Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software

<p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combina...

Full description

Bibliographic Details
Format: Article
Language:English
Published: Ivyspring International Publisher 2004-06-01
Series:International Journal of Medical Sciences
Subjects:
Online Access:http://www.medsci.org/v01p0063.htm
_version_ 1818473533791010816
collection DOAJ
description <p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R<sub>50</sub>. We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs.</p>
first_indexed 2024-04-14T04:25:03Z
format Article
id doaj.art-40258da57eb44445a87401a425c73d1d
institution Directory Open Access Journal
issn 1449-1907
language English
last_indexed 2024-04-14T04:25:03Z
publishDate 2004-06-01
publisher Ivyspring International Publisher
record_format Article
series International Journal of Medical Sciences
spelling doaj.art-40258da57eb44445a87401a425c73d1d2022-12-22T02:12:22ZengIvyspring International PublisherInternational Journal of Medical Sciences1449-19072004-06-01126375Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software<p>This work is a report on the commissioning of low energy electron beams of a medical linear accelerator for Monte Carlo dose calculation using NXEGS software (NXEGS version 1.0.10.0, NX Medical Software, LLC). A unique feature of NXEGS is automated commissioning, a process whereby a combination of analytic and Monte Carlo methods generates beam models from dosimetric data collected in a water phantom. This study uses NXEGS to commission 6, 9, and 12 MeV electron beams of a Varian Clinac 2100C using three applicators with standard inserts. Central axis depth-dose, primary axis and diagonal beam profiles, and output factors are the measurements necessary for commissioning of the code. We present a comparison of measured dose distributions with the distributions generated by NXEGS, using confidence limits on seven measures of error. We find that confidence limits are typically less than 3% or 3 mm, but increase with increasing source to surface distance (SSD) and depth at or beyond R<sub>50</sub>. We also investigate the dependence of NXEGS' performance on the size and composition of data used to commission the program, finding a weak dependence on number of dose profiles in the data set, but finding also that commissioning data need be measured at only two SSDs.</p>http://www.medsci.org/v01p0063.htmMonte Carloelectron beam commissioningNXEGS
spellingShingle Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
International Journal of Medical Sciences
Monte Carlo
electron beam commissioning
NXEGS
title Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_full Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_fullStr Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_full_unstemmed Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_short Monte Carlo Commissioning of Low Energy Electron Radiotherapy Beams using NXEGS Software
title_sort monte carlo commissioning of low energy electron radiotherapy beams using nxegs software
topic Monte Carlo
electron beam commissioning
NXEGS
url http://www.medsci.org/v01p0063.htm