Safe exposure distances for transcranial magnetic stimulation based on computer simulations

The results of a computer simulation examining the compliance of a given transcranial magnetic stimulation device to the 2010 International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines are presented. The objective was to update the safe distance estimates with the most current...

Full description

Bibliographic Details
Main Authors: Iam Palatnik de Sousa, Carlos R. H. Barbosa, Elisabeth Costa Monteiro
Format: Article
Language:English
Published: PeerJ Inc. 2018-06-01
Series:PeerJ
Subjects:
Online Access:https://peerj.com/articles/5034.pdf
_version_ 1827606848913539072
author Iam Palatnik de Sousa
Carlos R. H. Barbosa
Elisabeth Costa Monteiro
author_facet Iam Palatnik de Sousa
Carlos R. H. Barbosa
Elisabeth Costa Monteiro
author_sort Iam Palatnik de Sousa
collection DOAJ
description The results of a computer simulation examining the compliance of a given transcranial magnetic stimulation device to the 2010 International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines are presented. The objective was to update the safe distance estimates with the most current safety guidelines, as well as comparing these to values reported in previous publications. The 3D data generated was compared against results available in the literature, regarding the MCB-70 coil by Medtronic. Regarding occupational exposure, safe distances of 1.46 m and 0.96 m are derived from the simulation according to the 2003 and 2010 ICNIRP guidelines, respectively. These values are then compared to safe distances previously reported in other studies.
first_indexed 2024-03-09T06:46:15Z
format Article
id doaj.art-523fa05ad78a4f15ae633ce00dbefb6a
institution Directory Open Access Journal
issn 2167-8359
language English
last_indexed 2024-03-09T06:46:15Z
publishDate 2018-06-01
publisher PeerJ Inc.
record_format Article
series PeerJ
spelling doaj.art-523fa05ad78a4f15ae633ce00dbefb6a2023-12-03T10:35:23ZengPeerJ Inc.PeerJ2167-83592018-06-016e503410.7717/peerj.5034Safe exposure distances for transcranial magnetic stimulation based on computer simulationsIam Palatnik de SousaCarlos R. H. BarbosaElisabeth Costa MonteiroThe results of a computer simulation examining the compliance of a given transcranial magnetic stimulation device to the 2010 International Commission on Non-Ionizing Radiation Protection (ICNIRP) guidelines are presented. The objective was to update the safe distance estimates with the most current safety guidelines, as well as comparing these to values reported in previous publications. The 3D data generated was compared against results available in the literature, regarding the MCB-70 coil by Medtronic. Regarding occupational exposure, safe distances of 1.46 m and 0.96 m are derived from the simulation according to the 2003 and 2010 ICNIRP guidelines, respectively. These values are then compared to safe distances previously reported in other studies.https://peerj.com/articles/5034.pdfTranscranial magnetic stimulationSafetySpatial distributionExposure limitsMetrology
spellingShingle Iam Palatnik de Sousa
Carlos R. H. Barbosa
Elisabeth Costa Monteiro
Safe exposure distances for transcranial magnetic stimulation based on computer simulations
PeerJ
Transcranial magnetic stimulation
Safety
Spatial distribution
Exposure limits
Metrology
title Safe exposure distances for transcranial magnetic stimulation based on computer simulations
title_full Safe exposure distances for transcranial magnetic stimulation based on computer simulations
title_fullStr Safe exposure distances for transcranial magnetic stimulation based on computer simulations
title_full_unstemmed Safe exposure distances for transcranial magnetic stimulation based on computer simulations
title_short Safe exposure distances for transcranial magnetic stimulation based on computer simulations
title_sort safe exposure distances for transcranial magnetic stimulation based on computer simulations
topic Transcranial magnetic stimulation
Safety
Spatial distribution
Exposure limits
Metrology
url https://peerj.com/articles/5034.pdf
work_keys_str_mv AT iampalatnikdesousa safeexposuredistancesfortranscranialmagneticstimulationbasedoncomputersimulations
AT carlosrhbarbosa safeexposuredistancesfortranscranialmagneticstimulationbasedoncomputersimulations
AT elisabethcostamonteiro safeexposuredistancesfortranscranialmagneticstimulationbasedoncomputersimulations