Summary: | Several manufacturers have developed systems that automatically measure the amount of blood in culture bottles. We compared the volumes measured automatically by the Virtuo instrument (bioMérieux, France) (height-based volumes) and those calculated by weighing the bottles. In all, 150 pairs of aerobic and anaerobic blood culture bottles (BacT/ALERT FA/FN Plus, bioMérieux) were randomly selected over two periods to compare the height- and weight-based volumes and analyze the effect of foam. We also estimated the limit of detection (LOD) and the cut-off value for 5 mL equine blood. The mean height-based volume was approximately 0.67 mL greater than the weight-based volume, particularly in anaerobic culture bottles. Foam did not have a significant effect. The LOD for the automatic height-based volume of equine blood was 0.2–0.4 mL. The 5 mL cut-off was 4–4.2 mL. Therefore, when reporting or monitoring blood volume within culture bottles in the laboratory, these performance characteristics should be adequately considered.
|