Assessment of Patient Dose from CT Examinations in Khorasan, Iran
Introduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, ther...
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
Mashhad University of Medical Sciences
2013-03-01
|
Series: | Iranian Journal of Medical Physics |
Subjects: | |
Online Access: | http://ijmp.mums.ac.ir/article_465_7dd2b6fb16de1612b04084752cdf967a.pdf |
_version_ | 1818538746840088576 |
---|---|
author | Mohammad Taghi Bahreyni Toossi Mohammad Bahrami |
author_facet | Mohammad Taghi Bahreyni Toossi Mohammad Bahrami |
author_sort | Mohammad Taghi Bahreyni Toossi |
collection | DOAJ |
description | Introduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, there are potential risks in addition to the clear potential benefits. Materials and Methods Two reference dose quantities have been defined in order to promote the use of good technique in CT. These are weighted CT dose index (CTDIw) in (mGy) for a single slice in serial scanning or per rotation in helical scanning, and dose–length product (DLP) per complete examination (mGy.cm), All measurements were performed using a pencil shaped ionization chamber introduced into polymethyl methacrylate cylindrical brain and body phantoms. This survey was performed on 7 CT scanners in Khorasan Province-Iran. Results DLP for brain, chest, abdomen and pelvic examinations had a range of 255 - 1026, 76-1277, 48-737, 69-854 mGy.cm, respectively. Conclusion The results obtained in this study show that the DLP values obtained in this province are lower than European Commission reference dose levels (EC RDL), in other words performance of all the scanners were satisfactory. |
first_indexed | 2024-12-11T21:32:57Z |
format | Article |
id | doaj.art-2c1b5ef18e514f26b0f4e3ab48936b61 |
institution | Directory Open Access Journal |
issn | 2345-3672 2345-3672 |
language | English |
last_indexed | 2024-12-11T21:32:57Z |
publishDate | 2013-03-01 |
publisher | Mashhad University of Medical Sciences |
record_format | Article |
series | Iranian Journal of Medical Physics |
spelling | doaj.art-2c1b5ef18e514f26b0f4e3ab48936b612022-12-22T00:50:06ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722013-03-019423323810.22038/ijmp.2013.465465Assessment of Patient Dose from CT Examinations in Khorasan, IranMohammad Taghi Bahreyni Toossi0Mohammad Bahrami1Medical Physics Research Center, School of Medicine, Mashhad University of Medical Sciences, Mashhad, Iran Department of Medical Physics, School of Medicine, Mashhad University of Medical Sciences, Mashhad, IranDepartment of Medical Physics, School of Medicine, Mashhad University of Medical Sciences, Mashhad, IranIntroduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, there are potential risks in addition to the clear potential benefits. Materials and Methods Two reference dose quantities have been defined in order to promote the use of good technique in CT. These are weighted CT dose index (CTDIw) in (mGy) for a single slice in serial scanning or per rotation in helical scanning, and dose–length product (DLP) per complete examination (mGy.cm), All measurements were performed using a pencil shaped ionization chamber introduced into polymethyl methacrylate cylindrical brain and body phantoms. This survey was performed on 7 CT scanners in Khorasan Province-Iran. Results DLP for brain, chest, abdomen and pelvic examinations had a range of 255 - 1026, 76-1277, 48-737, 69-854 mGy.cm, respectively. Conclusion The results obtained in this study show that the DLP values obtained in this province are lower than European Commission reference dose levels (EC RDL), in other words performance of all the scanners were satisfactory.http://ijmp.mums.ac.ir/article_465_7dd2b6fb16de1612b04084752cdf967a.pdfDLPCTIonization Chamber |
spellingShingle | Mohammad Taghi Bahreyni Toossi Mohammad Bahrami Assessment of Patient Dose from CT Examinations in Khorasan, Iran Iranian Journal of Medical Physics DLP CT Ionization Chamber |
title | Assessment of Patient Dose from CT Examinations in Khorasan, Iran |
title_full | Assessment of Patient Dose from CT Examinations in Khorasan, Iran |
title_fullStr | Assessment of Patient Dose from CT Examinations in Khorasan, Iran |
title_full_unstemmed | Assessment of Patient Dose from CT Examinations in Khorasan, Iran |
title_short | Assessment of Patient Dose from CT Examinations in Khorasan, Iran |
title_sort | assessment of patient dose from ct examinations in khorasan iran |
topic | DLP CT Ionization Chamber |
url | http://ijmp.mums.ac.ir/article_465_7dd2b6fb16de1612b04084752cdf967a.pdf |
work_keys_str_mv | AT mohammadtaghibahreynitoossi assessmentofpatientdosefromctexaminationsinkhorasaniran AT mohammadbahrami assessmentofpatientdosefromctexaminationsinkhorasaniran |