An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring

One potential method to estimate noninvasive cuffless blood pressure (BP) is through measurement of pulse wave velocity (PWV), which can be characterized by measuring the distance and the transit time of the pulse between two arterial sites. To obtain the pulse waveform, bioimpedance (BI) measuremen...

Full description

Bibliographic Details
Main Authors: Toan Huu Huynh, Roozbeh Jafari, Wan-Young Chung
Format: Article
Language:English
Published: MDPI AG 2018-06-01
Series:Sensors
Subjects:
Online Access:http://www.mdpi.com/1424-8220/18/7/2095
_version_ 1811186051703635968
author Toan Huu Huynh
Roozbeh Jafari
Wan-Young Chung
author_facet Toan Huu Huynh
Roozbeh Jafari
Wan-Young Chung
author_sort Toan Huu Huynh
collection DOAJ
description One potential method to estimate noninvasive cuffless blood pressure (BP) is through measurement of pulse wave velocity (PWV), which can be characterized by measuring the distance and the transit time of the pulse between two arterial sites. To obtain the pulse waveform, bioimpedance (BI) measurement is a promising approach because it continuously reflects the change in BP through the change in the arterial cross-sectional area. Several studies have investigated BI channels in a vertical direction with electrodes located along the wrist and the finger to calculate PWV and convert to BP; however, the measurement systems were relatively large in size. In order to reduce the total device size for use in a PWV-based BP smartwatch, this study proposes and examines a horizontal BI structure. The BI device is also designed to apply in a very small body area. Our proposed structure is based on two sets of four-electrode BI interface attached around the wrist. The effectiveness of our system and approach is evaluated on 15 human subjects; the PWV values are obtained with various distances between two BI channels to assess the efficacy. The results show that our BI system can monitor pulse rate efficiently in only a 0.5 × 1.75 cm2 area of the body. The correlation of pulse rate from the proposed design against the reference is 0.98 ± 0.07 (p < 0.001). Our structure yields higher detection ratios for PWV measurements of 99.0 ± 2.2%, 99.0 ± 2.1%, and 94.8 ± 3.7% at 1, 2, and 3 cm between two BI channels, respectively. The measured PWVs correlate well with the BP standard device at 0.81 ± 0.08 and 0.84 ± 0.07 with low root-mean-squared-errors at 7.47 ± 2.15 mmHg and 5.17 ± 1.81 mmHg for SBP and DBP, respectively. Our results inform future designs of smart watches capable of measuring blood pressure.
first_indexed 2024-04-11T13:39:45Z
format Article
id doaj.art-5cf2b3d954b3405e99112651e92b70ba
institution Directory Open Access Journal
issn 1424-8220
language English
last_indexed 2024-04-11T13:39:45Z
publishDate 2018-06-01
publisher MDPI AG
record_format Article
series Sensors
spelling doaj.art-5cf2b3d954b3405e99112651e92b70ba2022-12-22T04:21:18ZengMDPI AGSensors1424-82202018-06-01187209510.3390/s18072095s18072095An Accurate Bioimpedance Measurement System for Blood Pressure MonitoringToan Huu Huynh0Roozbeh Jafari1Wan-Young Chung2Department of Electronic Engineering, Pukyong National University, Busan 48513, KoreaDepartments of Biomedical Engineering, Computer Science and Engineering, and Electrical and Computer Engineering, Texas A&M University, College Station, TX 77843, USADepartment of Electronic Engineering, Pukyong National University, Busan 48513, KoreaOne potential method to estimate noninvasive cuffless blood pressure (BP) is through measurement of pulse wave velocity (PWV), which can be characterized by measuring the distance and the transit time of the pulse between two arterial sites. To obtain the pulse waveform, bioimpedance (BI) measurement is a promising approach because it continuously reflects the change in BP through the change in the arterial cross-sectional area. Several studies have investigated BI channels in a vertical direction with electrodes located along the wrist and the finger to calculate PWV and convert to BP; however, the measurement systems were relatively large in size. In order to reduce the total device size for use in a PWV-based BP smartwatch, this study proposes and examines a horizontal BI structure. The BI device is also designed to apply in a very small body area. Our proposed structure is based on two sets of four-electrode BI interface attached around the wrist. The effectiveness of our system and approach is evaluated on 15 human subjects; the PWV values are obtained with various distances between two BI channels to assess the efficacy. The results show that our BI system can monitor pulse rate efficiently in only a 0.5 × 1.75 cm2 area of the body. The correlation of pulse rate from the proposed design against the reference is 0.98 ± 0.07 (p < 0.001). Our structure yields higher detection ratios for PWV measurements of 99.0 ± 2.2%, 99.0 ± 2.1%, and 94.8 ± 3.7% at 1, 2, and 3 cm between two BI channels, respectively. The measured PWVs correlate well with the BP standard device at 0.81 ± 0.08 and 0.84 ± 0.07 with low root-mean-squared-errors at 7.47 ± 2.15 mmHg and 5.17 ± 1.81 mmHg for SBP and DBP, respectively. Our results inform future designs of smart watches capable of measuring blood pressure.http://www.mdpi.com/1424-8220/18/7/2095blood pressurebioimpedancepulse wave velocitypulse transit timewearable structure
spellingShingle Toan Huu Huynh
Roozbeh Jafari
Wan-Young Chung
An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
Sensors
blood pressure
bioimpedance
pulse wave velocity
pulse transit time
wearable structure
title An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
title_full An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
title_fullStr An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
title_full_unstemmed An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
title_short An Accurate Bioimpedance Measurement System for Blood Pressure Monitoring
title_sort accurate bioimpedance measurement system for blood pressure monitoring
topic blood pressure
bioimpedance
pulse wave velocity
pulse transit time
wearable structure
url http://www.mdpi.com/1424-8220/18/7/2095
work_keys_str_mv AT toanhuuhuynh anaccuratebioimpedancemeasurementsystemforbloodpressuremonitoring
AT roozbehjafari anaccuratebioimpedancemeasurementsystemforbloodpressuremonitoring
AT wanyoungchung anaccuratebioimpedancemeasurementsystemforbloodpressuremonitoring
AT toanhuuhuynh accuratebioimpedancemeasurementsystemforbloodpressuremonitoring
AT roozbehjafari accuratebioimpedancemeasurementsystemforbloodpressuremonitoring
AT wanyoungchung accuratebioimpedancemeasurementsystemforbloodpressuremonitoring