Near-infrared metalens for high-resolution and deep focus optical coherence tomography

High-resolution endoscopic optical imaging is a crucial technique in biological imaging to examine the inside organs. There is a trade-off between lateral resolution and depth of focus in such applications. Traditional Optical Coherence Tomography provides an increased depth range but falls short of...

Full description

Bibliographic Details
Main Authors: Mohammad Haerinia, Bowen Zheng, Aaron Hutchins, Hong Tang, Hang Li, Yunxi Dong, Wei Guo, Hualiang Zhang
Format: Article
Language:English
Published: AIP Publishing LLC 2022-12-01
Series:AIP Advances
Online Access:http://dx.doi.org/10.1063/5.0131170
Description
Summary:High-resolution endoscopic optical imaging is a crucial technique in biological imaging to examine the inside organs. There is a trade-off between lateral resolution and depth of focus in such applications. Traditional Optical Coherence Tomography provides an increased depth range but falls short of desired resolution. The combination of both higher resolution and larger imaging depth of focus of metalens can improve the clinical utility of endoscopic optical imaging. In this work, we designed, analyzed, and fabricated a 500 µm diameter metalens operating at 1300 nm to achieve high resolution and large imaging depth of focus, therefore, addressing this need. The full width at half maximum and depth of focus for the proposed metalens are 3.10 and 286 µm, respectively.
ISSN:2158-3226