Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure

Portable X-ray fluorescence (p-XRF) devices are commonly utilized to analyze the chemical composition of various materials, such as archaeological pottery and siliceous substances. The discussion regarding the suitability of this method for such samples is ongoing, as the data are often said to be q...

Full description

Bibliographic Details
Main Author: Michaela Schauer
Format: Article
Language:English
Published: Elsevier 2024-04-01
Series:Data in Brief
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2352340923009538
_version_ 1827313721569968128
author Michaela Schauer
author_facet Michaela Schauer
author_sort Michaela Schauer
collection DOAJ
description Portable X-ray fluorescence (p-XRF) devices are commonly utilized to analyze the chemical composition of various materials, such as archaeological pottery and siliceous substances. The discussion regarding the suitability of this method for such samples is ongoing, as the data are often said to be quantitatively unreliable. Nevertheless, the development of coefficient corrections (coefcors) offers a means to transparently demonstrate the quality and comparability of p-XRF data. These coefcors are established by comparing p-XRF measurements with laboratory methods derived from the same sample set. The Frankfurt Keramiklabor (Frankfurt Procedure) procedure serves as the basis for the Munich Procedure, which outlines criteria for assessing the quality of linear regressions applied to p-XRF data. The Munich Procedure provides clear benchmarks for the coefficient of determination (r²) and the relative standard error of the estimate (rSEE). Additionally, it includes a robustness test using a bootstrap method, as detailed in the R-scripts of this article. The Munich Procedure is applied to datasets generated using the Niton XL3t No. 97390 instrument used by the Dept. of Culture Sciences, Ludwig-Maximilians-Universität München (LMU), between 2017 and 2023. Measurements were conducted in the TestAllGeo mode, using an 8 mm collimator and a 300-second measurement time in air at Goethe-Universität Frankfurt. Data processing took place at LMU. Four datasets were utilized to develop four coefcors for comparing p-XRF and WD-XRF values of the Frankfurt pottery standard set (coefcor I-IV). Additionally, five coefcors were created to adjust p-XRF data from the same instrument before and after an instrument reset (coefcor ItoII, ItoIII, IItoIII, IIItoII, IVtoIII). The instrument provides analytical data, which is used to generate the dataset for developing coefficients using R-scripts. Processed data, including calculated criteria and factors (slope and intercept) for each chemical element, are presented with relevant graphical output. These coefcor factors can be utilized to correct empirical data obtained with Niton No. 97390 by selecting the appropriate coefcor based on the measurement date. This process is also outlined in the R-scripts. To assist users in applying the Munich Procedure to their own data, an R-script is provided that offers a more detailed explanation of the calculations. While the coefcors given here are specific to the instrument and therefore provide a basis for understanding the data quality of (as yet unpublished) studies using the Niton No. 97390, the method itself can be universally applied to develop coefcors for other p-XRF instruments.
first_indexed 2024-03-07T23:38:12Z
format Article
id doaj.art-4314b5ea29e843a8bb0486feeb85790e
institution Directory Open Access Journal
issn 2352-3409
language English
last_indexed 2024-04-24T22:20:10Z
publishDate 2024-04-01
publisher Elsevier
record_format Article
series Data in Brief
spelling doaj.art-4314b5ea29e843a8bb0486feeb85790e2024-03-20T06:09:31ZengElsevierData in Brief2352-34092024-04-0153109914Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedureMichaela Schauer0Department für Kulturwissenschaften, Geschwister-Scholl-Platz 1, Ludwig-Maximilians-Universität München, München 80539, Germany; Universität Wien, Franz-Klein-Gasse 1, Vienna Institute for Archaeological Science (VIAS), Wien 1190, Austria; Human Evolution and Archaeological Sciences (HEAS), Djerassiplatz 1, Universität Wien, Wien 1030, Austria; Correspondence to: Universität Wien, Franz-Klein-Gasse 1, Vienna Institute for Archaeological Science (VIAS), Wien 1190, Austria.Portable X-ray fluorescence (p-XRF) devices are commonly utilized to analyze the chemical composition of various materials, such as archaeological pottery and siliceous substances. The discussion regarding the suitability of this method for such samples is ongoing, as the data are often said to be quantitatively unreliable. Nevertheless, the development of coefficient corrections (coefcors) offers a means to transparently demonstrate the quality and comparability of p-XRF data. These coefcors are established by comparing p-XRF measurements with laboratory methods derived from the same sample set. The Frankfurt Keramiklabor (Frankfurt Procedure) procedure serves as the basis for the Munich Procedure, which outlines criteria for assessing the quality of linear regressions applied to p-XRF data. The Munich Procedure provides clear benchmarks for the coefficient of determination (r²) and the relative standard error of the estimate (rSEE). Additionally, it includes a robustness test using a bootstrap method, as detailed in the R-scripts of this article. The Munich Procedure is applied to datasets generated using the Niton XL3t No. 97390 instrument used by the Dept. of Culture Sciences, Ludwig-Maximilians-Universität München (LMU), between 2017 and 2023. Measurements were conducted in the TestAllGeo mode, using an 8 mm collimator and a 300-second measurement time in air at Goethe-Universität Frankfurt. Data processing took place at LMU. Four datasets were utilized to develop four coefcors for comparing p-XRF and WD-XRF values of the Frankfurt pottery standard set (coefcor I-IV). Additionally, five coefcors were created to adjust p-XRF data from the same instrument before and after an instrument reset (coefcor ItoII, ItoIII, IItoIII, IIItoII, IVtoIII). The instrument provides analytical data, which is used to generate the dataset for developing coefficients using R-scripts. Processed data, including calculated criteria and factors (slope and intercept) for each chemical element, are presented with relevant graphical output. These coefcor factors can be utilized to correct empirical data obtained with Niton No. 97390 by selecting the appropriate coefcor based on the measurement date. This process is also outlined in the R-scripts. To assist users in applying the Munich Procedure to their own data, an R-script is provided that offers a more detailed explanation of the calculations. While the coefcors given here are specific to the instrument and therefore provide a basis for understanding the data quality of (as yet unpublished) studies using the Niton No. 97390, the method itself can be universally applied to develop coefcors for other p-XRF instruments.http://www.sciencedirect.com/science/article/pii/S2352340923009538p-XRFTransparencyQuality controlArchaeologyPotteryCorrection factor
spellingShingle Michaela Schauer
Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
Data in Brief
p-XRF
Transparency
Quality control
Archaeology
Pottery
Correction factor
title Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
title_full Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
title_fullStr Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
title_full_unstemmed Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
title_short Coefficient corrections for portable X-ray fluorescence data of the Niton XL3t No. 97390 (coefcor I-IV) developed according to the Munich procedure
title_sort coefficient corrections for portable x ray fluorescence data of the niton xl3t no 97390 coefcor i iv developed according to the munich procedure
topic p-XRF
Transparency
Quality control
Archaeology
Pottery
Correction factor
url http://www.sciencedirect.com/science/article/pii/S2352340923009538
work_keys_str_mv AT michaelaschauer coefficientcorrectionsforportablexrayfluorescencedataofthenitonxl3tno97390coefcoriivdevelopedaccordingtothemunichprocedure