Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital
Study of collimation field difference suitability test on analog and digital mammography at leading hospital in Makassar from 2016-2018. This test was carried out by using four coins of the same size placed on the edge of the collimator lamp field, one different size coin was placed on the table edg...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Lambung Mangkurat University Press
2019-08-01
|
Series: | Jurnal Fisika Flux |
Subjects: | |
Online Access: | https://ppjp.ulm.ac.id/journal/index.php/f/article/view/6344 |
_version_ | 1818475875909238784 |
---|---|
author | Fitriani Fitriani Bualkar Abdullah Dahlang Tahir |
author_facet | Fitriani Fitriani Bualkar Abdullah Dahlang Tahir |
author_sort | Fitriani Fitriani |
collection | DOAJ |
description | Study of collimation field difference suitability test on analog and digital mammography at leading hospital in Makassar from 2016-2018. This test was carried out by using four coins of the same size placed on the edge of the collimator lamp field, one different size coin was placed on the table edge of the chest wall section, and one coin was placed on the surface under the compression pedal to coincide with the edge of the chest wall. The results of the p value obtained for each measurement point, both for the difference of the light collimation field toward the X-ray beam and the difference of X-ray beam collimation field toward image receptor have the same value of 0.000 (p < 0.05). This value states that there are significant differences in the collimation field difference from 2016-2018. The results of collimation field difference suitability test on analog and digital mammography are still within the tolerance limits recommended by BAPETEN, for the difference of the light collimation field toward X-ray beam ≤ 1% SID and the difference of X-ray beam collimation field toward image receptor ≤ 2% SID. This shows that mammography is feasible and able to operate safely for patients and non-patients. |
first_indexed | 2024-12-10T09:18:21Z |
format | Article |
id | doaj.art-3092e34581f34a21824c432bbf2b19ee |
institution | Directory Open Access Journal |
issn | 1829-796X 2541-1713 |
language | English |
last_indexed | 2024-12-10T09:18:21Z |
publishDate | 2019-08-01 |
publisher | Lambung Mangkurat University Press |
record_format | Article |
series | Jurnal Fisika Flux |
spelling | doaj.art-3092e34581f34a21824c432bbf2b19ee2022-12-22T01:54:46ZengLambung Mangkurat University PressJurnal Fisika Flux1829-796X2541-17132019-08-011629710210.20527/flux.v16i2.63445419Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan DigitalFitriani Fitriani0Bualkar Abdullah1Dahlang Tahir2Jurusan Fisika Universitas HasanuddinJurusan Fisika Universitas HasanuddinJurusan Fisika Universitas HasanuddinStudy of collimation field difference suitability test on analog and digital mammography at leading hospital in Makassar from 2016-2018. This test was carried out by using four coins of the same size placed on the edge of the collimator lamp field, one different size coin was placed on the table edge of the chest wall section, and one coin was placed on the surface under the compression pedal to coincide with the edge of the chest wall. The results of the p value obtained for each measurement point, both for the difference of the light collimation field toward the X-ray beam and the difference of X-ray beam collimation field toward image receptor have the same value of 0.000 (p < 0.05). This value states that there are significant differences in the collimation field difference from 2016-2018. The results of collimation field difference suitability test on analog and digital mammography are still within the tolerance limits recommended by BAPETEN, for the difference of the light collimation field toward X-ray beam ≤ 1% SID and the difference of X-ray beam collimation field toward image receptor ≤ 2% SID. This shows that mammography is feasible and able to operate safely for patients and non-patients.https://ppjp.ulm.ac.id/journal/index.php/f/article/view/6344analogcollimationdigitalmammographyx-ray. |
spellingShingle | Fitriani Fitriani Bualkar Abdullah Dahlang Tahir Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital Jurnal Fisika Flux analog collimation digital mammography x-ray. |
title | Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital |
title_full | Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital |
title_fullStr | Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital |
title_full_unstemmed | Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital |
title_short | Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital |
title_sort | studi uji kesesuaian selisih lapangan kolimasi pada pesawat mammografi analog dan digital |
topic | analog collimation digital mammography x-ray. |
url | https://ppjp.ulm.ac.id/journal/index.php/f/article/view/6344 |
work_keys_str_mv | AT fitrianifitriani studiujikesesuaianselisihlapangankolimasipadapesawatmammografianalogdandigital AT bualkarabdullah studiujikesesuaianselisihlapangankolimasipadapesawatmammografianalogdandigital AT dahlangtahir studiujikesesuaianselisihlapangankolimasipadapesawatmammografianalogdandigital |