (ICD)/cardiac resynchronization therapy defibrillator (CRT-D) function [1,2]. However, induction of ventricular fibrillation may result in hemodynamic...

Full description

Bibliographic Details
Main Authors: Toshiko Nakai, MD, Kimie Ohkubo, MD, Yasuo Okumura, MD, Satoshi Kunimoto, MD, Yuji Kasamaki, MD, Ichiro Watanabe, MD, Atsushi Hirayama, MD
Format: Article
Language:English
Published: Wiley 2012-04-01
Series:Journal of Arrhythmia
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S188042761200035X
_version_ 1818944814204321792
author Toshiko Nakai, MD
Kimie Ohkubo, MD
Yasuo Okumura, MD
Satoshi Kunimoto, MD
Yuji Kasamaki, MD
Ichiro Watanabe, MD
Atsushi Hirayama, MD
author_facet Toshiko Nakai, MD
Kimie Ohkubo, MD
Yasuo Okumura, MD
Satoshi Kunimoto, MD
Yuji Kasamaki, MD
Ichiro Watanabe, MD
Atsushi Hirayama, MD
author_sort Toshiko Nakai, MD
collection DOAJ
description Defibrillation threshold (DFT) testing is usually recommended after device implantation to confirm appropriate implantable cardioverter defibrillator (ICD)/cardiac resynchronization therapy defibrillator (CRT-D) function [1,2]. However, induction of ventricular fibrillation may result in hemodynamic compromise, and cardioversion itself may cause myocardial injury [3,4]. We report on a CRT-D patient with acute myocardial infarction who died due to multiple organ failure 1 day after DFT testing. Our case emphasizes the importance of deciding whether DFT testing should be performed for patients with very severe heart failure in the acute stage of myocardial infarction.
first_indexed 2024-12-20T07:49:12Z
format Article
id doaj.art-21c55db1fecd4dc7ad9ce202d31117a8
institution Directory Open Access Journal
issn 1880-4276
language English
last_indexed 2024-12-20T07:49:12Z
publishDate 2012-04-01
publisher Wiley
record_format Article
series Journal of Arrhythmia
spelling doaj.art-21c55db1fecd4dc7ad9ce202d31117a82022-12-21T19:47:52ZengWileyJournal of Arrhythmia1880-42762012-04-0128211411610.1016/j.joa.2012.03.009Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarctionToshiko Nakai, MDKimie Ohkubo, MDYasuo Okumura, MDSatoshi Kunimoto, MDYuji Kasamaki, MDIchiro Watanabe, MDAtsushi Hirayama, MDDefibrillation threshold (DFT) testing is usually recommended after device implantation to confirm appropriate implantable cardioverter defibrillator (ICD)/cardiac resynchronization therapy defibrillator (CRT-D) function [1,2]. However, induction of ventricular fibrillation may result in hemodynamic compromise, and cardioversion itself may cause myocardial injury [3,4]. We report on a CRT-D patient with acute myocardial infarction who died due to multiple organ failure 1 day after DFT testing. Our case emphasizes the importance of deciding whether DFT testing should be performed for patients with very severe heart failure in the acute stage of myocardial infarction.http://www.sciencedirect.com/science/article/pii/S188042761200035XCRT-DDefibrillation threshold testing
spellingShingle Toshiko Nakai, MD
Kimie Ohkubo, MD
Yasuo Okumura, MD
Satoshi Kunimoto, MD
Yuji Kasamaki, MD
Ichiro Watanabe, MD
Atsushi Hirayama, MD
Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
Journal of Arrhythmia
CRT-D
Defibrillation threshold testing
title Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
title_full Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
title_fullStr Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
title_full_unstemmed Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
title_short Risk of defibrillation threshold testing in severe heart failure patient: A case of cardiac resynchronization therapy (CRT-D) with acute myocardial infarction
title_sort risk of defibrillation threshold testing in severe heart failure patient a case of cardiac resynchronization therapy crt d with acute myocardial infarction
topic CRT-D
Defibrillation threshold testing
url http://www.sciencedirect.com/science/article/pii/S188042761200035X
work_keys_str_mv AT toshikonakaimd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT kimieohkubomd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT yasuookumuramd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT satoshikunimotomd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT yujikasamakimd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT ichirowatanabemd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction
AT atsushihirayamamd riskofdefibrillationthresholdtestinginsevereheartfailurepatientacaseofcardiacresynchronizationtherapycrtdwithacutemyocardialinfarction