Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
<p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combina...
Format: | Article |
---|---|
Language: | English |
Published: |
Ivyspring International Publisher
2004-06-01
|
Series: | International Journal of Medical Sciences |
Subjects: | |
Online Access: | http://www.medsci.org/v01p0063.htm |
_version_ | 1818473533791010816 |
---|---|
collection | DOAJ |
description | <p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R<sub>50</sub>. We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs.</p> |
first_indexed | 2024-04-14T04:25:03Z |
format | Article |
id | doaj.art-40258da57eb44445a87401a425c73d1d |
institution | Directory Open Access Journal |
issn | 1449-1907 |
language | English |
last_indexed | 2024-04-14T04:25:03Z |
publishDate | 2004-06-01 |
publisher | Ivyspring International Publisher |
record_format | Article |
series | International Journal of Medical Sciences |
spelling | doaj.art-40258da57eb44445a87401a425c73d1d2022-12-22T02:12:22ZengIvyspring International PublisherInternational Journal of Medical Sciences1449-19072004-06-01126375Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software<p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R<sub>50</sub>. We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs.</p>http://www.medsci.org/v01p0063.htmMonte Carloelectron beam commissioningNXEGS |
spellingShingle | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software International Journal of Medical Sciences Monte Carlo electron beam commissioning NXEGS |
title | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software |
title_full | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software |
title_fullStr | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software |
title_full_unstemmed | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software |
title_short | Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software |
title_sort | monte carlo commissioning of low energy electron radiotherapy beams using nxegs software |
topic | Monte Carlo electron beam commissioning NXEGS |
url | http://www.medsci.org/v01p0063.htm |