Collapse of the Balloon-Expandable Stent in the Common Iliac Artery Due to Minor External Compression in a Lean Patient
A 60-year-old lean man with claudication due to severe stenosis of the right common iliac artery (CIA) underwent balloon-expandable stent deployment. Nineteen months later, claudication and lower limb ischemia in the patient recurred after minor external pressure and computed tomography indicated co...
Main Authors: | , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
The Korean Society of Radiology
2017-08-01
|
Series: | 대한영상의학회지 |
Subjects: | |
Online Access: | https://doi.org/10.3348/jksr.2017.77.2.121 |
Summary: | A 60-year-old lean man with claudication due to severe stenosis of the right common iliac artery (CIA) underwent balloon-expandable stent deployment. Nineteen months later, claudication and lower limb ischemia in the patient recurred after minor external pressure and computed tomography indicated collapse of the right CIA stent. We performed balloon angioplasty again and self-expandable stent deployment was performed, and the patient's symptoms were relieved. We suggest that the CIA in a lean person could be susceptible to external pressure. |
---|---|