Comparison of X-ray Radiant Power Absolute Measurement between a Free-Air Ionization Chamber and a Cryogenic Electrical Substitution Radiometer
Absolute measurement of radiant power in the X-ray region is essential for many applications in astrophysics, spectroscopy, and X-ray diagnostics. Comparison between different measuring methods is an effective way to check their reliability. In the present work, a comparison of X-ray radiant power a...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2023-01-01
|
Series: | Sensors |
Subjects: | |
Online Access: | https://www.mdpi.com/1424-8220/23/2/1006 |
Summary: | Absolute measurement of radiant power in the X-ray region is essential for many applications in astrophysics, spectroscopy, and X-ray diagnostics. Comparison between different measuring methods is an effective way to check their reliability. In the present work, a comparison of X-ray radiant power absolute measurement between a free-air ionization chamber and a cryogenic electrical substitution radiometer was performed at Beijing Synchrotron Radiation Facility. The absolute radiant power obtained by these two methods were mutually compared via a transfer standard detector’s spectral responsivity at a photon energy of 10 keV. The result of the comparison showed that the difference was 0.47%. A conclusion was reached that the free-air ionization chamber and the cryogenic electrical substitution radiometer agreed within the combined relative uncertainty of 3.35%. |
---|---|
ISSN: | 1424-8220 |