Mechanical Behaviour and Primary Stability of a Self-Condensing Implant: A Laboratory Critical Simulation of a Severe Maxillary Atrophy on Polyurethane Lamina

Background: Posterior maxillary atrophies could emerge after the loss of teeth, trauma, infections, or lesions that often require regenerative approaches. In these critical conditions, the achievement of implant primary stability represents a clinical challenge in the operative practice. Therefore,...

Full description

Bibliographic Details
Main Authors: Luca Comuzzi, Giovanna Iezzi, Alessandra Lucchese, Natalia Di Pietro, Pierluigi Balice, Camillo D’Arcangelo, Adriano Piattelli, Margherita Tumedei
Format: Article
Language:English
Published: MDPI AG 2022-01-01
Series:Applied Sciences
Subjects:
Online Access:https://www.mdpi.com/2076-3417/12/3/966
Description
Summary:Background: Posterior maxillary atrophies could emerge after the loss of teeth, trauma, infections, or lesions that often require regenerative approaches. In these critical conditions, the achievement of implant primary stability represents a clinical challenge in the operative practice. Therefore, a two-stage approach is often preferred with a delay of the rehabilitation time and a consistent increasing of the biological and the operative costs. The aim of this study was to evaluate the mechanical behaviour of a self-condenser implant compared to a standard implant in a critical simulation on different thicknesses and densities of polyurethane lamina. Materials and methods: A total of two implant models were tested: a self-condensing device (test) and a standard implant (control). The study evaluated the insertion torque and the pull-out strength values of the test and control implants inserted in different sizes (1, 2, and 3 mm) and density polyurethane lamina (10, 20, and 30 pcf) for a total of 320 experimental sites. Results: In total, 320 experimental sites were produced in the polyurethane samples. A statistically significant difference of insertion and pull-out torque values between the test and control Implants was found in the different bone densities (<i>p</i> < 0.05). The insertion and pull-out torque values were always higher for the test implants in all experimental conditions. In all bone densities, the insertion torque values were higher than the pull-out torque values. The self-condenser dental implant design evaluated in this in vitro study showed a high level of stability in all experimental conditions. Conclusions: The test implant could represent a useful tool for a one-stage surgical approach in the presence of limited residual native bone as an alternative to a delayed technique.
ISSN:2076-3417