In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry
BACKGROUND Usually, laser ablation inductively coupled plasma-mass spectrometry (LA-ICP-MS) analysis is equipped with a closed ablation cell. It cannot be used for non-destructive analysis of the large-size samples which exceeds the size of the cell. OBJECTIVES To realize directly in situ non-destru...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Science Press, PR China
2021-01-01
|
Series: | Yankuang ceshi |
Subjects: | |
Online Access: | http://www.ykcs.ac.cn/en/article/doi/10.15898/j.cnki.11-2131/td.202005240075 |
_version_ | 1797953786911653888 |
---|---|
author | ZHOU Fan LI Ming CHAI Xin-na HU Zhao-chu LUO Tao HU Sheng-hong |
author_facet | ZHOU Fan LI Ming CHAI Xin-na HU Zhao-chu LUO Tao HU Sheng-hong |
author_sort | ZHOU Fan |
collection | DOAJ |
description | BACKGROUND Usually, laser ablation inductively coupled plasma-mass spectrometry (LA-ICP-MS) analysis is equipped with a closed ablation cell. It cannot be used for non-destructive analysis of the large-size samples which exceeds the size of the cell. OBJECTIVES To realize directly in situ non-destructive analysis of major and trace elements in the large-size samples. METHODS Based on the self-designed open sample collection port, combined with the gas exchange device, an open LA-ICP-MS analysis method was established. Taking the large-size ceramic disc sample as an example, samples can be directly determined for major and trace elements in an air environment without being broken. RESULTS Analyte aerosols produced by laser ablation were collected and sucked by open sample collector. The air in the aerosols was replaced via transfer tube by high purity argon in a gas exchange device, the analyte aerosols were then transferred into ICP-MS for detection. After non-destructive analyses by this method, large-size ceramic samples were broken and analyzed by traditional LA-ICP-MS. Relative deviations of most of 51 elements detected by the two methods were less than 10%. Only part of elements (such as P, Be, Sc, Y, La, Sm, Eu, Dy, Hf, W, etc.) had relative deviations higher than 20%, because of the extremely low contents. CONCLUSIONS The open LA-ICP-MS method is suitable for in situ non-destructive determination of major and trace elements in the large-size samples. |
first_indexed | 2024-04-10T23:08:39Z |
format | Article |
id | doaj.art-293325221ff8491982a04b08e73eb384 |
institution | Directory Open Access Journal |
issn | 0254-5357 |
language | English |
last_indexed | 2024-04-10T23:08:39Z |
publishDate | 2021-01-01 |
publisher | Science Press, PR China |
record_format | Article |
series | Yankuang ceshi |
spelling | doaj.art-293325221ff8491982a04b08e73eb3842023-01-13T08:58:55ZengScience Press, PR ChinaYankuang ceshi0254-53572021-01-01401334110.15898/j.cnki.11-2131/td.202005240075yk202005240075In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass SpectrometryZHOU Fan0LI Ming1CHAI Xin-na2HU Zhao-chu3LUO Tao4HU Sheng-hong5State Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences(Wuhan), Wuhan 430074, ChinaState Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences(Wuhan), Wuhan 430074, ChinaState Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences(Wuhan), Wuhan 430074, ChinaState Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences(Wuhan), Wuhan 430074, ChinaState Key Laboratory of Geological Processes and Mineral Resources, China University of Geosciences(Wuhan), Wuhan 430074, ChinaSchool of Earth Sciences, China University of Geosciences(Wuhan), Wuhan 430074, ChinaBACKGROUND Usually, laser ablation inductively coupled plasma-mass spectrometry (LA-ICP-MS) analysis is equipped with a closed ablation cell. It cannot be used for non-destructive analysis of the large-size samples which exceeds the size of the cell. OBJECTIVES To realize directly in situ non-destructive analysis of major and trace elements in the large-size samples. METHODS Based on the self-designed open sample collection port, combined with the gas exchange device, an open LA-ICP-MS analysis method was established. Taking the large-size ceramic disc sample as an example, samples can be directly determined for major and trace elements in an air environment without being broken. RESULTS Analyte aerosols produced by laser ablation were collected and sucked by open sample collector. The air in the aerosols was replaced via transfer tube by high purity argon in a gas exchange device, the analyte aerosols were then transferred into ICP-MS for detection. After non-destructive analyses by this method, large-size ceramic samples were broken and analyzed by traditional LA-ICP-MS. Relative deviations of most of 51 elements detected by the two methods were less than 10%. Only part of elements (such as P, Be, Sc, Y, La, Sm, Eu, Dy, Hf, W, etc.) had relative deviations higher than 20%, because of the extremely low contents. CONCLUSIONS The open LA-ICP-MS method is suitable for in situ non-destructive determination of major and trace elements in the large-size samples.http://www.ykcs.ac.cn/en/article/doi/10.15898/j.cnki.11-2131/td.202005240075open laser ablationlarge-size ceramic samplesnon-destructive analysisin situ |
spellingShingle | ZHOU Fan LI Ming CHAI Xin-na HU Zhao-chu LUO Tao HU Sheng-hong In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry Yankuang ceshi open laser ablation large-size ceramic samples non-destructive analysis in situ |
title | In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry |
title_full | In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry |
title_fullStr | In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry |
title_full_unstemmed | In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry |
title_short | In-situ Non-destructive Determination of Major and Trace Elements in Large Size Ceramic Samples by Open Laser Ablation Inductively Coupled Plasma-Mass Spectrometry |
title_sort | in situ non destructive determination of major and trace elements in large size ceramic samples by open laser ablation inductively coupled plasma mass spectrometry |
topic | open laser ablation large-size ceramic samples non-destructive analysis in situ |
url | http://www.ykcs.ac.cn/en/article/doi/10.15898/j.cnki.11-2131/td.202005240075 |
work_keys_str_mv | AT zhoufan insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry AT liming insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry AT chaixinna insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry AT huzhaochu insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry AT luotao insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry AT hushenghong insitunondestructivedeterminationofmajorandtraceelementsinlargesizeceramicsamplesbyopenlaserablationinductivelycoupledplasmamassspectrometry |