Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
Purpose: To employ the microDiamond and the microSilicon detector (mDD and mSD, both PTW-Freiburg, Germany) to determine the dose rate around a HDR 192Ir brachytherapy source (model mHDR-v2r, Elekta AB, Sweden).Methods: The detectors were calibrated with a 60Co beam at the PTW Calibration Laboratory...
Main Authors: | Giulio Rossi, Thomas Failing, Mark Gainey, Michael Kollefrath, Frank Hensley, Klemens Zink, Dimos Baltas |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2023-11-01
|
Series: | Zeitschrift für Medizinische Physik |
Subjects: | |
Online Access: | http://www.sciencedirect.com/science/article/pii/S0939388922000861 |