Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
Hakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the proces...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Dove Medical Press
2015-01-01
|
Series: | Vascular Health and Risk Management |
Online Access: | http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM |
_version_ | 1818379826509119488 |
---|---|
author | Takahashi H Yoshika M Yokoi T |
author_facet | Takahashi H Yoshika M Yokoi T |
author_sort | Takahashi H |
collection | DOAJ |
description | Hakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the process of clinical diagnosis, and it is recognized as having great clinical utility. To measure BP as accurately as possible, innovative techniques have been incorporated into home BP measurement devices. Objective: The present study aimed to evaluate the performance of the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z), which are equipped with functions to detect irregular pulses and arm movement that lead to inaccurate BP readings. Methods: A team of three trained medical doctors validated the performance of these devices by comparing the data alternatively obtained from both devices with those from a standard mercury sphygmomanometer. Results: The magnitude of the difference in BP readings between the tested device and the standard mercury sphygmomanometer in the Omron BP765 and BP760N was within the range of ±3 mmHg (mean) allowed by the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2009 guidelines. Conclusion: The Omron BP765 and BP760N were found useful for the self-measurement of BP at home, and their performance fulfilled the requirement of the ANSI/AAMI/ISO 81060-2:2009 guidelines. Keywords: blood pressure, self-measurement, device, validation, ANSI/AAMI/ISO 81060-2:2009 |
first_indexed | 2024-12-14T02:08:57Z |
format | Article |
id | doaj.art-333ea81f7b524a29bd529861df6713b9 |
institution | Directory Open Access Journal |
issn | 1178-2048 |
language | English |
last_indexed | 2024-12-14T02:08:57Z |
publishDate | 2015-01-01 |
publisher | Dove Medical Press |
record_format | Article |
series | Vascular Health and Risk Management |
spelling | doaj.art-333ea81f7b524a29bd529861df6713b92022-12-21T23:20:49ZengDove Medical PressVascular Health and Risk Management1178-20482015-01-012015default495319896Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)Takahashi HYoshika MYokoi THakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the process of clinical diagnosis, and it is recognized as having great clinical utility. To measure BP as accurately as possible, innovative techniques have been incorporated into home BP measurement devices. Objective: The present study aimed to evaluate the performance of the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z), which are equipped with functions to detect irregular pulses and arm movement that lead to inaccurate BP readings. Methods: A team of three trained medical doctors validated the performance of these devices by comparing the data alternatively obtained from both devices with those from a standard mercury sphygmomanometer. Results: The magnitude of the difference in BP readings between the tested device and the standard mercury sphygmomanometer in the Omron BP765 and BP760N was within the range of ±3 mmHg (mean) allowed by the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2009 guidelines. Conclusion: The Omron BP765 and BP760N were found useful for the self-measurement of BP at home, and their performance fulfilled the requirement of the ANSI/AAMI/ISO 81060-2:2009 guidelines. Keywords: blood pressure, self-measurement, device, validation, ANSI/AAMI/ISO 81060-2:2009http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM |
spellingShingle | Takahashi H Yoshika M Yokoi T Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) Vascular Health and Risk Management |
title | Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) |
title_full | Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) |
title_fullStr | Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) |
title_full_unstemmed | Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) |
title_short | Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z) |
title_sort | validation of two automatic devices for the self measurement of blood pressure according to the ansi aami iso81060 2 2009 guidelines the omron bp765 hem 7311 zsa and the omron bp760n hem 7320 z |
url | http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM |
work_keys_str_mv | AT takahashih validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z AT yoshikam validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z AT yokoit validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z |