THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA

Objectives In this research, we aimed to develop a thermoelectric device (TED) for intracavitary hypothermia, to carry out experimental studies into its functional characteristics, as well as to develop treatment procedures using this device.Method A TED design for intracavitary hypothermia was deve...

Full description

Bibliographic Details
Main Authors: T. A. Ismailov, O. V. Evdulov, M. A. Khazamova, T. A. Ragimova, R. A.-M. Magomadov
Format: Article
Language:Russian
Published: Dagestan State Technical University 2019-11-01
Series:Вестник Дагестанского государственного технического университета: Технические науки
Subjects:
Online Access:https://vestnik.dgtu.ru/jour/article/view/689
_version_ 1797697808461987840
author T. A. Ismailov
O. V. Evdulov
M. A. Khazamova
T. A. Ragimova
R. A.-M. Magomadov
author_facet T. A. Ismailov
O. V. Evdulov
M. A. Khazamova
T. A. Ragimova
R. A.-M. Magomadov
author_sort T. A. Ismailov
collection DOAJ
description Objectives In this research, we aimed to develop a thermoelectric device (TED) for intracavitary hypothermia, to carry out experimental studies into its functional characteristics, as well as to develop treatment procedures using this device.Method A TED design for intracavitary hypothermia was developed, consisting of two thermoelectric batteries (TEBs) interconnected using an allmetal heat pipeline equipped with an influencing applicator and a liquid heat exchanger. Experimental studies of the device were carried out on a designed bench, which included a climatic chamber, a source of electrical energy and a temperature meter connected to a PC. Results Temperature changes at the control points of the device were recorded over time. Changes in the time taken to stabilise temperatures of the device tip, heat pipe and heat sink system at different TEB supply currents were investigated. Possible methods for treating some ENT diseases using the developed device are described.Conclusion It is shown that the temperature of the applicator is stabilised after approximately 4-4.5 minutes without load. At the same time, an increase in the current of the additional TEB from 5 to 12 A at the main TEB current of 23 A reduces the temperature from 237 K to 224 K. In the process of carrying out the procedures, the necessary level of a decrease in the temperature of a biological object can be reached at the supply current of the main and additional TEBs equal to 23 and 12 A, respectively, after 2.5 minutes. The full potential of the TED device will be realised through the developed of methods for treating ENT diseases.
first_indexed 2024-03-12T03:45:22Z
format Article
id doaj.art-3e8e893aeeb14b09ace2766a65105c59
institution Directory Open Access Journal
issn 2073-6185
2542-095X
language Russian
last_indexed 2024-03-12T03:45:22Z
publishDate 2019-11-01
publisher Dagestan State Technical University
record_format Article
series Вестник Дагестанского государственного технического университета: Технические науки
spelling doaj.art-3e8e893aeeb14b09ace2766a65105c592023-09-03T12:57:59ZrusDagestan State Technical UniversityВестник Дагестанского государственного технического университета: Технические науки2073-61852542-095X2019-11-01463203110.21822/2073-6185-2019-46-3-20-31522THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIAT. A. Ismailov0O. V. Evdulov1M. A. Khazamova2T. A. Ragimova3R. A.-M. Magomadov4Дагестанский государственный технический университетДагестанский государственный технический университетДагестанский государственный технический университетДагестанский государственный технический университетГрозненский государственный нефтяной технический университетObjectives In this research, we aimed to develop a thermoelectric device (TED) for intracavitary hypothermia, to carry out experimental studies into its functional characteristics, as well as to develop treatment procedures using this device.Method A TED design for intracavitary hypothermia was developed, consisting of two thermoelectric batteries (TEBs) interconnected using an allmetal heat pipeline equipped with an influencing applicator and a liquid heat exchanger. Experimental studies of the device were carried out on a designed bench, which included a climatic chamber, a source of electrical energy and a temperature meter connected to a PC. Results Temperature changes at the control points of the device were recorded over time. Changes in the time taken to stabilise temperatures of the device tip, heat pipe and heat sink system at different TEB supply currents were investigated. Possible methods for treating some ENT diseases using the developed device are described.Conclusion It is shown that the temperature of the applicator is stabilised after approximately 4-4.5 minutes without load. At the same time, an increase in the current of the additional TEB from 5 to 12 A at the main TEB current of 23 A reduces the temperature from 237 K to 224 K. In the process of carrying out the procedures, the necessary level of a decrease in the temperature of a biological object can be reached at the supply current of the main and additional TEBs equal to 23 and 12 A, respectively, after 2.5 minutes. The full potential of the TED device will be realised through the developed of methods for treating ENT diseases.https://vestnik.dgtu.ru/jour/article/view/689гипотермиявнутренняя полостьтермоэлектрическое устройствотермоэлектрическая батареяохлаждающее воздействиетеплообменэкспериментальный стенднатурные испытанияизмерениелечебные процедуры
spellingShingle T. A. Ismailov
O. V. Evdulov
M. A. Khazamova
T. A. Ragimova
R. A.-M. Magomadov
THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
Вестник Дагестанского государственного технического университета: Технические науки
гипотермия
внутренняя полость
термоэлектрическое устройство
термоэлектрическая батарея
охлаждающее воздействие
теплообмен
экспериментальный стенд
натурные испытания
измерение
лечебные процедуры
title THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
title_full THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
title_fullStr THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
title_full_unstemmed THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
title_short THERMOELECTRIC DEVICE FOR INTRACAVITARY HYPOTHERMIA
title_sort thermoelectric device for intracavitary hypothermia
topic гипотермия
внутренняя полость
термоэлектрическое устройство
термоэлектрическая батарея
охлаждающее воздействие
теплообмен
экспериментальный стенд
натурные испытания
измерение
лечебные процедуры
url https://vestnik.dgtu.ru/jour/article/view/689
work_keys_str_mv AT taismailov thermoelectricdeviceforintracavitaryhypothermia
AT ovevdulov thermoelectricdeviceforintracavitaryhypothermia
AT makhazamova thermoelectricdeviceforintracavitaryhypothermia
AT taragimova thermoelectricdeviceforintracavitaryhypothermia
AT rammagomadov thermoelectricdeviceforintracavitaryhypothermia