Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)

Hakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the proces...

Full description

Bibliographic Details
Main Authors: Takahashi H, Yoshika M, Yokoi T
Format: Article
Language:English
Published: Dove Medical Press 2015-01-01
Series:Vascular Health and Risk Management
Online Access:http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM
_version_ 1818379826509119488
author Takahashi H
Yoshika M
Yokoi T
author_facet Takahashi H
Yoshika M
Yokoi T
author_sort Takahashi H
collection DOAJ
description Hakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the process of clinical diagnosis, and it is recognized as having great clinical utility. To measure BP as accurately as possible, innovative techniques have been incorporated into home BP measurement devices. Objective: The present study aimed to evaluate the performance of the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z), which are equipped with functions to detect irregular pulses and arm movement that lead to inaccurate BP readings. Methods: A team of three trained medical doctors validated the performance of these devices by comparing the data alternatively obtained from both devices with those from a standard mercury sphygmomanometer. Results: The magnitude of the difference in BP readings between the tested device and the standard mercury sphygmomanometer in the Omron BP765 and BP760N was within the range of ±3 mmHg (mean) allowed by the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2009 guidelines. Conclusion: The Omron BP765 and BP760N were found useful for the self-measurement of BP at home, and their performance fulfilled the requirement of the ANSI/AAMI/ISO 81060-2:2009 guidelines. Keywords: blood pressure, self-measurement, device, validation, ANSI/AAMI/ISO 81060-2:2009
first_indexed 2024-12-14T02:08:57Z
format Article
id doaj.art-333ea81f7b524a29bd529861df6713b9
institution Directory Open Access Journal
issn 1178-2048
language English
last_indexed 2024-12-14T02:08:57Z
publishDate 2015-01-01
publisher Dove Medical Press
record_format Article
series Vascular Health and Risk Management
spelling doaj.art-333ea81f7b524a29bd529861df6713b92022-12-21T23:20:49ZengDove Medical PressVascular Health and Risk Management1178-20482015-01-012015default495319896Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)Takahashi HYoshika MYokoi THakuo Takahashi, Masamichi Yoshika, Toyohiko Yokoi Department of Clinical Sciences and Laboratory Medicine, Kansai Medical University, Hirakata Osaka, Japan Background: Allowing patients to measure their blood pressure (BP) at home will be the standard for evaluating the disease state as the process of clinical diagnosis, and it is recognized as having great clinical utility. To measure BP as accurately as possible, innovative techniques have been incorporated into home BP measurement devices. Objective: The present study aimed to evaluate the performance of the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z), which are equipped with functions to detect irregular pulses and arm movement that lead to inaccurate BP readings. Methods: A team of three trained medical doctors validated the performance of these devices by comparing the data alternatively obtained from both devices with those from a standard mercury sphygmomanometer. Results: The magnitude of the difference in BP readings between the tested device and the standard mercury sphygmomanometer in the Omron BP765 and BP760N was within the range of ±3 mmHg (mean) allowed by the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2009 guidelines. Conclusion: The Omron BP765 and BP760N were found useful for the self-measurement of BP at home, and their performance fulfilled the requirement of the ANSI/AAMI/ISO 81060-2:2009 guidelines. Keywords: blood pressure, self-measurement, device, validation, ANSI/AAMI/ISO 81060-2:2009http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM
spellingShingle Takahashi H
Yoshika M
Yokoi T
Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
Vascular Health and Risk Management
title Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
title_full Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
title_fullStr Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
title_full_unstemmed Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
title_short Validation of two automatic devices for the self-measurement of blood pressure according to the ANSI/AAMI/ISO81060-2:2009 guidelines: the Omron BP765 (HEM-7311-ZSA) and the Omron BP760N (HEM-7320-Z)
title_sort validation of two automatic devices for the self measurement of blood pressure according to the ansi aami iso81060 2 2009 guidelines the omron bp765 hem 7311 zsa and the omron bp760n hem 7320 z
url http://www.dovepress.com/validation-of-two-automatic-devices-for-the-self-measurement-of-blood--peer-reviewed-article-VHRM
work_keys_str_mv AT takahashih validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z
AT yoshikam validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z
AT yokoit validationoftwoautomaticdevicesfortheselfmeasurementofbloodpressureaccordingtotheansiaamiiso8106022009guidelinestheomronbp765hem7311zsaandtheomronbp760nhem7320z