Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report

Fracture and loosening of the implant can pose a serious impact as the fractured segment remaining inside the implant may prevent it from functioning efficiently. The harmony between the implant supported prosthesis and implant could be disrupted, leading to failure of the upper structure. Fistula f...

Full description

Bibliographic Details
Main Authors: Leblebicioglu Kurtulus Ikbal, Kuru Emel, Canakci Gulfesan
Format: Article
Language:English
Published: JCDR Research and Publications Private Limited 2018-03-01
Series:Journal of Clinical and Diagnostic Research
Subjects:
Online Access:https://jcdr.net/articles/PDF/11311/31092_CE[Ra1]_CE[EK]_F(SL)_PF1(EK_SL)_PFA(EK_SL)_PB(EK_SL)_PN(AP).pdf
_version_ 1818855636867219456
author Leblebicioglu Kurtulus Ikbal
Kuru Emel
Canakci Gulfesan
author_facet Leblebicioglu Kurtulus Ikbal
Kuru Emel
Canakci Gulfesan
author_sort Leblebicioglu Kurtulus Ikbal
collection DOAJ
description Fracture and loosening of the implant can pose a serious impact as the fractured segment remaining inside the implant may prevent it from functioning efficiently. The harmony between the implant supported prosthesis and implant could be disrupted, leading to failure of the upper structure. Fistula formation and soft tissue inflammation are frequently associated with loose abutment screws. This case report involved an evaluation of the screw threads of a loosened abutment compared with those of a fractured abutment removed from the same patient by Scanning Electron Microscopy (SEM). According to the results of the SEM analysis, deformities in the screw threads of the loosened abutment were evaluated, and whether a loosened abutment should be rescrewed was considered. In this case report an unused abutment screw, broken screw, and loosened abutment screw were analysed by SEM. In the loosened abutment screw, the thread was deformed in the concavities to a greater extent, than in the convexities. The metal structure appeared strained without any cracks. In the fractured abutment screw, the metal structure in the convexities appeared strained with small cracks present at grain boundaries. Crack growth appeared to be related to the high number of crack coalescence events. A significant difference was observed between screw threads with abutment loosening and screw threads without abutment loosening. Therefore, a loosened abutment should not be rescrewed because it increases the risk of abutment fracture.
first_indexed 2024-12-19T08:11:46Z
format Article
id doaj.art-33bfd31f48714a64b33de512091afbce
institution Directory Open Access Journal
issn 2249-782X
0973-709X
language English
last_indexed 2024-12-19T08:11:46Z
publishDate 2018-03-01
publisher JCDR Research and Publications Private Limited
record_format Article
series Journal of Clinical and Diagnostic Research
spelling doaj.art-33bfd31f48714a64b33de512091afbce2022-12-21T20:29:37ZengJCDR Research and Publications Private LimitedJournal of Clinical and Diagnostic Research2249-782X0973-709X2018-03-01123ZD04ZD0610.7860/JCDR/2018/31092.11311Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case ReportLeblebicioglu Kurtulus Ikbal0Kuru Emel1Canakci Gulfesan2Prosthodontist, Kayseri Nimet Bayraktar Oral and Dental Health Hospital, Kayseri, Turkey.Prosthodontist, Bitlis Oral and Dental Health Hospital, Bitlis, Turkey.Assistant Professor, Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Trakya University, Edirne, Turkey.Fracture and loosening of the implant can pose a serious impact as the fractured segment remaining inside the implant may prevent it from functioning efficiently. The harmony between the implant supported prosthesis and implant could be disrupted, leading to failure of the upper structure. Fistula formation and soft tissue inflammation are frequently associated with loose abutment screws. This case report involved an evaluation of the screw threads of a loosened abutment compared with those of a fractured abutment removed from the same patient by Scanning Electron Microscopy (SEM). According to the results of the SEM analysis, deformities in the screw threads of the loosened abutment were evaluated, and whether a loosened abutment should be rescrewed was considered. In this case report an unused abutment screw, broken screw, and loosened abutment screw were analysed by SEM. In the loosened abutment screw, the thread was deformed in the concavities to a greater extent, than in the convexities. The metal structure appeared strained without any cracks. In the fractured abutment screw, the metal structure in the convexities appeared strained with small cracks present at grain boundaries. Crack growth appeared to be related to the high number of crack coalescence events. A significant difference was observed between screw threads with abutment loosening and screw threads without abutment loosening. Therefore, a loosened abutment should not be rescrewed because it increases the risk of abutment fracture.https://jcdr.net/articles/PDF/11311/31092_CE[Ra1]_CE[EK]_F(SL)_PF1(EK_SL)_PFA(EK_SL)_PB(EK_SL)_PN(AP).pdfdental implantsfatiguefractography
spellingShingle Leblebicioglu Kurtulus Ikbal
Kuru Emel
Canakci Gulfesan
Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
Journal of Clinical and Diagnostic Research
dental implants
fatigue
fractography
title Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
title_full Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
title_fullStr Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
title_full_unstemmed Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
title_short Scanning Electron Microscopic Analysis of the Screw Threads of Loosened and Fractured Abutments: A Case Report
title_sort scanning electron microscopic analysis of the screw threads of loosened and fractured abutments a case report
topic dental implants
fatigue
fractography
url https://jcdr.net/articles/PDF/11311/31092_CE[Ra1]_CE[EK]_F(SL)_PF1(EK_SL)_PFA(EK_SL)_PB(EK_SL)_PN(AP).pdf
work_keys_str_mv AT leblebicioglukurtulusikbal scanningelectronmicroscopicanalysisofthescrewthreadsofloosenedandfracturedabutmentsacasereport
AT kuruemel scanningelectronmicroscopicanalysisofthescrewthreadsofloosenedandfracturedabutmentsacasereport
AT canakcigulfesan scanningelectronmicroscopicanalysisofthescrewthreadsofloosenedandfracturedabutmentsacasereport