Ascorbate Oxidase-Based Amperometric Biosensor for L-Ascorbic Acid Determination in Beverages

A novel biosensor for l-ascorbic acid determination in different beverages was elaborated. The ascorbate oxidase enzyme (AAO) from Cucurbita sp., EC 1.10.3.3, was immobilized on a screen-printed carbon electrode with poly(ethylene glycol) (400) diglycidyl ether (PEGDGE) as a crosslinking agent. The...

Full description

Bibliographic Details
Main Authors: Gábor Csiffáry, Péter Fűtő, Nóra Adányi, Attila Kiss
Format: Article
Language:English
Published: University of Zagreb Faculty of Food Technology and Biotechnology 2016-01-01
Series:Food Technology and Biotechnology
Subjects:
Online Access:https://hrcak.srce.hr/file/228597
Description
Summary:A novel biosensor for l-ascorbic acid determination in different beverages was elaborated. The ascorbate oxidase enzyme (AAO) from Cucurbita sp., EC 1.10.3.3, was immobilized on a screen-printed carbon electrode with poly(ethylene glycol) (400) diglycidyl ether (PEGDGE) as a crosslinking agent. The standards and samples were measured first with a blank electrode. An inert protein, bovine serum albumin (BSA), was immobilized on the surface of this electrode with PEGDGE. The BSA mass was equivalent to the mass of 10 U of AAO enzyme immobilized on the electrodes (0.021 mg). The linear measuring range for l-ascorbic acid was between 5 and 150 μmol/L. As l-ascorbic acid is a vital vitamin and a common antioxidant used in food industry, fruit juices and vitamin C effervescent tablets were examined. The results were compared to HPLC measurements.
ISSN:1330-9862
1334-2606