The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study
Purpose: Accurate refraction is arguably the most important parameter for a successful laser vision correction surgery and is based on a combination of manifest and cycloplegic refraction. Wavefront-based objective refraction may be useful in the evaluation of patients. So far, the reliability of ob...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2022-01-01
|
Series: | Indian Journal of Ophthalmology |
Subjects: | |
Online Access: | http://www.ijo.in/article.asp?issn=0301-4738;year=2022;volume=70;issue=5;spage=1533;epage=1537;aulast=Balparda |
_version_ | 1828820794534264832 |
---|---|
author | Kepa Balparda Andrea Acevedo-Urrego Laura Andrea Silva-Quintero Tatiana Herrera-Chalarca |
author_facet | Kepa Balparda Andrea Acevedo-Urrego Laura Andrea Silva-Quintero Tatiana Herrera-Chalarca |
author_sort | Kepa Balparda |
collection | DOAJ |
description | Purpose: Accurate refraction is arguably the most important parameter for a successful laser vision correction surgery and is based on a combination of manifest and cycloplegic refraction. Wavefront-based objective refraction may be useful in the evaluation of patients. So far, the reliability of objective refraction as measured using the Pentacam® AXL Wave has not been published in the literature. Methods: This was a prospective study including a total of 168 eyes belonging to 84 young non-presbyopic patients evaluated for refractive surgery. Pentacam® AXL Wave full sequence was taken for all patients. Then, a clinician who was unaware of the objective refraction results performed a full physical examination, including manifest refraction starting from an autorefractometer value. All refraction values were transferred to astigmatic power vectors as per the Thibos method. Reliability of the different vectors and a unifying blur value were compared using Spearman correlation, Bland–Altman plot, and intraclass correlation coefficient. Results: The mean age was 28.8 ± 5.4 years, with a female preponderance (60.7%). The correlation between both eyes was high. The difference in M vector between subjective and objective refraction was 0.16 D, while the difference was 0.04 and 0.01 D for the J0 and J45 vectors, respectively. Paired samples Student t was non-significant for all comparisons. Spearman rho correlations were high (0.666–0.924, all P < 0.001). Intraclass correlation coefficients were also high (0.890–0.966). Bland–Altman plots did not demonstrate any systematic errors. Conclusion: Wavefront-based refractive refraction obtained using the Pentacam® AXL Wave is highly agreeable and correlated with measurements obtained by manifest subjective refraction. |
first_indexed | 2024-12-12T12:41:27Z |
format | Article |
id | doaj.art-29f7237c676e4ca69a62ca54468e130f |
institution | Directory Open Access Journal |
issn | 0301-4738 1998-3689 |
language | English |
last_indexed | 2024-12-12T12:41:27Z |
publishDate | 2022-01-01 |
publisher | Wolters Kluwer Medknow Publications |
record_format | Article |
series | Indian Journal of Ophthalmology |
spelling | doaj.art-29f7237c676e4ca69a62ca54468e130f2022-12-22T00:24:13ZengWolters Kluwer Medknow PublicationsIndian Journal of Ophthalmology0301-47381998-36892022-01-017051533153710.4103/ijo.IJO_3006_21The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective studyKepa BalpardaAndrea Acevedo-UrregoLaura Andrea Silva-QuinteroTatiana Herrera-ChalarcaPurpose: Accurate refraction is arguably the most important parameter for a successful laser vision correction surgery and is based on a combination of manifest and cycloplegic refraction. Wavefront-based objective refraction may be useful in the evaluation of patients. So far, the reliability of objective refraction as measured using the Pentacam® AXL Wave has not been published in the literature. Methods: This was a prospective study including a total of 168 eyes belonging to 84 young non-presbyopic patients evaluated for refractive surgery. Pentacam® AXL Wave full sequence was taken for all patients. Then, a clinician who was unaware of the objective refraction results performed a full physical examination, including manifest refraction starting from an autorefractometer value. All refraction values were transferred to astigmatic power vectors as per the Thibos method. Reliability of the different vectors and a unifying blur value were compared using Spearman correlation, Bland–Altman plot, and intraclass correlation coefficient. Results: The mean age was 28.8 ± 5.4 years, with a female preponderance (60.7%). The correlation between both eyes was high. The difference in M vector between subjective and objective refraction was 0.16 D, while the difference was 0.04 and 0.01 D for the J0 and J45 vectors, respectively. Paired samples Student t was non-significant for all comparisons. Spearman rho correlations were high (0.666–0.924, all P < 0.001). Intraclass correlation coefficients were also high (0.890–0.966). Bland–Altman plots did not demonstrate any systematic errors. Conclusion: Wavefront-based refractive refraction obtained using the Pentacam® AXL Wave is highly agreeable and correlated with measurements obtained by manifest subjective refraction.http://www.ijo.in/article.asp?issn=0301-4738;year=2022;volume=70;issue=5;spage=1533;epage=1537;aulast=Balpardaobjective refractionpentacamrefractive surgical proceduressubjective refraction |
spellingShingle | Kepa Balparda Andrea Acevedo-Urrego Laura Andrea Silva-Quintero Tatiana Herrera-Chalarca The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study Indian Journal of Ophthalmology objective refraction pentacam refractive surgical procedures subjective refraction |
title | The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study |
title_full | The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study |
title_fullStr | The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study |
title_full_unstemmed | The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study |
title_short | The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study |
title_sort | pentacam r axl wave provides a reliable wavefront based objective refraction when compared to manifest subjective refraction a prospective study |
topic | objective refraction pentacam refractive surgical procedures subjective refraction |
url | http://www.ijo.in/article.asp?issn=0301-4738;year=2022;volume=70;issue=5;spage=1533;epage=1537;aulast=Balparda |
work_keys_str_mv | AT kepabalparda thepentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT andreaacevedourrego thepentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT lauraandreasilvaquintero thepentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT tatianaherrerachalarca thepentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT kepabalparda pentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT andreaacevedourrego pentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT lauraandreasilvaquintero pentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy AT tatianaherrerachalarca pentacamaxlwaveprovidesareliablewavefrontbasedobjectiverefractionwhencomparedtomanifestsubjectiverefractionaprospectivestudy |