Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study

We detail an extensive characterisation study on a previously described dual amperometric H2O2 biosensor consisting of H2O2 detection (blank) and degradation (catalase) electrodes. In vitro investigations demonstrated excellent H2O2 sensitivity and selectivity against the interferent, ascorbic acid....

Full description

Bibliographic Details
Main Authors: Caroline H. Reid, Niall J. Finnerty
Format: Article
Language:English
Published: MDPI AG 2017-07-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/17/7/1596
_version_ 1811185909512536064
author Caroline H. Reid
Niall J. Finnerty
author_facet Caroline H. Reid
Niall J. Finnerty
author_sort Caroline H. Reid
collection DOAJ
description We detail an extensive characterisation study on a previously described dual amperometric H2O2 biosensor consisting of H2O2 detection (blank) and degradation (catalase) electrodes. In vitro investigations demonstrated excellent H2O2 sensitivity and selectivity against the interferent, ascorbic acid. Ex vivo studies were performed to mimic physiological conditions prior to in vivo deployment. Exposure to brain tissue homogenate identified reliable sensitivity and selectivity recordings up to seven days for both blank and catalase electrodes. Furthermore, there was no compromise in pre- and post-implanted catalase electrode sensitivity in ex vivo mouse brain. In vivo investigations performed in anaesthetised mice confirmed the ability of the H2O2 biosensor to detect increases in amperometric current following locally perfused/infused H2O2 and antioxidant inhibitors mercaptosuccinic acid and sodium azide. Subsequent recordings in freely moving mice identified negligible effects of control saline and sodium ascorbate interference injections on amperometric H2O2 current. Furthermore, the stability of the amperometric current was confirmed over a five-day period and analysis of 24-h signal recordings identified the absence of diurnal variations in amperometric current. Collectively, these findings confirm the biosensor current responds in vivo to increasing exogenous and endogenous H2O2 and tentatively supports measurement of H2O2 dynamics in freely moving NOD SCID mice.
first_indexed 2024-04-11T13:38:27Z
format Article
id doaj.art-382fc3b7236e4491893d6dad1ee64d2d
institution Directory Open Access Journal
issn 1424-8220
language English
last_indexed 2024-04-11T13:38:27Z
publishDate 2017-07-01
publisher MDPI AG
record_format Article
series Sensors
spelling doaj.art-382fc3b7236e4491893d6dad1ee64d2d2022-12-22T04:21:22ZengMDPI AGSensors1424-82202017-07-01177159610.3390/s17071596s17071596Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation StudyCaroline H. Reid0Niall J. Finnerty1Chemistry Department, Maynooth University, Maynooth W23 F2H6, County Kildare, IrelandChemistry Department, Maynooth University, Maynooth W23 F2H6, County Kildare, IrelandWe detail an extensive characterisation study on a previously described dual amperometric H2O2 biosensor consisting of H2O2 detection (blank) and degradation (catalase) electrodes. In vitro investigations demonstrated excellent H2O2 sensitivity and selectivity against the interferent, ascorbic acid. Ex vivo studies were performed to mimic physiological conditions prior to in vivo deployment. Exposure to brain tissue homogenate identified reliable sensitivity and selectivity recordings up to seven days for both blank and catalase electrodes. Furthermore, there was no compromise in pre- and post-implanted catalase electrode sensitivity in ex vivo mouse brain. In vivo investigations performed in anaesthetised mice confirmed the ability of the H2O2 biosensor to detect increases in amperometric current following locally perfused/infused H2O2 and antioxidant inhibitors mercaptosuccinic acid and sodium azide. Subsequent recordings in freely moving mice identified negligible effects of control saline and sodium ascorbate interference injections on amperometric H2O2 current. Furthermore, the stability of the amperometric current was confirmed over a five-day period and analysis of 24-h signal recordings identified the absence of diurnal variations in amperometric current. Collectively, these findings confirm the biosensor current responds in vivo to increasing exogenous and endogenous H2O2 and tentatively supports measurement of H2O2 dynamics in freely moving NOD SCID mice.https://www.mdpi.com/1424-8220/17/7/1596amperometryH2O2characterisationimmunocompromisedin vivoextracellular
spellingShingle Caroline H. Reid
Niall J. Finnerty
Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
Sensors
amperometry
H2O2
characterisation
immunocompromised
in vivo
extracellular
title Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
title_full Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
title_fullStr Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
title_full_unstemmed Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
title_short Real-Time Amperometric Recording of Extracellular H2O2 in the Brain of Immunocompromised Mice: An In Vitro, Ex Vivo and In Vivo Characterisation Study
title_sort real time amperometric recording of extracellular h2o2 in the brain of immunocompromised mice an in vitro ex vivo and in vivo characterisation study
topic amperometry
H2O2
characterisation
immunocompromised
in vivo
extracellular
url https://www.mdpi.com/1424-8220/17/7/1596
work_keys_str_mv AT carolinehreid realtimeamperometricrecordingofextracellularh2o2inthebrainofimmunocompromisedmiceaninvitroexvivoandinvivocharacterisationstudy
AT nialljfinnerty realtimeamperometricrecordingofextracellularh2o2inthebrainofimmunocompromisedmiceaninvitroexvivoandinvivocharacterisationstudy