Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam
Objectives: To research the repeatability and intercompatibility of keratometry values measured with Potec PRK-6000 autorefractometer, IOL Master, and Pentacam. Materials and Methods: In this prospective study, consecutive measurements were performed in two different sessions with the mentioned t...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Galenos Yayinevi
2014-05-01
|
Series: | Türk Oftalmoloji Dergisi |
Subjects: | |
Online Access: | http://www.oftalmoloji.org/article_6088/Repeatability-And-Comparison-Of-Keratometry-Values-Measured-With-Potec-Prk-6000-Autorefractometer-Iolmaster-And-Pentacam |
_version_ | 1797924239897001984 |
---|---|
author | Adem Türk Ceyhun Arıcı Osman Melih Ceylan Mehmet Kola |
author_facet | Adem Türk Ceyhun Arıcı Osman Melih Ceylan Mehmet Kola |
author_sort | Adem Türk |
collection | DOAJ |
description | Objectives: To research the repeatability and intercompatibility of keratometry values measured with Potec PRK-6000
autorefractometer, IOL Master, and Pentacam.
Materials and Methods: In this prospective study, consecutive measurements were performed in two different sessions with the
mentioned three devices on 110 eyes of 55 subjects who had no additional ocular pathology except for refraction error. The consistency
of flat and steep keratometry, average keratometry, and corneal astigmatism values obtained in both sessions was compared by using
intraclass correlation coefficient (ICC). The measurement differences between the devices were statistically compared as well.
Results: The mean age of the study subjects was 23.05±3.01 (18-30) years. ICC values of average keratometry measurements obtained
in the sessions were 0.996 for Potec PRK-6000 autorefractometer, 0.997 for IOL Master, and 0.999 for Pentacam. There was high
compatibility between the three devices in terms of average keratometry values in Bland-Altman analysis. However, there were
statistically significant differences between the devices in terms of parameters other than corneal astigmatism.
Conclusion: The repeatability of the three devices was found considerably high in keratometry measurements. However, it is not
appropriate for these devices to be substituted for each other in keratometry measurements. (Turk J Ophthalmol 2014; 44: 179-83) |
first_indexed | 2024-04-10T14:58:27Z |
format | Article |
id | doaj.art-2e78f26db7de4faa861aa8c01bc615ae |
institution | Directory Open Access Journal |
issn | 1300-0659 2147-2661 |
language | English |
last_indexed | 2024-04-10T14:58:27Z |
publishDate | 2014-05-01 |
publisher | Galenos Yayinevi |
record_format | Article |
series | Türk Oftalmoloji Dergisi |
spelling | doaj.art-2e78f26db7de4faa861aa8c01bc615ae2023-02-15T16:07:13ZengGalenos YayineviTürk Oftalmoloji Dergisi1300-06592147-26612014-05-0144317918310.4274/tjo.26086Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and PentacamAdem Türk0Ceyhun Arıcı1Osman Melih Ceylan2Mehmet Kola3Karadeniz Teknik Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Trabzon, Türkiyeİstanbul Üniversitesi, Cerrahpaşa Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, İstanbul, TürkiyeGülhane Askeri Tıp Akademisi, Göz Hastalıkları Anabilim Dalı, Ankara, TürkiyeKaradeniz Teknik Üniversitesi Tıp Fakültesi, Göz Hastalıkları Anabilim Dalı, Trabzon, TürkiyeObjectives: To research the repeatability and intercompatibility of keratometry values measured with Potec PRK-6000 autorefractometer, IOL Master, and Pentacam. Materials and Methods: In this prospective study, consecutive measurements were performed in two different sessions with the mentioned three devices on 110 eyes of 55 subjects who had no additional ocular pathology except for refraction error. The consistency of flat and steep keratometry, average keratometry, and corneal astigmatism values obtained in both sessions was compared by using intraclass correlation coefficient (ICC). The measurement differences between the devices were statistically compared as well. Results: The mean age of the study subjects was 23.05±3.01 (18-30) years. ICC values of average keratometry measurements obtained in the sessions were 0.996 for Potec PRK-6000 autorefractometer, 0.997 for IOL Master, and 0.999 for Pentacam. There was high compatibility between the three devices in terms of average keratometry values in Bland-Altman analysis. However, there were statistically significant differences between the devices in terms of parameters other than corneal astigmatism. Conclusion: The repeatability of the three devices was found considerably high in keratometry measurements. However, it is not appropriate for these devices to be substituted for each other in keratometry measurements. (Turk J Ophthalmol 2014; 44: 179-83)http://www.oftalmoloji.org/article_6088/Repeatability-And-Comparison-Of-Keratometry-Values-Measured-With-Potec-Prk-6000-Autorefractometer-Iolmaster-And-PentacamBiometryanterior eye segmentrefractive errorscornearefractometry |
spellingShingle | Adem Türk Ceyhun Arıcı Osman Melih Ceylan Mehmet Kola Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam Türk Oftalmoloji Dergisi Biometry anterior eye segment refractive errors cornea refractometry |
title | Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam |
title_full | Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam |
title_fullStr | Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam |
title_full_unstemmed | Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam |
title_short | Repeatability and Comparison of Keratometry Values Measured with Potec PRK-6000 Autorefractometer, IOLMaster, and Pentacam |
title_sort | repeatability and comparison of keratometry values measured with potec prk 6000 autorefractometer iolmaster and pentacam |
topic | Biometry anterior eye segment refractive errors cornea refractometry |
url | http://www.oftalmoloji.org/article_6088/Repeatability-And-Comparison-Of-Keratometry-Values-Measured-With-Potec-Prk-6000-Autorefractometer-Iolmaster-And-Pentacam |
work_keys_str_mv | AT ademturk repeatabilityandcomparisonofkeratometryvaluesmeasuredwithpotecprk6000autorefractometeriolmasterandpentacam AT ceyhunarıcı repeatabilityandcomparisonofkeratometryvaluesmeasuredwithpotecprk6000autorefractometeriolmasterandpentacam AT osmanmelihceylan repeatabilityandcomparisonofkeratometryvaluesmeasuredwithpotecprk6000autorefractometeriolmasterandpentacam AT mehmetkola repeatabilityandcomparisonofkeratometryvaluesmeasuredwithpotecprk6000autorefractometeriolmasterandpentacam |