Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission.
<h4>Background</h4>To investigate the effects of B1-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI).<h4>Methods</h4>...
Main Authors: | , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2022-01-01
|
Series: | PLoS ONE |
Online Access: | https://doi.org/10.1371/journal.pone.0270689 |
_version_ | 1818193406107582464 |
---|---|
author | Stefan Herz Maria R Stefanescu David Lohr Patrick Vogel Aleksander Kosmala Maxim Terekhov Andreas M Weng Jan-Peter Grunz Thorsten A Bley Laura M Schreiber |
author_facet | Stefan Herz Maria R Stefanescu David Lohr Patrick Vogel Aleksander Kosmala Maxim Terekhov Andreas M Weng Jan-Peter Grunz Thorsten A Bley Laura M Schreiber |
author_sort | Stefan Herz |
collection | DOAJ |
description | <h4>Background</h4>To investigate the effects of B1-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI).<h4>Methods</h4>Stenosis phantoms with different grades of stenosis (0%, 20%, 40%, 60%, 80%, and 100%; 5 mm inner vessel diameter) were produced using 3D printing (clear resin). Phantoms were imaged with four different concentrations of diluted Gd-DOTA representing established arterial concentrations after intravenous injection in humans. Samples were centrally positioned in a thorax phantom of 30 cm diameter filled with a custom-made liquid featuring dielectric properties of muscle tissue. MRI was performed on a 7 T whole-body system. 2D-gradient-echo sequences were acquired with an 8-channel transmit 16-channel receive (8 Tx / 16 Rx) cardiac array prototype coil with and without pTX mode. Measurements were compared to those obtained with identical scan parameters using a commercially available 1 Tx / 16 Rx single transmit coil (sTX). To assess reproducibility, measurements (n = 15) were repeated at different horizontal angles with respect to the B0-field.<h4>Results</h4>B1-shimming and pTX markedly improved flip angle homogeneity across the thorax phantom yielding a distinctly increased signal-to-noise ratio (SNR) averaged over a whole slice relative to non-manipulated RF fields. Images without B1-shimming showed shading artifacts due to local B1+-field inhomogeneities, which hampered stenosis quantification in severe cases. In contrast, B1-shimming and pTX provided superior image homogeneity. Compared with a conventional sTX coil higher grade stenoses (60% and 80%) were graded significantly (p<0.01) more precise. Mild to moderate grade stenoses did not show significant differences. Overall, SNR was distinctly higher with B1-shimming and pTX than with the conventional sTX coil (inside the stenosis phantoms 14%, outside the phantoms 32%). Both full and half concentration (10.2 mM and 5.1 mM) of a conventional Gd-DOTA dose for humans were equally suitable for stenosis evaluation in this phantom study.<h4>Conclusions</h4>B1-shimming and pTX at 7 T can distinctly improve image homogeneity and therefore provide considerably more accurate MR image analysis, which is beneficial for imaging of small vessel structures. |
first_indexed | 2024-12-12T00:45:53Z |
format | Article |
id | doaj.art-5544386ee28c43e09f1aec32e1c7c01f |
institution | Directory Open Access Journal |
issn | 1932-6203 |
language | English |
last_indexed | 2024-12-12T00:45:53Z |
publishDate | 2022-01-01 |
publisher | Public Library of Science (PLoS) |
record_format | Article |
series | PLoS ONE |
spelling | doaj.art-5544386ee28c43e09f1aec32e1c7c01f2022-12-22T00:44:07ZengPublic Library of Science (PLoS)PLoS ONE1932-62032022-01-01176e027068910.1371/journal.pone.0270689Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission.Stefan HerzMaria R StefanescuDavid LohrPatrick VogelAleksander KosmalaMaxim TerekhovAndreas M WengJan-Peter GrunzThorsten A BleyLaura M Schreiber<h4>Background</h4>To investigate the effects of B1-shimming and radiofrequency (RF) parallel transmission (pTX) on the visualization and quantification of the degree of stenosis in a coronary artery phantom using 7 Tesla (7 T) magnetic resonance imaging (MRI).<h4>Methods</h4>Stenosis phantoms with different grades of stenosis (0%, 20%, 40%, 60%, 80%, and 100%; 5 mm inner vessel diameter) were produced using 3D printing (clear resin). Phantoms were imaged with four different concentrations of diluted Gd-DOTA representing established arterial concentrations after intravenous injection in humans. Samples were centrally positioned in a thorax phantom of 30 cm diameter filled with a custom-made liquid featuring dielectric properties of muscle tissue. MRI was performed on a 7 T whole-body system. 2D-gradient-echo sequences were acquired with an 8-channel transmit 16-channel receive (8 Tx / 16 Rx) cardiac array prototype coil with and without pTX mode. Measurements were compared to those obtained with identical scan parameters using a commercially available 1 Tx / 16 Rx single transmit coil (sTX). To assess reproducibility, measurements (n = 15) were repeated at different horizontal angles with respect to the B0-field.<h4>Results</h4>B1-shimming and pTX markedly improved flip angle homogeneity across the thorax phantom yielding a distinctly increased signal-to-noise ratio (SNR) averaged over a whole slice relative to non-manipulated RF fields. Images without B1-shimming showed shading artifacts due to local B1+-field inhomogeneities, which hampered stenosis quantification in severe cases. In contrast, B1-shimming and pTX provided superior image homogeneity. Compared with a conventional sTX coil higher grade stenoses (60% and 80%) were graded significantly (p<0.01) more precise. Mild to moderate grade stenoses did not show significant differences. Overall, SNR was distinctly higher with B1-shimming and pTX than with the conventional sTX coil (inside the stenosis phantoms 14%, outside the phantoms 32%). Both full and half concentration (10.2 mM and 5.1 mM) of a conventional Gd-DOTA dose for humans were equally suitable for stenosis evaluation in this phantom study.<h4>Conclusions</h4>B1-shimming and pTX at 7 T can distinctly improve image homogeneity and therefore provide considerably more accurate MR image analysis, which is beneficial for imaging of small vessel structures.https://doi.org/10.1371/journal.pone.0270689 |
spellingShingle | Stefan Herz Maria R Stefanescu David Lohr Patrick Vogel Aleksander Kosmala Maxim Terekhov Andreas M Weng Jan-Peter Grunz Thorsten A Bley Laura M Schreiber Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. PLoS ONE |
title | Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. |
title_full | Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. |
title_fullStr | Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. |
title_full_unstemmed | Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. |
title_short | Effects of image homogeneity on stenosis visualization at 7 T in a coronary artery phantom study: With and without B1-shimming and parallel transmission. |
title_sort | effects of image homogeneity on stenosis visualization at 7 t in a coronary artery phantom study with and without b1 shimming and parallel transmission |
url | https://doi.org/10.1371/journal.pone.0270689 |
work_keys_str_mv | AT stefanherz effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT mariarstefanescu effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT davidlohr effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT patrickvogel effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT aleksanderkosmala effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT maximterekhov effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT andreasmweng effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT janpetergrunz effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT thorstenabley effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission AT lauramschreiber effectsofimagehomogeneityonstenosisvisualizationat7tinacoronaryarteryphantomstudywithandwithoutb1shimmingandparalleltransmission |