TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY

To ensure the safety of radiation oncology patients needed to provide a consistent functional characteristics of the medical linear electron accelerators, which affect the accuracy of dose delivery. To this end, their quality control procedures, which include the calibration of radiation output of t...

Full description

Bibliographic Details
Main Authors: Y. V. Tsitovich, M. G. Kiselev
Format: Article
Language:English
Published: Belarusian National Technical University 2015-12-01
Series:Pribory i Metody Izmerenij
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/228
_version_ 1797874819614638080
author Y. V. Tsitovich
M. G. Kiselev
author_facet Y. V. Tsitovich
M. G. Kiselev
author_sort Y. V. Tsitovich
collection DOAJ
description To ensure the safety of radiation oncology patients needed to provide a consistent functional characteristics of the medical linear electron accelerators, which affect the accuracy of dose delivery. To this end, their quality control procedures, which include the calibration of radiation output of the linear accelerator, the error in determining the dose reference value during which must not exceed 2 %, is provided. The aim is to develop a methodology for determining the error in determining this value, depending on the mechanical charachteristics of the linac’s gantry. To achieve the objectives have been carried out dosimetric measurements of Trilogy S/N 3567 linac dose distributions, on the basis of which dose errors depending on the accuracy setting the zero position of the gantry and the deviation of the gantry rotation isocenter were obtained. It was found that the greatest impact on the value of the error has gantry rotation isocenter deviation in a plane perpendicular to the plane of incidence of the radiation beam (up to 3,64% for the energy of 6 MeV). Dose errors caused by tilting the gantry and its isocenter deviation in the plane of incidence of the beam were highest for 18 MeV energy and reached –0,7 % and –0,9 % respectively. Thus, it is possible to express the results of periodic quality control of the linear accelerator ganty in terms of dose and use them to conduct a comprehensive assessment of the possibility of clinical use of a linear accelerator for oncology patients irradiation in case of development of techniques that allow to analyze the influence of the rest of its technical and dosimetric parameters for error in dose.
first_indexed 2024-04-10T01:38:05Z
format Article
id doaj.art-600884358cea4f8289f4b046cc962155
institution Directory Open Access Journal
issn 2220-9506
2414-0473
language English
last_indexed 2024-04-10T01:38:05Z
publishDate 2015-12-01
publisher Belarusian National Technical University
record_format Article
series Pribory i Metody Izmerenij
spelling doaj.art-600884358cea4f8289f4b046cc9621552023-03-13T09:14:46ZengBelarusian National Technical UniversityPribory i Metody Izmerenij2220-95062414-04732015-12-0162230238220TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRYY. V. Tsitovich0M. G. Kiselev1Республиканский научно-практический центр онкологии и медицинской радиологии им. Н.Н.АлександроваБелорусский национальный технический университетTo ensure the safety of radiation oncology patients needed to provide a consistent functional characteristics of the medical linear electron accelerators, which affect the accuracy of dose delivery. To this end, their quality control procedures, which include the calibration of radiation output of the linear accelerator, the error in determining the dose reference value during which must not exceed 2 %, is provided. The aim is to develop a methodology for determining the error in determining this value, depending on the mechanical charachteristics of the linac’s gantry. To achieve the objectives have been carried out dosimetric measurements of Trilogy S/N 3567 linac dose distributions, on the basis of which dose errors depending on the accuracy setting the zero position of the gantry and the deviation of the gantry rotation isocenter were obtained. It was found that the greatest impact on the value of the error has gantry rotation isocenter deviation in a plane perpendicular to the plane of incidence of the radiation beam (up to 3,64% for the energy of 6 MeV). Dose errors caused by tilting the gantry and its isocenter deviation in the plane of incidence of the beam were highest for 18 MeV energy and reached –0,7 % and –0,9 % respectively. Thus, it is possible to express the results of periodic quality control of the linear accelerator ganty in terms of dose and use them to conduct a comprehensive assessment of the possibility of clinical use of a linear accelerator for oncology patients irradiation in case of development of techniques that allow to analyze the influence of the rest of its technical and dosimetric parameters for error in dose.https://pimi.bntu.by/jour/article/view/228лучевая терапиякалибровка радиационного выхода линейного ускорителяошибки в опорном значении дозыштатив линейного ускорителя
spellingShingle Y. V. Tsitovich
M. G. Kiselev
TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
Pribory i Metody Izmerenij
лучевая терапия
калибровка радиационного выхода линейного ускорителя
ошибки в опорном значении дозы
штатив линейного ускорителя
title TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
title_full TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
title_fullStr TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
title_full_unstemmed TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
title_short TECHNIQUE OF ESTIMATION OF ERROR IN THE REFERENCE VALUE OF THE DOSE DURING THE LINEAR ACCELERATOR RADIATION OUTPUT CALIBRATION PROCEDURE. PART 1. DEPENDANCE OF THE MECHANICAL PARAMETERS OF LINAC’S GANTRY
title_sort technique of estimation of error in the reference value of the dose during the linear accelerator radiation output calibration procedure part 1 dependance of the mechanical parameters of linac s gantry
topic лучевая терапия
калибровка радиационного выхода линейного ускорителя
ошибки в опорном значении дозы
штатив линейного ускорителя
url https://pimi.bntu.by/jour/article/view/228
work_keys_str_mv AT yvtsitovich techniqueofestimationoferrorinthereferencevalueofthedoseduringthelinearacceleratorradiationoutputcalibrationprocedurepart1dependanceofthemechanicalparametersoflinacsgantry
AT mgkiselev techniqueofestimationoferrorinthereferencevalueofthedoseduringthelinearacceleratorradiationoutputcalibrationprocedurepart1dependanceofthemechanicalparametersoflinacsgantry