A Doherty Power Amplifier for Ultrasound Instrumentation

The ultrasound instrumentation uses linear power amplifiers with low power efficiency, generating unwanted heat and resulting in the deterioration of the echo signal quality of measured targets. Therefore, this study aims to develop a power amplifier scheme to increase power efficiency while maintai...

Full description

Bibliographic Details
Main Author: Hojong Choi
Format: Article
Language:English
Published: MDPI AG 2023-02-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/23/5/2406
_version_ 1827752187546042368
author Hojong Choi
author_facet Hojong Choi
author_sort Hojong Choi
collection DOAJ
description The ultrasound instrumentation uses linear power amplifiers with low power efficiency, generating unwanted heat and resulting in the deterioration of the echo signal quality of measured targets. Therefore, this study aims to develop a power amplifier scheme to increase power efficiency while maintaining appropriate echo signal quality. In communication systems, the Doherty power amplifier has shown relatively good power efficiency while producing high signal distortion. The same design scheme cannot be directly applied to ultrasound instrumentation. Therefore, the Doherty power amplifier needs to be re-designed. To verify the feasibility of the instrumentation, a Doherty power amplifier was designed to obtain high power efficiency. The measured gain, output 1-dB compression point, and power-added efficiency of the designed Doherty power amplifier were 33.71 dB, 35.71 dB<sub>m</sub>, and 57.24% at 25 MHz, respectively. In addition, the performance of the developed amplifier was measured and tested using the ultrasound transducer through the pulse-echo responses. The output power with 25 MHz, 5-cycle, and 43.06 dB<sub>m</sub> generated from the Doherty power amplifier was sent through the expander to the focused ultrasound transducer with 25 MHz and 0.5″ diameter. The detected signal was sent via a limiter. Afterwards, the signal was amplified by a 36.8 dB gain preamplifier, and then displayed in the oscilloscope. The measured peak-to-peak amplitude in the pulse-echo response with an ultrasound transducer was 0.9698 V. The data showed a comparable echo signal amplitude. Therefore, the designed Doherty power amplifier can improve the power efficiency used for medical ultrasound instrumentation.
first_indexed 2024-03-11T07:11:15Z
format Article
id doaj.art-3b90faf878f64bff92bd19c465fb5dbd
institution Directory Open Access Journal
issn 1424-8220
language English
last_indexed 2024-03-11T07:11:15Z
publishDate 2023-02-01
publisher MDPI AG
record_format Article
series Sensors
spelling doaj.art-3b90faf878f64bff92bd19c465fb5dbd2023-11-17T08:34:18ZengMDPI AGSensors1424-82202023-02-01235240610.3390/s23052406A Doherty Power Amplifier for Ultrasound InstrumentationHojong Choi0Department of Electronic Engineering, Gachon University, 1342 Seongnam-daero, Sujeong-gu, Seongnam 13120, Gyeonggi-do, Republic of KoreaThe ultrasound instrumentation uses linear power amplifiers with low power efficiency, generating unwanted heat and resulting in the deterioration of the echo signal quality of measured targets. Therefore, this study aims to develop a power amplifier scheme to increase power efficiency while maintaining appropriate echo signal quality. In communication systems, the Doherty power amplifier has shown relatively good power efficiency while producing high signal distortion. The same design scheme cannot be directly applied to ultrasound instrumentation. Therefore, the Doherty power amplifier needs to be re-designed. To verify the feasibility of the instrumentation, a Doherty power amplifier was designed to obtain high power efficiency. The measured gain, output 1-dB compression point, and power-added efficiency of the designed Doherty power amplifier were 33.71 dB, 35.71 dB<sub>m</sub>, and 57.24% at 25 MHz, respectively. In addition, the performance of the developed amplifier was measured and tested using the ultrasound transducer through the pulse-echo responses. The output power with 25 MHz, 5-cycle, and 43.06 dB<sub>m</sub> generated from the Doherty power amplifier was sent through the expander to the focused ultrasound transducer with 25 MHz and 0.5″ diameter. The detected signal was sent via a limiter. Afterwards, the signal was amplified by a 36.8 dB gain preamplifier, and then displayed in the oscilloscope. The measured peak-to-peak amplitude in the pulse-echo response with an ultrasound transducer was 0.9698 V. The data showed a comparable echo signal amplitude. Therefore, the designed Doherty power amplifier can improve the power efficiency used for medical ultrasound instrumentation.https://www.mdpi.com/1424-8220/23/5/2406Doherty power amplifierultrasound instrumentationultrasound transducer
spellingShingle Hojong Choi
A Doherty Power Amplifier for Ultrasound Instrumentation
Sensors
Doherty power amplifier
ultrasound instrumentation
ultrasound transducer
title A Doherty Power Amplifier for Ultrasound Instrumentation
title_full A Doherty Power Amplifier for Ultrasound Instrumentation
title_fullStr A Doherty Power Amplifier for Ultrasound Instrumentation
title_full_unstemmed A Doherty Power Amplifier for Ultrasound Instrumentation
title_short A Doherty Power Amplifier for Ultrasound Instrumentation
title_sort doherty power amplifier for ultrasound instrumentation
topic Doherty power amplifier
ultrasound instrumentation
ultrasound transducer
url https://www.mdpi.com/1424-8220/23/5/2406
work_keys_str_mv AT hojongchoi adohertypoweramplifierforultrasoundinstrumentation
AT hojongchoi dohertypoweramplifierforultrasoundinstrumentation