A case of two connected stents deployed during iStent inject W surgery

Abstract Background We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. Case presentation A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent...

Full description

Bibliographic Details
Main Authors: Ayaka Shimada, Sho Ichioka, Akiko Ishida, Sachiko Kaidzu, Masaki Tanito
Format: Article
Language:English
Published: BMC 2023-05-01
Series:BMC Ophthalmology
Subjects:
Online Access:https://doi.org/10.1186/s12886-023-02951-z
_version_ 1797827626295885824
author Ayaka Shimada
Sho Ichioka
Akiko Ishida
Sachiko Kaidzu
Masaki Tanito
author_facet Ayaka Shimada
Sho Ichioka
Akiko Ishida
Sachiko Kaidzu
Masaki Tanito
author_sort Ayaka Shimada
collection DOAJ
description Abstract Background We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. Case presentation A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. Conclusions Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery.
first_indexed 2024-04-09T12:51:21Z
format Article
id doaj.art-23d57fec99804f64898c8aaa2d479312
institution Directory Open Access Journal
issn 1471-2415
language English
last_indexed 2024-04-09T12:51:21Z
publishDate 2023-05-01
publisher BMC
record_format Article
series BMC Ophthalmology
spelling doaj.art-23d57fec99804f64898c8aaa2d4793122023-05-14T11:11:52ZengBMCBMC Ophthalmology1471-24152023-05-012311410.1186/s12886-023-02951-zA case of two connected stents deployed during iStent inject W surgeryAyaka Shimada0Sho Ichioka1Akiko Ishida2Sachiko Kaidzu3Masaki Tanito4Department of Ophthalmology, Shimane University Faculty of MedicineDepartment of Ophthalmology, Shimane University Faculty of MedicineDepartment of Ophthalmology, Shimane University Faculty of MedicineDepartment of Ophthalmology, Shimane University Faculty of MedicineDepartment of Ophthalmology, Shimane University Faculty of MedicineAbstract Background We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. Case presentation A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. Conclusions Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery.https://doi.org/10.1186/s12886-023-02951-zMinimally invasive glaucoma surgeryiStentSurgical complication
spellingShingle Ayaka Shimada
Sho Ichioka
Akiko Ishida
Sachiko Kaidzu
Masaki Tanito
A case of two connected stents deployed during iStent inject W surgery
BMC Ophthalmology
Minimally invasive glaucoma surgery
iStent
Surgical complication
title A case of two connected stents deployed during iStent inject W surgery
title_full A case of two connected stents deployed during iStent inject W surgery
title_fullStr A case of two connected stents deployed during iStent inject W surgery
title_full_unstemmed A case of two connected stents deployed during iStent inject W surgery
title_short A case of two connected stents deployed during iStent inject W surgery
title_sort case of two connected stents deployed during istent inject w surgery
topic Minimally invasive glaucoma surgery
iStent
Surgical complication
url https://doi.org/10.1186/s12886-023-02951-z
work_keys_str_mv AT ayakashimada acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT shoichioka acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT akikoishida acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT sachikokaidzu acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT masakitanito acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT ayakashimada caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT shoichioka caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT akikoishida caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT sachikokaidzu caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT masakitanito caseoftwoconnectedstentsdeployedduringistentinjectwsurgery