Piezoresistive microcantilever aptasensor for ricin detection and kinetic analysis

Up to now, there has been no report on target molecules detection by a piezoresistive microcantilever aptasensor. In order to evaluate the test performance and investigate the response dynamic characteristics of a piezoresistive microcantilever aptasensor, a novel method for ricin detection and kine...

Full description

Bibliographic Details
Main Authors: Zhi-Wei Liu, Zhao-Yang Tong, Bing Liu, Lan-Qun Hao, Xi-Hui Mu, Jin-Ping Zhang, Chuan Gao
Format: Article
Language:English
Published: AIP Publishing LLC 2015-04-01
Series:AIP Advances
Online Access:http://dx.doi.org/10.1063/1.4907996
Description
Summary:Up to now, there has been no report on target molecules detection by a piezoresistive microcantilever aptasensor. In order to evaluate the test performance and investigate the response dynamic characteristics of a piezoresistive microcantilever aptasensor, a novel method for ricin detection and kinetic analysis based on a piezoresistive microcantilever aptasensor was proposed, where ricin aptamer was immobilised on the microcantilever surface by biotin-avidin binding system. Results showed that the detection limit of ricin was 0.04μg L−1 (S/N ≥ 3). A linear relationship between the response voltage and the concentration of ricin in the range of 0.2μg L−1-40μg L−1 was obtained, with the linear regression equation of ΔUe = 0.904C + 5.852 (n = 5, R = 0.991, p < 0.001). The sensor showed no response for abrin, BSA, and could overcome the influence of complex environmental disruptors, indicating high specificity and good selectivity. Recovery and reproducibility in the result of simulated samples (simulated water, soil, and flour sample) determination met the analysis requirements, which was 90.5∼95.5% and 7.85%∼9.39%, respectively. On this basis, a reaction kinetic model based on ligand-receptor binding and the relationship with response voltage was established. The model could well reflect the dynamic response of the sensor. The correlation coefficient (R) was greater than or equal to 0.9456 (p < 0.001). Response voltage (ΔUe) and response time (t0) obtained from the fitting equation on different concentrations of ricin fitted well with the measured values.
ISSN:2158-3226