A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring

Abstract Background Automated measurement of blood pressure (BP) in designated BP kiosks have in recent years been introduced in primary care. If kiosk blood pressure (BP) monitoring provides results equivalent to in-office BP or daytime ambulatory BP monitoring (ABPM), follow-up of adult patients c...

Full description

Bibliographic Details
Main Authors: Gunnar Nilsson, Anna Lindam
Format: Article
Language:English
Published: BMC 2024-01-01
Series:BMC Cardiovascular Disorders
Subjects:
Online Access:https://doi.org/10.1186/s12872-023-03701-1
_version_ 1797363646236459008
author Gunnar Nilsson
Anna Lindam
author_facet Gunnar Nilsson
Anna Lindam
author_sort Gunnar Nilsson
collection DOAJ
description Abstract Background Automated measurement of blood pressure (BP) in designated BP kiosks have in recent years been introduced in primary care. If kiosk blood pressure (BP) monitoring provides results equivalent to in-office BP or daytime ambulatory BP monitoring (ABPM), follow-up of adult patients could be managed primarily by self-checks. We therefore designed a comparative trial and evaluated the diagnostic performance of kiosk- and office-based BP (nurse- versus physician-measured) compared with daytime ABPM. Methods A trial of automated BP monitoring in three settings: a designated BP kiosk, by nurses and physicians in clinic, and by ABPM. The primary outcome was systolic and diastolic BP, with respective diagnostic thresholds of ≥135 mmHg and/or ≥ 85 mmHg for daytime ABPM and kiosk BP and ≥ 140 mmHg and/or ≥ 90 mmHg for office BP (nurse- and physician-measured). Results Compared with daytime ABPM, mean systolic kiosk BP was higher by 6.2 mmHg (95% confidence interval [CI] 3.8–8.6) and diastolic by 7.9 mmHg (95% CI 6.2–9.6; p < 0.001). Mean systolic BP taken by nurses was similar to daytime ABPM values (+ 2.0 mmHg; 95% CI − 0.2–4.2; p = 0.071), but nurse-measured diastolic values were higher, by 7.2 mmHg (95% CI 5.9–9.6; p < 0.001). Mean systolic and diastolic physician-measured BPs were higher compared with daytime ABPM (systolic, by 7.6 mmHg [95% CI 4.5–10.2] and diastolic by 5.8 mmHg [95% CI 4.1–7.6]; p < 0.001). Receiver operating characteristic curves of BP monitoring across pairs of systolic/diastolic cut-off levels among the three settings, with daytime ABPM as reference, demonstrated overall similar diagnostic performance between kiosk and nurse-measured values and over the curve performance for physician-measured BP. Accuracy with nurse-measured BP was 69.2% (95% CI 60.0–77.4%), compared with 65.8% (95% CI 56.5–74.3%) for kiosk BP. Conclusions In this study kiosk BP monitoring was not comparable to daytime ABPM but could be an alternative to in-office BP monitoring by trained nurses. The diagnostic performance of kiosk and nurse-measured BP monitoring was similar and better than that of physician-measured BP. Trial registration ClinicalTrials.gov (NCT04488289) 27/07/2020.
first_indexed 2024-03-08T16:24:04Z
format Article
id doaj.art-2cd2da1566fe4b4bbf304c2279f5914f
institution Directory Open Access Journal
issn 1471-2261
language English
last_indexed 2024-03-08T16:24:04Z
publishDate 2024-01-01
publisher BMC
record_format Article
series BMC Cardiovascular Disorders
spelling doaj.art-2cd2da1566fe4b4bbf304c2279f5914f2024-01-07T12:09:50ZengBMCBMC Cardiovascular Disorders1471-22612024-01-012411910.1186/s12872-023-03701-1A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoringGunnar Nilsson0Anna Lindam1Department of Public Health and Clinical Medicine, Umeå UniversityDepartment of Public Health and Clinical Medicine, Unit of Research, Education, and Development, Östersund Hospital, Umeå UniversityAbstract Background Automated measurement of blood pressure (BP) in designated BP kiosks have in recent years been introduced in primary care. If kiosk blood pressure (BP) monitoring provides results equivalent to in-office BP or daytime ambulatory BP monitoring (ABPM), follow-up of adult patients could be managed primarily by self-checks. We therefore designed a comparative trial and evaluated the diagnostic performance of kiosk- and office-based BP (nurse- versus physician-measured) compared with daytime ABPM. Methods A trial of automated BP monitoring in three settings: a designated BP kiosk, by nurses and physicians in clinic, and by ABPM. The primary outcome was systolic and diastolic BP, with respective diagnostic thresholds of ≥135 mmHg and/or ≥ 85 mmHg for daytime ABPM and kiosk BP and ≥ 140 mmHg and/or ≥ 90 mmHg for office BP (nurse- and physician-measured). Results Compared with daytime ABPM, mean systolic kiosk BP was higher by 6.2 mmHg (95% confidence interval [CI] 3.8–8.6) and diastolic by 7.9 mmHg (95% CI 6.2–9.6; p < 0.001). Mean systolic BP taken by nurses was similar to daytime ABPM values (+ 2.0 mmHg; 95% CI − 0.2–4.2; p = 0.071), but nurse-measured diastolic values were higher, by 7.2 mmHg (95% CI 5.9–9.6; p < 0.001). Mean systolic and diastolic physician-measured BPs were higher compared with daytime ABPM (systolic, by 7.6 mmHg [95% CI 4.5–10.2] and diastolic by 5.8 mmHg [95% CI 4.1–7.6]; p < 0.001). Receiver operating characteristic curves of BP monitoring across pairs of systolic/diastolic cut-off levels among the three settings, with daytime ABPM as reference, demonstrated overall similar diagnostic performance between kiosk and nurse-measured values and over the curve performance for physician-measured BP. Accuracy with nurse-measured BP was 69.2% (95% CI 60.0–77.4%), compared with 65.8% (95% CI 56.5–74.3%) for kiosk BP. Conclusions In this study kiosk BP monitoring was not comparable to daytime ABPM but could be an alternative to in-office BP monitoring by trained nurses. The diagnostic performance of kiosk and nurse-measured BP monitoring was similar and better than that of physician-measured BP. Trial registration ClinicalTrials.gov (NCT04488289) 27/07/2020.https://doi.org/10.1186/s12872-023-03701-1Ambulatory blood pressureSelf-blood pressure monitoringClinical trial
spellingShingle Gunnar Nilsson
Anna Lindam
A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
BMC Cardiovascular Disorders
Ambulatory blood pressure
Self-blood pressure monitoring
Clinical trial
title A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
title_full A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
title_fullStr A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
title_full_unstemmed A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
title_short A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring
title_sort comparative trial of blood pressure monitoring in a self care kiosk in office and with ambulatory blood pressure monitoring
topic Ambulatory blood pressure
Self-blood pressure monitoring
Clinical trial
url https://doi.org/10.1186/s12872-023-03701-1
work_keys_str_mv AT gunnarnilsson acomparativetrialofbloodpressuremonitoringinaselfcarekioskinofficeandwithambulatorybloodpressuremonitoring
AT annalindam acomparativetrialofbloodpressuremonitoringinaselfcarekioskinofficeandwithambulatorybloodpressuremonitoring
AT gunnarnilsson comparativetrialofbloodpressuremonitoringinaselfcarekioskinofficeandwithambulatorybloodpressuremonitoring
AT annalindam comparativetrialofbloodpressuremonitoringinaselfcarekioskinofficeandwithambulatorybloodpressuremonitoring