Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation

<i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and no...

Full description

Bibliographic Details
Main Authors: William T. Hrinivich, Heng Li, Anh Tran, Sahaja Acharya, Matthew M. Ladra, Khadija Sheikh
Format: Article
Language:English
Published: MDPI AG 2023-05-01
Series:Cancers
Subjects:
Online Access:https://www.mdpi.com/2072-6694/15/11/2882
_version_ 1797597782018621440
author William T. Hrinivich
Heng Li
Anh Tran
Sahaja Acharya
Matthew M. Ladra
Khadija Sheikh
author_facet William T. Hrinivich
Heng Li
Anh Tran
Sahaja Acharya
Matthew M. Ladra
Khadija Sheikh
author_sort William T. Hrinivich
collection DOAJ
description <i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and normal tissue dose for image-guided pediatric craniospinal irradiation (CSI). <i>Methods</i>: A custom RSB was designed and manufactured from a 3.5 cm thick slab of polymethyl methacrylate (PMMA) to be placed directly under patients, on top of our existing couch top. The relative linear stopping power (RLSP) of the RSB was measured using a multi-layer ionization chamber, and output constancy was measured using an ion chamber. End-to-end tests were performed using the MRS and RSB approaches using an anthropomorphic phantom and radiochromic film measurements. Cone beam CT (CBCT) and 2D planar kV X-ray image quality were compared with and without the RSB present using image quality phantoms. CSI plans were produced using MRS and RSB approaches for two retrospective pediatric patients, and the resultant normal tissue doses were compared. <i>Results</i>: The RLSP of the RSB was found to be 1.163 and provided computed penumbra of 6.9 mm in the phantom compared to 11.8 mm using the MRS. Phantom measurements using the RSB demonstrated errors in output constancy, range, and penumbra of 0.3%, −0.8%, and 0.6 mm, respectively. The RSB reduced mean kidney and lung dose compared to the MRS by 57.7% and 46.3%, respectively. The RSB decreased mean CBCT image intensities by 86.8 HU but did not significantly impact CBCT or kV spatial resolution providing acceptable image quality for patient setup. <i>Conclusions</i>: A custom RSB for pediatric proton CSI was designed, manufactured, modeled in our TPS, and found to significantly reduce lateral proton beam penumbra compared to a standard MRS while maintaining CBCT and kV image-quality and is in routine use at our center.
first_indexed 2024-03-11T03:10:19Z
format Article
id doaj.art-2ff79be66a9b4f2996bfe4af5ca14963
institution Directory Open Access Journal
issn 2072-6694
language English
last_indexed 2024-03-11T03:10:19Z
publishDate 2023-05-01
publisher MDPI AG
record_format Article
series Cancers
spelling doaj.art-2ff79be66a9b4f2996bfe4af5ca149632023-11-18T07:37:51ZengMDPI AGCancers2072-66942023-05-011511288210.3390/cancers15112882Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal IrradiationWilliam T. Hrinivich0Heng Li1Anh Tran2Sahaja Acharya3Matthew M. Ladra4Khadija Sheikh5Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA<i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and normal tissue dose for image-guided pediatric craniospinal irradiation (CSI). <i>Methods</i>: A custom RSB was designed and manufactured from a 3.5 cm thick slab of polymethyl methacrylate (PMMA) to be placed directly under patients, on top of our existing couch top. The relative linear stopping power (RLSP) of the RSB was measured using a multi-layer ionization chamber, and output constancy was measured using an ion chamber. End-to-end tests were performed using the MRS and RSB approaches using an anthropomorphic phantom and radiochromic film measurements. Cone beam CT (CBCT) and 2D planar kV X-ray image quality were compared with and without the RSB present using image quality phantoms. CSI plans were produced using MRS and RSB approaches for two retrospective pediatric patients, and the resultant normal tissue doses were compared. <i>Results</i>: The RLSP of the RSB was found to be 1.163 and provided computed penumbra of 6.9 mm in the phantom compared to 11.8 mm using the MRS. Phantom measurements using the RSB demonstrated errors in output constancy, range, and penumbra of 0.3%, −0.8%, and 0.6 mm, respectively. The RSB reduced mean kidney and lung dose compared to the MRS by 57.7% and 46.3%, respectively. The RSB decreased mean CBCT image intensities by 86.8 HU but did not significantly impact CBCT or kV spatial resolution providing acceptable image quality for patient setup. <i>Conclusions</i>: A custom RSB for pediatric proton CSI was designed, manufactured, modeled in our TPS, and found to significantly reduce lateral proton beam penumbra compared to a standard MRS while maintaining CBCT and kV image-quality and is in routine use at our center.https://www.mdpi.com/2072-6694/15/11/2882craniospinal irradiationpediatric radiotherapyproton therapypencil beam scanningimage-guided radiotherapy
spellingShingle William T. Hrinivich
Heng Li
Anh Tran
Sahaja Acharya
Matthew M. Ladra
Khadija Sheikh
Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
Cancers
craniospinal irradiation
pediatric radiotherapy
proton therapy
pencil beam scanning
image-guided radiotherapy
title Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
title_full Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
title_fullStr Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
title_full_unstemmed Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
title_short Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
title_sort clinical characterization of a table mounted range shifter board for synchrotron based intensity modulated proton therapy for pediatric craniospinal irradiation
topic craniospinal irradiation
pediatric radiotherapy
proton therapy
pencil beam scanning
image-guided radiotherapy
url https://www.mdpi.com/2072-6694/15/11/2882
work_keys_str_mv AT williamthrinivich clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation
AT hengli clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation
AT anhtran clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation
AT sahajaacharya clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation
AT matthewmladra clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation
AT khadijasheikh clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation