A technique to retrieve an internalised embedded central venous catheter

Abstract Background Central venous catheters may become embedded due to the formation of adhesions between the indwelling catheter and the vein wall. Case presentation A 49-year-old patient with bacteraemia was referred for retrieval of an embedded internalised central venous dialysis catheter. Rece...

Full description

Bibliographic Details
Main Authors: John Finnegan, Pradeep Govender
Format: Article
Language:English
Published: SpringerOpen 2024-02-01
Series:CVIR Endovascular
Subjects:
Online Access:https://doi.org/10.1186/s42155-024-00436-8
_version_ 1797273089154744320
author John Finnegan
Pradeep Govender
author_facet John Finnegan
Pradeep Govender
author_sort John Finnegan
collection DOAJ
description Abstract Background Central venous catheters may become embedded due to the formation of adhesions between the indwelling catheter and the vein wall. Case presentation A 49-year-old patient with bacteraemia was referred for retrieval of an embedded internalised central venous dialysis catheter. Recently the catheter had been surgically ligated at the venotomy site internalising the intravascular catheter component, which precluded antegrade ballooning through the catheter hub. Seldinger technique was used to access the catheter lumen within the left internal jugular vein and through and through access was established across the catheter. Retrograde endoluminal balloon dilation was performed to disrupt adhesions and free the catheter. The catheter was snared over the wire and removed from the right femoral vein. Conclusion This case report outlines an effective, minimally invasive retrieval method in a rare case of an embedded internalised central venous catheter. Graphical Abstract
first_indexed 2024-03-07T14:38:36Z
format Article
id doaj.art-2df8db339ec441faac008e435466842a
institution Directory Open Access Journal
issn 2520-8934
language English
last_indexed 2024-03-07T14:38:36Z
publishDate 2024-02-01
publisher SpringerOpen
record_format Article
series CVIR Endovascular
spelling doaj.art-2df8db339ec441faac008e435466842a2024-03-05T20:28:23ZengSpringerOpenCVIR Endovascular2520-89342024-02-01711510.1186/s42155-024-00436-8A technique to retrieve an internalised embedded central venous catheterJohn Finnegan0Pradeep Govender1Department of Radiology, Tallaght University HospitalDepartment of Radiology, Tallaght University HospitalAbstract Background Central venous catheters may become embedded due to the formation of adhesions between the indwelling catheter and the vein wall. Case presentation A 49-year-old patient with bacteraemia was referred for retrieval of an embedded internalised central venous dialysis catheter. Recently the catheter had been surgically ligated at the venotomy site internalising the intravascular catheter component, which precluded antegrade ballooning through the catheter hub. Seldinger technique was used to access the catheter lumen within the left internal jugular vein and through and through access was established across the catheter. Retrograde endoluminal balloon dilation was performed to disrupt adhesions and free the catheter. The catheter was snared over the wire and removed from the right femoral vein. Conclusion This case report outlines an effective, minimally invasive retrieval method in a rare case of an embedded internalised central venous catheter. Graphical Abstracthttps://doi.org/10.1186/s42155-024-00436-8Venous interventionDialysis accessCentral venous access
spellingShingle John Finnegan
Pradeep Govender
A technique to retrieve an internalised embedded central venous catheter
CVIR Endovascular
Venous intervention
Dialysis access
Central venous access
title A technique to retrieve an internalised embedded central venous catheter
title_full A technique to retrieve an internalised embedded central venous catheter
title_fullStr A technique to retrieve an internalised embedded central venous catheter
title_full_unstemmed A technique to retrieve an internalised embedded central venous catheter
title_short A technique to retrieve an internalised embedded central venous catheter
title_sort technique to retrieve an internalised embedded central venous catheter
topic Venous intervention
Dialysis access
Central venous access
url https://doi.org/10.1186/s42155-024-00436-8
work_keys_str_mv AT johnfinnegan atechniquetoretrieveaninternalisedembeddedcentralvenouscatheter
AT pradeepgovender atechniquetoretrieveaninternalisedembeddedcentralvenouscatheter
AT johnfinnegan techniquetoretrieveaninternalisedembeddedcentralvenouscatheter
AT pradeepgovender techniquetoretrieveaninternalisedembeddedcentralvenouscatheter