An automatic immuno-microfluidic system integrating electrospun polystyrene microfibrous reactors for rapid detection of salivary cortisol

Summary: Conventional competitive enzyme-linked immunosorbent assay (ELISA) to measure the cortisol level in body fluid consumes a large amount of time, owing to complicated operations involved and requirement for precise control of reagent addition. We developed an automatic microfluidic system to ...

Full description

Bibliographic Details
Main Authors: Yecan Wang, Hiroshi Murakami, Toshihiro Kasama, Shigenobu Mitsuzawa, Satoru Shinkawa, Ryo Miyake, Madoka Takai
Format: Article
Language:English
Published: Elsevier 2023-10-01
Series:iScience
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2589004223018977
Description
Summary:Summary: Conventional competitive enzyme-linked immunosorbent assay (ELISA) to measure the cortisol level in body fluid consumes a large amount of time, owing to complicated operations involved and requirement for precise control of reagent addition. We developed an automatic microfluidic system to detect salivary cortisol rapidly, and an electrospun polystyrene (PS) microfiber-based reactor providing considerable binding sites for antibody immobilization, thus resolving the time limitations of competitive ELISA. Cortisol sample, horseradish peroxidase (HRP)-conjugated cortisol, and 3,3′,5,5′-tetramethylbenzidine (TMB) substrate were delivered to the PS reactor from containers in sequence by pumps automatically. The color variation due to oxidized TMB complex reflects the cortisol concentration level measured using an RGB phototransistor. In addition, the entire procedure from sample introduction to obtaining the photocurrent took only 15 min. This system can be implemented to quantify cortisol from 0.37 ng/mL to 30 ng/mL, and the limit of detection was estimated at 0.37 ng/mL.
ISSN:2589-0042