Evaluation of Bio-Rad D-10® and arkray adams HA-8160® HPLC analyzers in HbA1c measurement

Glycosylated hemoglobin A1c (HbA1c) is an important parameter used for the assessment of time-dependent glycemic status and the diagnosis and follow-up of diabetes. This study aimed to evaluate the analytical performance of two HPLC analyzers, the Bio-Rad D-10®, and the Arkray Adams HA-8160®. Accura...

Full description

Bibliographic Details
Main Authors: Ozgur Mehmet Yis, Mine Busra Pehlivan, Guler Bugdayci, Neslihan Yuce
Format: Article
Language:English
Published: Society of Turaz Bilim 2019-12-01
Series:Medicine Science
Subjects:
Online Access:http://www.ejmanager.com/fulltextpdf.php?mno=41188
Description
Summary:Glycosylated hemoglobin A1c (HbA1c) is an important parameter used for the assessment of time-dependent glycemic status and the diagnosis and follow-up of diabetes. This study aimed to evaluate the analytical performance of two HPLC analyzers, the Bio-Rad D-10®, and the Arkray Adams HA-8160®. Accuracy and imprecision studies were conducted, and a method comparison study was performed with 105 samples. Samples were collected on five consecutive days and measured on both analyzers within two hours. Bland-Altman and regression analyses were used for statistical evaluation of the data. Low and high-level within-day CV values were calculated as 1.22% and 0.60% for the Arkray analyzer and 1.2% and 0.30% for the Bio-Rad analyzer, respectively. They were calculated as 1.27% and 2.52% for the Arkray analyzer and 2.32% and 3.44% for the Bio-Rad analyzer, respectively, between days. The within-day CV values for both analyzers were below the limit of 2.5% specified by the International Federation of Clinical Chemistry (IFCC). The bias average for three months was 3.2% for the Arkray analyzer and 1.2% for the Bio-Rad analyzer. The Spearman correlation coefficient (r) was 0.973 (p [Med-Science 2019; 8(4.000): 953-6]
ISSN:2147-0634