Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India

A Helical Tomotherapy<sup> TM</sup> (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two o...

Full description

Bibliographic Details
Main Authors: Kinhikar Rajesh, Master Zubin, Dhote Dipak, Deshpande Deepak
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2009-01-01
Series:Journal of Medical Physics
Subjects:
Online Access:http://www.jmp.org.in/article.asp?issn=0971-6203;year=2009;volume=34;issue=2;spage=73;epage=79;aulast=Kinhikar
_version_ 1818346327543644160
author Kinhikar Rajesh
Master Zubin
Dhote Dipak
Deshpande Deepak
author_facet Kinhikar Rajesh
Master Zubin
Dhote Dipak
Deshpande Deepak
author_sort Kinhikar Rajesh
collection DOAJ
description A Helical Tomotherapy<sup> TM</sup> (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study are: (i) To estimate the dosimetric and general capability (TomoImage registration, reconstruction, contrast and spatial resolution, artifacts-free image and dose in TomoImage) of on-board MVCT detectors. (ii) To measure the dosimetric parameters (output and energy) following major repair. The MVCT detectors also estimated the rotational output constancy well. During this study, dosimetric tests were repeated after replacing MVCT detectors and the target. fixed-gantry/fixed-couch measurements were measured daily to investigate; the system stability. Thermoluminescense dosimeter (TLD) was used during both the measurements subsequently. The MVCT image quality with old and new detectors was comparable and hence acceptable clinically. The spatial resolution was optimal and the dose during TomoImage was 2 cGy (well within the manufacturer tolerance of 4 cGy). The results of lateral beam profiles showed an excellent agreement between the two normalized plots. The output from the rotational procedure revealed 99.7&#x0025; while the energy was consistent over a period of twelve months. The Hi-Art II system has maintained its calibration to within &#x002B;/- 2&#x0025; and energy to within &#x002B;/- 1.5&#x0025; over the initial twelve-month period. Based on the periodic measurements for rotational output and consistency in the lateral beam profile shape, the on-board detector proved to be a viable dosimetric quality assurance tool for IMRT with Tomotherapy. Tomotherapy was stable from the dosimetric point of view during the twelve-month period.
first_indexed 2024-12-13T17:16:30Z
format Article
id doaj.art-29b360800e0d4a1586de8f86f2e86cfe
institution Directory Open Access Journal
issn 0971-6203
1998-3913
language English
last_indexed 2024-12-13T17:16:30Z
publishDate 2009-01-01
publisher Wolters Kluwer Medknow Publications
record_format Article
series Journal of Medical Physics
spelling doaj.art-29b360800e0d4a1586de8f86f2e86cfe2022-12-21T23:37:25ZengWolters Kluwer Medknow PublicationsJournal of Medical Physics0971-62031998-39132009-01-013427379Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in IndiaKinhikar RajeshMaster ZubinDhote DipakDeshpande DeepakA Helical Tomotherapy<sup> TM</sup> (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study are: (i) To estimate the dosimetric and general capability (TomoImage registration, reconstruction, contrast and spatial resolution, artifacts-free image and dose in TomoImage) of on-board MVCT detectors. (ii) To measure the dosimetric parameters (output and energy) following major repair. The MVCT detectors also estimated the rotational output constancy well. During this study, dosimetric tests were repeated after replacing MVCT detectors and the target. fixed-gantry/fixed-couch measurements were measured daily to investigate; the system stability. Thermoluminescense dosimeter (TLD) was used during both the measurements subsequently. The MVCT image quality with old and new detectors was comparable and hence acceptable clinically. The spatial resolution was optimal and the dose during TomoImage was 2 cGy (well within the manufacturer tolerance of 4 cGy). The results of lateral beam profiles showed an excellent agreement between the two normalized plots. The output from the rotational procedure revealed 99.7&#x0025; while the energy was consistent over a period of twelve months. The Hi-Art II system has maintained its calibration to within &#x002B;/- 2&#x0025; and energy to within &#x002B;/- 1.5&#x0025; over the initial twelve-month period. Based on the periodic measurements for rotational output and consistency in the lateral beam profile shape, the on-board detector proved to be a viable dosimetric quality assurance tool for IMRT with Tomotherapy. Tomotherapy was stable from the dosimetric point of view during the twelve-month period.http://www.jmp.org.in/article.asp?issn=0971-6203;year=2009;volume=34;issue=2;spage=73;epage=79;aulast=KinhikarTomotherapymega voltage computed tomography detectorDosimetric stability
spellingShingle Kinhikar Rajesh
Master Zubin
Dhote Dipak
Deshpande Deepak
Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
Journal of Medical Physics
Tomotherapy
mega voltage computed tomography detector
Dosimetric stability
title Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
title_full Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
title_fullStr Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
title_full_unstemmed Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
title_short Initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post-repair dosimetric parameters of a first helical Hi-Art II tomotherapy machine in India
title_sort initial dosimetric experience with mega voltage computed tomography detectors and estimation of pre and post repair dosimetric parameters of a first helical hi art ii tomotherapy machine in india
topic Tomotherapy
mega voltage computed tomography detector
Dosimetric stability
url http://www.jmp.org.in/article.asp?issn=0971-6203;year=2009;volume=34;issue=2;spage=73;epage=79;aulast=Kinhikar
work_keys_str_mv AT kinhikarrajesh initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia
AT masterzubin initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia
AT dhotedipak initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia
AT deshpandedeepak initialdosimetricexperiencewithmegavoltagecomputedtomographydetectorsandestimationofpreandpostrepairdosimetricparametersofafirsthelicalhiartiitomotherapymachineinindia