Ion-selective electrode and anion gap range: What should the anion gap be?

Seyed-Ali Sadjadi, Rendell Manalo, Navin Jaipaul, James McMillan Jerry L Pettis Memorial Veterans Medical Center, Loma Linda University School of Medicine, Loma Linda, CA, USA Background: Using flame photometry technique in the 1970s, the normal value of anion gap (AG) was determined to be 12 &p...

Full description

Bibliographic Details
Main Authors: Sadjadi SA, Manalo R, Jaipaul N, McMillan J
Format: Article
Language:English
Published: Dove Medical Press 2013-06-01
Series:International Journal of Nephrology and Renovascular Disease
Online Access:http://www.dovepress.com/ion-selective-electrode-and-anion-gap-range-what-should-the-anion-gap--a13289
_version_ 1819017296850452480
author Sadjadi SA
Manalo R
Jaipaul N
McMillan J
author_facet Sadjadi SA
Manalo R
Jaipaul N
McMillan J
author_sort Sadjadi SA
collection DOAJ
description Seyed-Ali Sadjadi, Rendell Manalo, Navin Jaipaul, James McMillan Jerry L Pettis Memorial Veterans Medical Center, Loma Linda University School of Medicine, Loma Linda, CA, USA Background: Using flame photometry technique in the 1970s, the normal value of anion gap (AG) was determined to be 12 &plusmn; 4 meq/L. However, with introduction of the autoanalyzers using an ion-selective electrode (ISE), the anion gap value has fallen to lower levels. Methods: A retrospective study of US veterans from a single medical center was performed to determine the value of the anion gap in subjects with normal renal function and normal serum albumin and in patients with lactic acidosis and end-stage renal disease on dialysis. Results: In 409 patients with an estimated glomerular filtration rate &ge;60 mL/min/1.73 m2 body surface area and serum albumin &ge;4 g/dL, the mean AG was 7.2 &plusmn; 2 (range 3&ndash;11) meq/L. In 299 patients with lactic acidosis (lactate level &ge;4 meq/L) and 68 patients with end-stage renal disease on dialysis, the mean AG was 12.5 meq/L and 12.4 meq/L, respectively. A value <2 meq/L should be considered a low anion gap and a possible clue to drug intoxication and paraproteinemic disorders. Conclusion: With the advent of ISE for measurement of analytes, the value of the anion gap has fallen. Physicians need to be aware of the normal AG value in their respective institutions, and laboratories need to have an established value for AG based on the type of instrument they are using. Keywords: acidosis, electrolytes, ESRD
first_indexed 2024-12-21T03:01:17Z
format Article
id doaj.art-29eec5946ccf4286bd491accf2d7fa6c
institution Directory Open Access Journal
issn 1178-7058
language English
last_indexed 2024-12-21T03:01:17Z
publishDate 2013-06-01
publisher Dove Medical Press
record_format Article
series International Journal of Nephrology and Renovascular Disease
spelling doaj.art-29eec5946ccf4286bd491accf2d7fa6c2022-12-21T19:18:10ZengDove Medical PressInternational Journal of Nephrology and Renovascular Disease1178-70582013-06-012013default101105Ion-selective electrode and anion gap range: What should the anion gap be?Sadjadi SAManalo RJaipaul NMcMillan JSeyed-Ali Sadjadi, Rendell Manalo, Navin Jaipaul, James McMillan Jerry L Pettis Memorial Veterans Medical Center, Loma Linda University School of Medicine, Loma Linda, CA, USA Background: Using flame photometry technique in the 1970s, the normal value of anion gap (AG) was determined to be 12 &plusmn; 4 meq/L. However, with introduction of the autoanalyzers using an ion-selective electrode (ISE), the anion gap value has fallen to lower levels. Methods: A retrospective study of US veterans from a single medical center was performed to determine the value of the anion gap in subjects with normal renal function and normal serum albumin and in patients with lactic acidosis and end-stage renal disease on dialysis. Results: In 409 patients with an estimated glomerular filtration rate &ge;60 mL/min/1.73 m2 body surface area and serum albumin &ge;4 g/dL, the mean AG was 7.2 &plusmn; 2 (range 3&ndash;11) meq/L. In 299 patients with lactic acidosis (lactate level &ge;4 meq/L) and 68 patients with end-stage renal disease on dialysis, the mean AG was 12.5 meq/L and 12.4 meq/L, respectively. A value <2 meq/L should be considered a low anion gap and a possible clue to drug intoxication and paraproteinemic disorders. Conclusion: With the advent of ISE for measurement of analytes, the value of the anion gap has fallen. Physicians need to be aware of the normal AG value in their respective institutions, and laboratories need to have an established value for AG based on the type of instrument they are using. Keywords: acidosis, electrolytes, ESRDhttp://www.dovepress.com/ion-selective-electrode-and-anion-gap-range-what-should-the-anion-gap--a13289
spellingShingle Sadjadi SA
Manalo R
Jaipaul N
McMillan J
Ion-selective electrode and anion gap range: What should the anion gap be?
International Journal of Nephrology and Renovascular Disease
title Ion-selective electrode and anion gap range: What should the anion gap be?
title_full Ion-selective electrode and anion gap range: What should the anion gap be?
title_fullStr Ion-selective electrode and anion gap range: What should the anion gap be?
title_full_unstemmed Ion-selective electrode and anion gap range: What should the anion gap be?
title_short Ion-selective electrode and anion gap range: What should the anion gap be?
title_sort ion selective electrode and anion gap range what should the anion gap be
url http://www.dovepress.com/ion-selective-electrode-and-anion-gap-range-what-should-the-anion-gap--a13289
work_keys_str_mv AT sadjadisa ionselectiveelectrodeandaniongaprangewhatshouldtheaniongapbe
AT manalor ionselectiveelectrodeandaniongaprangewhatshouldtheaniongapbe
AT jaipauln ionselectiveelectrodeandaniongaprangewhatshouldtheaniongapbe
AT mcmillanj ionselectiveelectrodeandaniongaprangewhatshouldtheaniongapbe