Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy

Purpose: To compare the sensitivity of ArcCHECK (AC), portal dosimetry (PD), and an in-house logfile-based system (LF) to multileaf collimators (MLC) aperture errors and the ability to identify these errors. Methods and Materials: For 12 retrospective original head and neck volumetric modulated arc...

Full description

Bibliographic Details
Main Authors: Jinyou Hu MS, Shaoxian Gu MS, Ningyu Wang MS, Fengjie Cui MS, Shengyuan Zhang MS, Chuou Yin MS, Yunzhu Cai MS, Chengjun Gou PhD, Lian Zou PhD, Zhangwen Wu PhD
Format: Article
Language:English
Published: SAGE Publishing 2022-09-01
Series:Technology in Cancer Research & Treatment
Online Access:https://doi.org/10.1177/15330338221114499
_version_ 1811267878239862784
author Jinyou Hu MS
Shaoxian Gu MS
Ningyu Wang MS
Fengjie Cui MS
Shengyuan Zhang MS
Chuou Yin MS
Yunzhu Cai MS
Chengjun Gou PhD
Lian Zou PhD
Zhangwen Wu PhD
author_facet Jinyou Hu MS
Shaoxian Gu MS
Ningyu Wang MS
Fengjie Cui MS
Shengyuan Zhang MS
Chuou Yin MS
Yunzhu Cai MS
Chengjun Gou PhD
Lian Zou PhD
Zhangwen Wu PhD
author_sort Jinyou Hu MS
collection DOAJ
description Purpose: To compare the sensitivity of ArcCHECK (AC), portal dosimetry (PD), and an in-house logfile-based system (LF) to multileaf collimators (MLC) aperture errors and the ability to identify these errors. Methods and Materials: For 12 retrospective original head and neck volumetric modulated arc therapy (VMAT) plans, MLC aperture errors of ± 0.4mm, ± 1.2mm, ± 2mm, and ± 3mm were introduced for each plan, resulting in 96 plans with errors. AC, PD, and LF were used for the gamma evaluation at 3%/3mm, 3%/2mm, and 2%/2mm criteria. Gradient analysis was used to evaluate the sensitivity to MLC aperture errors. The area under the curve (AUC) obtained from the receiver operating characteristic (ROC) curve was used to evaluate the ability to identify MLC aperture errors and dose errors, and the optimal cut-off value to identify the error was obtained. Results: The gamma pass rate (%GP) of LF had the smallest descent gradient as the MLC error increases in any case. The descent gradient of PD was larger than AC, except for the case at the 2%/2mm criteria. For the 3%/3mm criteria, the MLC aperture errors that can be perfectly identified by AC, PD, and LF were ± 3mm, ± 2mm, and ± 1.2mm, respectively, and the average percent dose error (%DEs) of dose metrics in targets that can be perfectly identified were 4% to 5%, 3% to 4%, and 2% to 3%, respectively. For the 3%/2mm criteria, the errors that AC, PD, and LF can perfectly identify were the same as the 3%/3mm criteria. For the 2%/2mm criteria, AC can perfectly identify the MLC error of ± 2mm and the %DE of 3% to 4%. PD and LF can identify the MLC error of ± 1.2mm and the %DE of 2% to 3%. Conclusion: Different patient-specific quality assurance (PSQA) systems have different sensitivity and recognition abilities to MLC aperture errors. Institutions should formulate their own customized %GP limits based on their PSQA process through ROC or other methods.
first_indexed 2024-04-12T21:10:27Z
format Article
id doaj.art-5e5f4bb62fb449e58148766f686010c5
institution Directory Open Access Journal
issn 1533-0338
language English
last_indexed 2024-04-12T21:10:27Z
publishDate 2022-09-01
publisher SAGE Publishing
record_format Article
series Technology in Cancer Research & Treatment
spelling doaj.art-5e5f4bb62fb449e58148766f686010c52022-12-22T03:16:36ZengSAGE PublishingTechnology in Cancer Research & Treatment1533-03382022-09-012110.1177/15330338221114499Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc TherapyJinyou Hu MS0Shaoxian Gu MS1Ningyu Wang MS2Fengjie Cui MS3Shengyuan Zhang MS4Chuou Yin MS5Yunzhu Cai MS6Chengjun Gou PhD7Lian Zou PhD8Zhangwen Wu PhD9 Cancer Center, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, China Cancer Center, , Chengdu, China Key Laboratory of Radiation Physics and Technology of the Ministry of Education, Institute of Nuclear Science and Technology, , Chengdu, ChinaPurpose: To compare the sensitivity of ArcCHECK (AC), portal dosimetry (PD), and an in-house logfile-based system (LF) to multileaf collimators (MLC) aperture errors and the ability to identify these errors. Methods and Materials: For 12 retrospective original head and neck volumetric modulated arc therapy (VMAT) plans, MLC aperture errors of ± 0.4mm, ± 1.2mm, ± 2mm, and ± 3mm were introduced for each plan, resulting in 96 plans with errors. AC, PD, and LF were used for the gamma evaluation at 3%/3mm, 3%/2mm, and 2%/2mm criteria. Gradient analysis was used to evaluate the sensitivity to MLC aperture errors. The area under the curve (AUC) obtained from the receiver operating characteristic (ROC) curve was used to evaluate the ability to identify MLC aperture errors and dose errors, and the optimal cut-off value to identify the error was obtained. Results: The gamma pass rate (%GP) of LF had the smallest descent gradient as the MLC error increases in any case. The descent gradient of PD was larger than AC, except for the case at the 2%/2mm criteria. For the 3%/3mm criteria, the MLC aperture errors that can be perfectly identified by AC, PD, and LF were ± 3mm, ± 2mm, and ± 1.2mm, respectively, and the average percent dose error (%DEs) of dose metrics in targets that can be perfectly identified were 4% to 5%, 3% to 4%, and 2% to 3%, respectively. For the 3%/2mm criteria, the errors that AC, PD, and LF can perfectly identify were the same as the 3%/3mm criteria. For the 2%/2mm criteria, AC can perfectly identify the MLC error of ± 2mm and the %DE of 3% to 4%. PD and LF can identify the MLC error of ± 1.2mm and the %DE of 2% to 3%. Conclusion: Different patient-specific quality assurance (PSQA) systems have different sensitivity and recognition abilities to MLC aperture errors. Institutions should formulate their own customized %GP limits based on their PSQA process through ROC or other methods.https://doi.org/10.1177/15330338221114499
spellingShingle Jinyou Hu MS
Shaoxian Gu MS
Ningyu Wang MS
Fengjie Cui MS
Shengyuan Zhang MS
Chuou Yin MS
Yunzhu Cai MS
Chengjun Gou PhD
Lian Zou PhD
Zhangwen Wu PhD
Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
Technology in Cancer Research & Treatment
title Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
title_full Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
title_fullStr Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
title_full_unstemmed Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
title_short Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
title_sort sensitivity of three patient specific quality assurance systems to mlc aperture errors with volumetric modulated arc therapy
url https://doi.org/10.1177/15330338221114499
work_keys_str_mv AT jinyouhums sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT shaoxiangums sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT ningyuwangms sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT fengjiecuims sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT shengyuanzhangms sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT chuouyinms sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT yunzhucaims sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT chengjungouphd sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT lianzouphd sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy
AT zhangwenwuphd sensitivityofthreepatientspecificqualityassurancesystemstomlcapertureerrorswithvolumetricmodulatedarctherapy