Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation
<i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and no...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2023-05-01
|
Series: | Cancers |
Subjects: | |
Online Access: | https://www.mdpi.com/2072-6694/15/11/2882 |
_version_ | 1797597782018621440 |
---|---|
author | William T. Hrinivich Heng Li Anh Tran Sahaja Acharya Matthew M. Ladra Khadija Sheikh |
author_facet | William T. Hrinivich Heng Li Anh Tran Sahaja Acharya Matthew M. Ladra Khadija Sheikh |
author_sort | William T. Hrinivich |
collection | DOAJ |
description | <i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and normal tissue dose for image-guided pediatric craniospinal irradiation (CSI). <i>Methods</i>: A custom RSB was designed and manufactured from a 3.5 cm thick slab of polymethyl methacrylate (PMMA) to be placed directly under patients, on top of our existing couch top. The relative linear stopping power (RLSP) of the RSB was measured using a multi-layer ionization chamber, and output constancy was measured using an ion chamber. End-to-end tests were performed using the MRS and RSB approaches using an anthropomorphic phantom and radiochromic film measurements. Cone beam CT (CBCT) and 2D planar kV X-ray image quality were compared with and without the RSB present using image quality phantoms. CSI plans were produced using MRS and RSB approaches for two retrospective pediatric patients, and the resultant normal tissue doses were compared. <i>Results</i>: The RLSP of the RSB was found to be 1.163 and provided computed penumbra of 6.9 mm in the phantom compared to 11.8 mm using the MRS. Phantom measurements using the RSB demonstrated errors in output constancy, range, and penumbra of 0.3%, −0.8%, and 0.6 mm, respectively. The RSB reduced mean kidney and lung dose compared to the MRS by 57.7% and 46.3%, respectively. The RSB decreased mean CBCT image intensities by 86.8 HU but did not significantly impact CBCT or kV spatial resolution providing acceptable image quality for patient setup. <i>Conclusions</i>: A custom RSB for pediatric proton CSI was designed, manufactured, modeled in our TPS, and found to significantly reduce lateral proton beam penumbra compared to a standard MRS while maintaining CBCT and kV image-quality and is in routine use at our center. |
first_indexed | 2024-03-11T03:10:19Z |
format | Article |
id | doaj.art-2ff79be66a9b4f2996bfe4af5ca14963 |
institution | Directory Open Access Journal |
issn | 2072-6694 |
language | English |
last_indexed | 2024-03-11T03:10:19Z |
publishDate | 2023-05-01 |
publisher | MDPI AG |
record_format | Article |
series | Cancers |
spelling | doaj.art-2ff79be66a9b4f2996bfe4af5ca149632023-11-18T07:37:51ZengMDPI AGCancers2072-66942023-05-011511288210.3390/cancers15112882Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal IrradiationWilliam T. Hrinivich0Heng Li1Anh Tran2Sahaja Acharya3Matthew M. Ladra4Khadija Sheikh5Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USADepartment of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA<i>Purpose</i>: To report our design, manufacturing, commissioning and initial clinical experience with a table-mounted range shifter board (RSB) intended to replace the machine-mounted range shifter (MRS) in a synchrotron-based pencil beam scanning (PBS) system to reduce penumbra and normal tissue dose for image-guided pediatric craniospinal irradiation (CSI). <i>Methods</i>: A custom RSB was designed and manufactured from a 3.5 cm thick slab of polymethyl methacrylate (PMMA) to be placed directly under patients, on top of our existing couch top. The relative linear stopping power (RLSP) of the RSB was measured using a multi-layer ionization chamber, and output constancy was measured using an ion chamber. End-to-end tests were performed using the MRS and RSB approaches using an anthropomorphic phantom and radiochromic film measurements. Cone beam CT (CBCT) and 2D planar kV X-ray image quality were compared with and without the RSB present using image quality phantoms. CSI plans were produced using MRS and RSB approaches for two retrospective pediatric patients, and the resultant normal tissue doses were compared. <i>Results</i>: The RLSP of the RSB was found to be 1.163 and provided computed penumbra of 6.9 mm in the phantom compared to 11.8 mm using the MRS. Phantom measurements using the RSB demonstrated errors in output constancy, range, and penumbra of 0.3%, −0.8%, and 0.6 mm, respectively. The RSB reduced mean kidney and lung dose compared to the MRS by 57.7% and 46.3%, respectively. The RSB decreased mean CBCT image intensities by 86.8 HU but did not significantly impact CBCT or kV spatial resolution providing acceptable image quality for patient setup. <i>Conclusions</i>: A custom RSB for pediatric proton CSI was designed, manufactured, modeled in our TPS, and found to significantly reduce lateral proton beam penumbra compared to a standard MRS while maintaining CBCT and kV image-quality and is in routine use at our center.https://www.mdpi.com/2072-6694/15/11/2882craniospinal irradiationpediatric radiotherapyproton therapypencil beam scanningimage-guided radiotherapy |
spellingShingle | William T. Hrinivich Heng Li Anh Tran Sahaja Acharya Matthew M. Ladra Khadija Sheikh Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation Cancers craniospinal irradiation pediatric radiotherapy proton therapy pencil beam scanning image-guided radiotherapy |
title | Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation |
title_full | Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation |
title_fullStr | Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation |
title_full_unstemmed | Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation |
title_short | Clinical Characterization of a Table Mounted Range Shifter Board for Synchrotron-Based Intensity Modulated Proton Therapy for Pediatric Craniospinal Irradiation |
title_sort | clinical characterization of a table mounted range shifter board for synchrotron based intensity modulated proton therapy for pediatric craniospinal irradiation |
topic | craniospinal irradiation pediatric radiotherapy proton therapy pencil beam scanning image-guided radiotherapy |
url | https://www.mdpi.com/2072-6694/15/11/2882 |
work_keys_str_mv | AT williamthrinivich clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation AT hengli clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation AT anhtran clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation AT sahajaacharya clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation AT matthewmladra clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation AT khadijasheikh clinicalcharacterizationofatablemountedrangeshifterboardforsynchrotronbasedintensitymodulatedprotontherapyforpediatriccraniospinalirradiation |